510(k) Report

PMA · Approved

Recalled · Class II · 2019

Medtronic Mosaic Porcine Bioprosthetic Heart Valve

P990064PMAApproved Jul 14, 2000280-day reviewClass III21 CFR 870.3925Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic, Inc. · DYE Replacement Heart-Valve

Approved on Jul 14, 2000 through premarket approval, which rests on the device’s own clinical evidence after a 280-day review. It has been recalled 3 times.

Decided
Jul 14, 2000
Review
280 days
Class
Class III
Product code
DYE
Recalls
3

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P990064 has 89 supplements since 2000: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2000: 22001: 22002: 22004: 22006: 52007: 52008: 32009: 42010: 72011: 32012: 42013: 82014: 72015: 22016: 32017: 82018: 32019: 32020: 32021: 12022: 12023: 32024: 12025: 52026: 2
2000high 8 (2013) · low 0 (2003)2026
View FIG. 1 as a table
YearSupplements approved
20002
20012
20022
20042
20065
20075
20083
20094
20107
20113
20124
20138
20147
20152
20163
20178
20183
20193
20203
20211
20221
20233
20241
20255
20262
FIG. 1 — Supplements to P990064, 2000–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 21, 2019Class IIZ-1880-2019Labeling mix-upsFifteen affected devices were distributed worldwide, which were mislabeled with an incorrect size. In all cases, the actual surgical heart valves are larger than what was indicated on the boxes and jar labels.
May 21, 2019Class IIZ-1881-2019Labeling mix-upsFifteen affected devices were distributed worldwide, which were mislabeled with an incorrect size. In all cases, the actual surgical heart valves are larger than what was indicated on the boxes and jar labels.
Jan 26, 2009Not classifiedZ-1225-2009StorageMedtronic identified several heart valves (Mosaic Porcine Bioprosthesis with Cinch & Hancock II Bioprosthesis with Cinch) that were inadvertently distributed after exposure to temperatures exceeding the 77 F (25 C) maximum storage…

No ancestor or descendant has a Class I recall.

How does its category compare?

All DYE devices

DYE Replacement Heart-Valve: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DYE

13in 2025

2007: 282008: 342009: 312010: 472011: 222012: 582013: 662014: 382015: 132016: 222017: 342018: 272019: 162020: 132021: 72022: 82023: 112024: 72025: 132026 (partial year): 7
2007high 66 (2013) · low 7 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DYE

Too few DYE 510(k)s in any year since 2007 to give a median.

Recalls per year · DYE

2in 2009

2009: 2
2007high 2 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DYE

3,106in 2025

2007: 192008: 162009: 9102010: 1,0712011: 1,2212012: 1,0152013: 1,2192014: 1,2822015: 1,2422016: 1,7442017: 2,2842018: 3,1662019: 2,8202020: 2,4662021: 2,8332022: 2,5662023: 3,4892024: 3,4352025: 3,106
2007high 3,489 (2023) · low 16 (2008)2025

MAUDE reports naming product code DYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200728—019
200834—016
200931—2910
201047—01,071
201122—01,221
201258—01,015
201366—01,219
201438—01,282
201513—01,242
201622—01,744
201734—02,284
201827—03,166
201916—02,820
202013—02,466
20217—02,833
20228—02,566
202311—03,489
20247—03,435
202513—03,106
20267—0—
Cite this page

510(k) Report. “P990064: Medtronic Mosaic Porcine Bioprosthetic Heart Valve.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P990064

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026