510(k) Report

PMA · Approved

Recalled · Class II · 2020

Ats Open Pivot Bileaflet Heart Valve

P990046PMAApproved Oct 13, 2000442-day reviewClass IIIRecalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Ats Medical, Inc. · LWQ Heart-Valve, Mechanical

Approved on Oct 13, 2000 through premarket approval, which rests on the device’s own clinical evidence after a 442-day review. It has been recalled 2 times.

Decided
Oct 13, 2000
Review
442 days
Class
Class III
Product code
LWQ
Recalls
2

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P990046 has 59 supplements since 2001: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2001: 32002: 22003: 32004: 12006: 12007: 42008: 12009: 32010: 12011: 12012: 42013: 72014: 32015: 32016: 62017: 42018: 12020: 22021: 22022: 52023: 2
2001high 7 (2013) · low 0 (2005)2023
View FIG. 1 as a table
YearSupplements approved
20013
20022
20033
20041
20061
20074
20081
20093
20101
20111
20124
20137
20143
20153
20166
20174
20181
20202
20212
20225
20232
FIG. 1 — Supplements to P990046, 2001–2023Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LWQ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 11, 2020Class IIZ-1533-2020Labeling mix-upsMedtronic received a complaint from China that the label on the box for an Open Pivot Heart Valve was labeled as Aortic, but the product was Mitral. The valve itself was correctly labeled as Mitral.
Feb 18, 2004Not classifiedZ-0634-04OtherNormally functioning prosthesis may have been placed on the valve holder mechanism in reverse position.

No ancestor or descendant has a Class I recall.

How does its category compare?

All LWQ devices

LWQ Heart-Valve, Mechanical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LWQ

9in 2025

2007: 122008: 152009: 82010: 62011: 162012: 212013: 172014: 112015: 142016: 132017: 112018: 82019: 42020: 62021: 92022: 82023: 92024: 62025: 92026 (partial year): 5
2007high 21 (2012) · low 4 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LWQ

Too few LWQ 510(k)s in any year since 2007 to give a median.

Recalls per year · LWQ

1so far in 2026, the most since 2007

2026 (partial year): 1
2007high 1 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LWQ

330in 2025

2007: 882008: 902009: 912010: 1192011: 1392012: 1222013: 1342014: 1342015: 1392016: 2022017: 2992018: 4342019: 3502020: 3722021: 2872022: 2972023: 4552024: 4042025: 330
2007high 455 (2023) · low 88 (2007)2025

MAUDE reports naming product code LWQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712—088
200815—090
20098—091
20106—0119
201116—0139
201221—0122
201317—0134
201411—0134
201514—0139
201613—0202
201711—0299
20188—0434
20194—0350
20206—0372
20219—0287
20228—0297
20239—0455
20246—0404
20259—0330
20265—1—
Cite this page

510(k) Report. “P990046: Ats Open Pivot Bileaflet Heart Valve.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P990046

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026