510(k) Report

PMA · Approved

Recalled · 2011

Pathway Anti-HCR-2/Ncu (4B5) Rabbit Monoclonal Primary Antibody

P990081PMAApproved Nov 28, 2000351-day reviewClass IIIRecalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ventana Medical Systems, Inc. · MVC System, Test, Her-2/Neu, Ihc

Approved on Nov 28, 2000 through premarket approval, which rests on the device’s own clinical evidence after a 351-day review. It has been recalled 3 times.

Decided
Nov 28, 2000
Review
351 days
Class
Class III
Product code
MVC
Recalls
3

PMA approvals rest on their own clinical evidence and do not claim a predicate, so this family tree starts here.

P990081 has 57 supplements since 2002: changes to the approved device, its labelling or its manufacturing, each approved separately.

PMA supplements approved per year

2002: 12005: 12006: 12007: 42010: 12011: 22012: 32013: 102014: 62015: 32016: 12017: 22018: 12019: 32020: 22021: 12022: 52023: 22024: 22025: 42026: 2
2002high 10 (2013) · low 0 (2003)2026
View FIG. 1 as a table
YearSupplements approved
20021
20051
20061
20074
20101
20112
20123
201310
20146
20153
20161
20172
20181
20193
20202
20211
20225
20232
20242
20254
20262
FIG. 1 — Supplements to P990081, 2002–2026Source: FDA PMA database

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MVC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 16, 2011Not classifiedZ-2717-2011PendingThe recall was initiated because Ventana Medical Systems, Inc. (Ventana), has identified the presence of precipitate in some DAB chromogen dispensers in two lots of ultraView Universal DAB Detection Kit (part number 760-500, lot A05885…
May 16, 2011Not classifiedZ-2716-2011PendingThe recall was initiated because Ventana Medical Systems, Inc. (Ventana), has identified the presence of precipitate in some DAB chromogen dispensers in two lots of ultraView Universal DAB Detection Kit (part number 760-500, lot A05885…
Sep 17, 2009Not classifiedZ-1146-2010Nonconforming Material/ComponentIncorrect result: Light staining inconsistencies with lot number 680957 of the Ventana PATHWAY HER-2/neu (4B5) Rabbit Monoclonal Primary Antibody.

No ancestor or descendant has a Class I recall.

How does its category compare?

All MVC devices

MVC System, Test, Her-2/Neu, Ihc: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MVC

4in 2025

2007: 42010: 62011: 62012: 52013: 132014: 72015: 72016: 32017: 52018: 22019: 72020: 62021: 12022: 32023: 22024: 32025: 42026 (partial year): 5
2007high 13 (2013) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MVC

Too few MVC 510(k)s in any year since 2007 to give a median.

Recalls per year · MVC

1in 2009

2009: 1
2007high 1 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MVC

2in 2021

2009: 42012: 22013: 22014: 12015: 42016: 32018: 22021: 2
2007high 4 (2009) · low 0 (2007)2025

MAUDE reports naming product code MVC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074—00
20080—00
20090—14
20106—00
20116—00
20125—02
201313—02
20147—01
20157—04
20163—03
20175—00
20182—02
20197—00
20206—00
20211—02
20223—00
20232—00
20243—00
20254—00
20265—0—
Cite this page

510(k) Report. “P990081: Pathway Anti-HCR-2/Ncu (4B5) Rabbit Monoclonal Primary Antibody.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/P990081

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026