De Novo · Granted
GrantedAPAS Compact with Urine Analysis Module
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Clever Culture Systems AG · PPU Microbial Colony Image Assessment System
Granted on Oct 6, 2016 as a new device type through De Novo after a 287-day review. 4 later devices descend from it.
- Decided
- Oct 6, 2016
- Review
- 287 days
- Class
- Class II
- Product code
- PPU
- Recalls
- None
4 later devices descend from it, 1 citing it directly.
- This deviceDEN150059 APAS Compact with Urine Analysis ModuleClever Culture Systems AG · 2016
- Descendant · cites it directly · 3 of its ownK183648 APAS Independence with Urine Analysis ModuleClever Culture Systems AG · 2019
- Descendant · 2 generations on · 2 of its ownK200839 APAS Independence with IC Chromogenic MRSA BD Analysis Module; APAS…Clever Culture Systems · 2021
- Descendant · 2 generations on · 0 of its ownK251511 PhenoMATRIXCopan Wasp Srl · 2026
- Descendant · 3 generations on · 0 of its ownK213280 BD Kiestra Methicillin-resistant Staphylococcus aureus (MRSA)…Bd Kiestra B.V. · 2023
Most influential descendants
Showing 4 of 4 · See all- K183648APAS Independence with Urine Analysis ModuleClever Culture Systems AG · 2019 · cites it directly3descendants
- K200839APAS Independence with IC Chromogenic MRSA BD Analysis Module; APAS Independence with IC…Clever Culture Systems · 2021 · 2 generations on2descendants
- K251511PhenoMATRIXCopan Wasp Srl · 2026 · 2 generations on0descendants
- K213280BD Kiestra Methicillin-resistant Staphylococcus aureus (MRSA) Application, BD Kiestra…Bd Kiestra B.V. · 2023 · 3 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | DEN150059 | APAS Compact with Urine Analysis Module | Clever Culture Systems AG | 2016 | — |
| +1 | K183648 | APAS Independence with Urine Analysis Module | Clever Culture Systems AG | 2019 | — |
| +2 | K200839 | APAS Independence with IC Chromogenic MRSA BD Analysis Module; APAS… | Clever Culture Systems | 2021 | — |
| +2 | K251511 | PhenoMATRIX | Copan Wasp Srl | 2026 | — |
| +3 | K213280 | BD Kiestra Methicillin-resistant Staphylococcus aureus (MRSA) Application, BD… | Bd Kiestra B.V. | 2023 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2019 | 1 | 0 |
| 2021 | 1 | 0 |
| 2023 | 1 | 0 |
| 2026 | 1 | 0 |
Sources: FDA De Novo database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in PPU? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new PPU device
- #1 K251511 PhenoMATRIXscore 11.3 · 3 generations · no Class I recall
Does it have a predicate?
De Novo decisions do not claim a predicate device.
What descends from it? (4)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K183648 | APAS Independence with Urine Analysis Module | 2019 | 3 |
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K183648 (Clever Culture Systems AG, 2019) has gone furthest, with 3 descendants of its own.
How does its category compare?
All PPU devicesPPU Microbial Colony Image Assessment System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.
FDA decisions per year · PPU
1in 2019
2026 to Oct 25, outlined.
Median 510(k) review, days · PPU
Too few PPU 510(k)s in any year since 2007 to give a median.
Recalls per year · PPU
No recall names PPU since 2007.
Adverse event reports per year · PPU
No MAUDE report names PPU since 2007.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 0 | — | 0 | 0 |
| 2012 | 0 | — | 0 | 0 |
| 2013 | 0 | — | 0 | 0 |
| 2014 | 0 | — | 0 | 0 |
| 2015 | 0 | — | 0 | 0 |
| 2016 | 1 | — | 0 | 0 |
| 2017 | 0 | — | 0 | 0 |
| 2018 | 0 | — | 0 | 0 |
| 2019 | 1 | — | 0 | 0 |
| 2020 | 0 | — | 0 | 0 |
| 2021 | 0 | — | 0 | 0 |
| 2022 | 0 | — | 0 | 0 |
| 2023 | 0 | — | 0 | 0 |
| 2024 | 0 | — | 0 | 0 |
| 2025 | 0 | — | 0 | 0 |
| 2026 | 1 | — | 0 | — |
What is it indicated for?
The APAS Compact is an in vitro diagnostic system comprised of an instrument for automated imaging of agar culture plates and a software analysis module for the following use: The APAS Compact, when using its urine analysis module, automates urine culture plate imaging and interpretation to detect the presence or absence of microbial growth on sheep blood and MacConkey agar culture plates that are inoculated with a 1µL sample volume. The APAS Compact…
Read the full statement (127 words)
The APAS Compact is an in vitro diagnostic system comprised of an instrument for automated imaging of agar culture plates and a software analysis module for the following use:
The APAS Compact, when using its urine analysis module, automates urine culture plate imaging and interpretation to detect the presence or absence of microbial growth on sheep blood and MacConkey agar culture plates that are inoculated with a 1µL sample volume. The APAS Compact, when using its urine analysis module, provides a semi-quantitative assessment of colony counts that are used as an aid in the diagnosis of urinary tract infection. All urine culture plates that are identified as positive for growth by the APAS Compact, when using its urine analysis module, must be reviewed by a trained microbiologist.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/DEN150059 · printed Nov 6, 2026
Cite this page
510(k) Report. “DEN150059: APAS Compact with Urine Analysis Module.” Lineage record from FDA De Novo data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/DEN150059
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026