510(k) Report

PMA · Approved

Recalled · 2007

Ladarvision Excimer Laser System

P970043PMAApproved Nov 2, 1998423-day reviewClass IIIRecalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Alcon Laboratories · LZS Excimer Laser System

Approved on Nov 2, 1998 through premarket approval, which rests on the device’s own clinical evidence after a 423-day review. It has been recalled 2 times.

Decided
Nov 2, 1998
Review
423 days
Class
Class III
Product code
LZS
Recalls
2

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P970043 has 23 supplements since 1999: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

1999: 42000: 52002: 22003: 32004: 22005: 12006: 6
1999high 6 (2006) · low 0 (2001)2006
View FIG. 1 as a table
YearSupplements approved
19994
20005
20022
20033
20042
20051
20066
FIG. 1 — Supplements to P970043, 1999–2006Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LZS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 21, 2007Not classifiedZ-0854-2007OtherObserved 'central islands' in patients following custom myopia laser procedures using the LADAR6000 Excimer Laser. (On May 11, 2007, Alcon notified consignees that the device software programs for CustomCornea Myopia and CustomCornea…
Aug 1, 2005Not classifiedZ-1147-06OtherAlcon has identified a software error associated with the use of the Measuring Mode (ruler tool) on the LADARVision4000 Excimer Laser Systems. Use of this feature after alignment of the horizontal reference line will negate compensation…

No ancestor or descendant has a Class I recall.

How does its category compare?

All LZS devices

LZS Excimer Laser System: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LZS

6in 2025

2007: 42008: 62009: 32010: 32011: 62012: 62013: 142014: 182015: 142016: 42017: 102018: 112019: 132020: 62021: 72023: 82024: 102025: 62026 (partial year): 2
2007high 18 (2014) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZS

Too few LZS 510(k)s in any year since 2007 to give a median.

Recalls per year · LZS

1in 2021

2007: 32008: 12010: 12011: 112012: 12013: 42021: 1
2007high 11 (2011) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LZS

260in 2025

2007: 5162008: 5882009: 2662010: 6982011: 4632012: 4772013: 9442014: 1,8032015: 1,1462016: 8082017: 1,3602018: 1,6012019: 1,0632020: 5652021: 2662022: 1522023: 1282024: 912025: 260
2007high 1,803 (2014) · low 91 (2024)2025

MAUDE reports naming product code LZS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074—3516
20086—1588
20093—0266
20103—1698
20116—11463
20126—1477
201314—4944
201418—01,803
201514—01,146
20164—0808
201710—01,360
201811—01,601
201913—01,063
20206—0565
20217—1266
20220—0152
20238—0128
202410—091
20256—0260
20262—0—
Cite this page

510(k) Report. “P970043: Ladarvision Excimer Laser System.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P970043

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026