VNS Therapy System

P970003PMAApproved Jul 16, 1997170-day reviewClass IIIRecalls 30 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

LivaNova USA, Inc. · LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy

Approved on Jul 16, 1997 through premarket approval, which rests on the device’s own clinical evidence after a 170-day review. It has a Class I recall. It has been recalled 30 times, including a Class I recall.

Decided
Jul 16, 1997
Review
170 days
Class
Class III
Product code
LYJ
Recalls
30 (1 Class I)

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P970003 has 238 supplements since 1997: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

1997: 41998: 71999: 102000: 32001: 62002: 62003: 62004: 32005: 72006: 102007: 212008: 92009: 32010: 102011: 262012: 102013: 132014: 62015: 142016: 182017: 92018: 32019: 72020: 62021: 12022: 32023: 52024: 32025: 62026: 3
1997high 26 (2011) · low 1 (2021)2026
View FIG. 1 as a table
YearSupplements approved
19974
19987
199910
20003
20016
20026
20036
20043
20057
200610
200721
20089
20093
201010
201126
201210
201313
20146
201514
201618
20179
20183
20197
20206
20211
20223
20235
20243
20256
20263
FIG. 1 — Supplements to P970003, 1997–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 3, 2025Class IIZ-1068-2025Device DesignTheir is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.
Jan 3, 2025Class IIZ-1067-2025Device DesignTheir is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.
Dec 12, 2024Class IIZ-1072-2025Device DesignVagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or…
Dec 2, 2021Class IIZ-0532-2022Process change controlA manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to…
Sep 28, 2020Class IIZ-0462-2021Process controlDue to an extra digit being inadvertently added to the serial number of a subset of implantable pulse generators, they become not compatible with therapy programmers with software version 11.0.4.
Feb 4, 2020Class IIZ-1563-2020Software Design ChangeFalse positive warning may occur after: 1) VNS Generator interrogated at 0mA normal output current, 2) Generator programmed to non-0mA output current, 3) In-session re-interrogation performed. Users instructed to lower output current and…
Aug 22, 2019Class IZ-0561-2020Device DesignSentiva generators may reset and become disabled within 60 days of enabling. Physicians can re-enable but devices will continue to be susceptible to resets. If a device experiences this issue, this may lead to additional surgery…
Aug 13, 2019Class IIZ-1118-2020Process controlFirm identified a subset of its generators that were sterilized one additional sterilization cycle, which does not meet the firm's quality specifications.
Nov 16, 2018Class IIZ-0346-2020Software designLead impedance values reported by the affected VNS generator will be higher compared to those reported by previous models. This is due to a change in the timing of when affected VNS generator takes the lead impedance measurement during…
Nov 16, 2018Class IIZ-0663-2020Software designThis recall is being initiated due to reports that that the therapy programming tablet with software version 1.5 errantly performs a normal mode diagnostic test instead of the selected system diagnostic test on Model 102 and Model 102R…
Nov 16, 2018Class IIZ-0869-2020Manufacturing material removalThis recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106 generators. *Note this…
Nov 16, 2018Class IIZ-0868-2020Manufacturing material removalThis recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106 generators. *Note this…
Jan 24, 2018Class IIZ-2572-2018Software design (manufacturing process)Unintended warning message displayed on generators programmed with a Model 3000 v.1.0.2.2 programmer.
Dec 2, 2017Class IIZ-0555-2018Employee errorTwo Model 3000 Programmers were distributed in error by prior to FDA approval of version 1.0.2.2 software.
Nov 13, 2017Class IIZ-2255-2018Device DesignCertain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than…
Dec 18, 2015Class IIZ-0659-2016Device DesignRecall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled.
Nov 18, 2015Class IIZ-0645-2016Manufacturing material removalCertain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit the potential for decreased battery longevity.
Oct 7, 2015Class IIZ-0280-2016Manufacturing material removalCertain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit potential for decreased battery longevity.
Nov 20, 2014Class IIZ-0858-2015Labeling mix-upsThe recalled product was distributed with an incorrect serial number printed on the device's label.
Aug 12, 2011Not classifiedZ-0210-2012Nonconforming Material/ComponentThe devices are being recalled because the output current delivered to the vagus nerve is less than the design intent and there is a potential charge imbalance at the lead cathode and generator-can during stimulation.
May 6, 2011Not classifiedZ-0005-2012Software design (manufacturing process)An investigation was initiated based on a report from the field in which an Intensive Follow-up Indicator (IFI) message was unexpectedly received by a medical professional when using Model 250 version 8.0 software to interrogate a…
May 6, 2011Not classifiedZ-0006-2012Software design (manufacturing process)An investigation was initiated based on a report from the field in which an Intensive Follow-up Indicator (IFI) message was unexpectedly received by a medical professional when using Model 250 version 8.0 software to interrogate a…
Apr 6, 2010Not classifiedZ-1591-2010Software designBattery life projection is inaccurate.
Feb 7, 2008Not classifiedZ-0994-2009Software designReset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference resulting in the loss of stimulation.
Feb 7, 2008Not classifiedZ-0995-2009Software designReset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference, resulting in the loss of stimulation.
Aug 2, 2007Not classifiedZ-0171-2008Process controlDissolution/Fractures to the leads of the VNS Therapy System
Aug 2, 2007Not classifiedZ-0172-2008Process controlDissolution/Fractures to the leads of the VNS Therapy System
Aug 2, 2007Not classifiedZ-0173-2008Process controlDissolution/Fractures to the leads of the VNS Therapy System
Oct 17, 2006Not classifiedZ-0342-2007OtherDuring programming, pulse generator may be inadvertently set to 8.0 mA output, regardless of the mA range selected by the clinician
Oct 17, 2006Not classifiedZ-0341-2007OtherDuring programming, pulse generator may be inadvertently set to 8.0 mA output, regardless of the mA range selected by the clinician

No ancestor or descendant has a Class I recall.

How does its category compare?

All LYJ devices

LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LYJ

6in 2025

2007: 212008: 92009: 32010: 102011: 242015: 22016: 142017: 92018: 32019: 72020: 62021: 12022: 32023: 52024: 32025: 62026 (partial year): 3
2007high 24 (2011) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LYJ

Too few LYJ 510(k)s in any year since 2007 to give a median.

Recalls per year · LYJ

2in 2025

2007: 42008: 22009: 22010: 12011: 32014: 12015: 32017: 12018: 42019: 1, 1 Class I2020: 22021: 12025: 2
2007high 4 (2007) · low 0 (2012)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · LYJ

1,426in 2025

2007: 1,8892008: 2,7582009: 2,6882010: 2,7492011: 3,0422012: 3,3642013: 3,8092014: 3,3652015: 3,2652016: 2,8512017: 2,0742018: 2,3162019: 2,4322020: 1,6882021: 1,7642022: 1,5842023: 1,7642024: 1,5242025: 1,426
2007high 3,809 (2013) · low 1,426 (2025)2025

MAUDE reports naming product code LYJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200721—41,889
20089—22,758
20093—22,688
201010—12,749
201124—33,042
20120—03,364
20130—03,809
20140—13,365
20152—33,265
201614—02,851
20179—12,074
20183—42,316
20197—1 (1)2,432
20206—21,688
20211—11,764
20223—01,584
20235—01,764
20243—01,524
20256—21,426
20263—0—
Cite this page

510(k) Report. “P970003: VNS Therapy System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P970003