510(k) Report

PMA · Approved

Recalled · Class II · 2026

Clarion Multi-Strategy Cochlear Implant

P960058PMAApproved Jun 26, 1997178-day reviewClass IIIRecalls 17

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Advanced Bionics · MCM Implant, Cochlear

Approved on Jun 26, 1997 through premarket approval, which rests on the device’s own clinical evidence after a 178-day review. It has been recalled 17 times.

Decided
Jun 26, 1997
Review
178 days
Class
Class III
Product code
MCM
Recalls
17

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P960058 has 149 supplements since 1997: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

1997: 11998: 41999: 52000: 42001: 12002: 92003: 42004: 62005: 62006: 52007: 32008: 42009: 52010: 62011: 32012: 62013: 72014: 52015: 72016: 22017: 72018: 32019: 152020: 62021: 22022: 32023: 52024: 42025: 42026: 7
1997high 15 (2019) · low 1 (1997)2026
View FIG. 1 as a table
YearSupplements approved
19971
19984
19995
20004
20011
20029
20034
20046
20056
20065
20073
20084
20095
20106
20113
20126
20137
20145
20157
20162
20177
20183
201915
20206
20212
20223
20235
20244
20254
20267
FIG. 1 — Supplements to P960058, 1997–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MCM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 30, 2026Class IIZ-2283-2026Under Investigation by firmDue to incorrect shelf-life expiration date.
Mar 23, 2026Class IIZ-1950-2026Under Investigation by firmThe non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by not requiring the use of a tool to replace the button/coin cell battery.
Mar 5, 2025Class IIZ-1813-2025Under Investigation by firmDue to incorrect the battery charger being packaged and distributed. Product labeling does not match the intended product as specified.
Feb 17, 2020Class IIZ-2033-2020Device Designhearing performance degradation due to body-fluid entering the device.
Feb 17, 2020Class IIZ-2032-2020Device Designhearing performance degradation due to body-fluid entering the device.
Apr 4, 2019Class IIIZ-1441-2020Software in the Use EnvironmentThe manufacturer received complaints that customers were attempting to install the fitting software using the supplied USB drive, an error message was received, and installation failed. The failed installation has not impact to the…
Mar 18, 2019Class IIIZ-1259-2020OtherThree materials used in the makeup of the implant were mistakenly left off the list of bio-compatible materials in the firm's previously approved submission to the Pharmaceuticals and Medical Devices Agency (PMDA) of Japan.
Feb 6, 2019Class IIZ-1085-2019Process controlThe sound processors were loaded with the incorrect firmware.
Feb 6, 2019Class IIZ-1084-2019Process controlThe sound processors were loaded with the incorrect firmware.
Feb 6, 2019Class IIZ-1083-2019Process controlThe sound processors were loaded with the incorrect firmware.
Jun 10, 2009Not classifiedZ-2039-2009Process controlThe recall was initiated after Advanced Bionics became aware of fourteen (14) devices manufactured between May 27, 2005 and June 30, 2005 that are confirmed to have or potentially have an issue related to the internal magnet. These…
Jul 24, 2008Not classifiedZ-0956-2009Employee errorMistuned headpiece - Advanced Bionics recently became aware of a tuning issue with some Platinum Series and S-Series headpieces shipped from May 2007 through April 2008. Tuning refers to the process through which the headpiece's…
Jul 24, 2008Not classifiedZ-0958-2009Employee errorMistuned headpiece - Advanced Bionics recently became aware of a tuning issue with some Platinum Series and S-Series headpieces shipped from May 2007 through April 2008. Tuning refers to the process through which the headpiece's…
Jul 24, 2008Not classifiedZ-0959-2009Employee errorMistuned headpiece - Advanced Bionics recently became aware of a tuning issue with some Platinum Series and S-Series headpieces shipped from May 2007 through April 2008. Tuning refers to the process through which the headpiece's…
Jul 24, 2008Not classifiedZ-0957-2009Employee errorMistuned headpiece - Advanced Bionics recently became aware of a tuning issue with some Platinum Series and S-Series headpieces shipped from May 2007 through April 2008. Tuning refers to the process through which the headpiece's…
Mar 8, 2006Not classifiedZ-0759-06OtherCertain HiRes 90K cochlear implants could fail as a result of elevated moisture levels. The unimplanted HiRes 90K implants listed are being removed from the market. Patients are to be followed for signs of device failure: intermittent…
Mar 8, 2006Not classifiedZ-0758-06OtherCertain HiRes 90K cochlear implants could fail as a result of elevated moisture levels. The unimplanted HiRes 90K implants listed are being removed from the market. Patients are to be followed for signs of device failure: intermittent…

No ancestor or descendant has a Class I recall.

How does its category compare?

All MCM devices

MCM Implant, Cochlear: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MCM

17in 2025

2007: 172008: 172009: 192010: 212011: 272012: 212013: 282014: 232015: 232016: 252017: 312018: 202019: 352020: 252021: 202022: 242023: 182024: 172025: 172026 (partial year): 13
2007high 35 (2019) · low 17 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MCM

Too few MCM 510(k)s in any year since 2007 to give a median.

Recalls per year · MCM

2in 2020

2007: 12008: 42009: 22010: 22011: 12020: 22026 (partial year): 2
2007high 4 (2008) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MCM

5,418in 2025

2007: 1,6392008: 1,7932009: 1,8972010: 1,7732011: 2,2932012: 3,0782013: 2,6892014: 2,6762015: 3,1052016: 2,7852017: 3,1222018: 3,0542019: 3,4052020: 4,4032021: 5,2602022: 5,5022023: 5,7272024: 5,6022025: 5,418
2007high 5,727 (2023) · low 1,639 (2007)2025

MAUDE reports naming product code MCM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200717—11,639
200817—41,793
200919—21,897
201021—21,773
201127—12,293
201221—03,078
201328—02,689
201423—02,676
201523—03,105
201625—02,785
201731—03,122
201820—03,054
201935—03,405
202025—24,403
202120—05,260
202224—05,502
202318—05,727
202417—05,602
202517—05,418
202613—2—
Cite this page

510(k) Report. “P960058: Clarion Multi-Strategy Cochlear Implant.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P960058

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026