510(k) Report

De Novo · Granted

Granted

3M™ Attest™ eBowie-Dick Test Card 10135 3M™ Attest™ eBowie-Dick Card Holder 10135CH 3M™ Attest™ eBowie-Dick Auto-reader 1190

DEN230068De NovoGranted May 21, 2024235-day reviewClass II21 CFR 880.2800

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

3M Company · JOJ Indicator, Physical/Chemical Sterilization Process

Granted on May 21, 2024 as a new device type through De Novo after a 235-day review.

Decided
May 21, 2024
Review
235 days
Class
Class II
Product code
JOJ
Recalls
None

De Novo grants create a new device type without a predicate, so this family tree starts here.

FIG. 1 — Lineage of DEN230068Source: FDA De Novo database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA De Novo database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JOJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JOJ device

  1. #1 K261992 3M™ Attest™ Steam Chemical Integrator (1243A…score 13.3 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

Does it have a predicate?

De Novo decisions do not claim a predicate device.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

How does its category compare?

All JOJ devices

JOJ Indicator, Physical/Chemical Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · JOJ

6in 2025

2007: 82008: 92009: 72010: 62011: 132012: 72013: 62014: 62015: 82016: 42017: 72018: 52019: 122020: 72021: 42022: 82023: 42024: 122025: 62026 (partial year): 3
2007high 13 (2011) · low 4 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JOJ

90in 2025

2007: 140 days median2008: 57 days median2009: 188 days median2010: 115 days median2011: 151 days median2012: 224 days median2013: 212 days median2014: 198 days median2015: 188 days median2016: 150 days median2017: 172 days median2018: 150 days median2019: 38 days median2020: 108 days median2021: 87 days median2022: 90 days median2023: 53 days median2024: 71 days median2025: 90 days median2026 (partial year): 30 days median
2007high 224 (2012) · low 38 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JOJ

1in 2023

2008: 22009: 22011: 22021: 12023: 1
2007high 2 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JOJ

12in 2025

2007: 12008: 12009: 92010: 32011: 12012: 82013: 62014: 112015: 72016: 112017: 352018: 122019: 82020: 62021: 292022: 232023: 392024: 252025: 12
2007high 39 (2023) · low 1 (2007)2025

MAUDE reports naming product code JOJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007814001
200895721
2009718829
2010611503
20111315121
2012722408
2013621206
20146198011
2015818807
20164150011
20177172035
20185150012
2019123808
2020710806
2021487129
2022890023
2023453139
20241271025
2025690012
20263300—

What is it indicated for?

3M™ Attest eBowie Dick Test Card and 3M™ Attest eBowie-Dick Card Holder As a Sterilizer Monitoring Device, the 3M™ Attest™ eBowie-Dick Test Card 10135 and 3M™ Attest™ eBowie-Dick Card Holder 10135CH is used in conjunction with the 3M™ Attest™ eBowie-Dick Auto-reader 1190 to monitor the equipment performance of steam sterilizers each day, as well as after a steam sterilizer repair. The 3M™ Attest™ eBowie-Dick Test Card 10135 will detect steam penetration…

Read the full statement (153 words)

3M™ Attest eBowie Dick Test Card and 3M™ Attest eBowie-Dick Card Holder

As a Sterilizer Monitoring Device, the 3M™ Attest™ eBowie-Dick Test Card 10135 and 3M™ Attest™ eBowie-Dick Card Holder 10135CH is used in conjunction with the 3M™ Attest™ eBowie-Dick Auto-reader 1190 to monitor the equipment performance of steam sterilizers each day, as well as after a steam sterilizer repair. The 3M™ Attest™ eBowie-Dick Test Card 10135 will detect steam penetration as a means of measuring air removal efficiency. The 3M™ Attest™ eBowie-Dick Test Card 10135 and 3M™ Attest™ eBowie-Dick Card Holder 10135CH is designed for testing air removal efficiency of 132°C (270°F), 134°C (273°F), and 135°C (275°F) for 3.5 minute test cycles in dynamic-air-removal steam sterilizers with chamber volumes greater than 54L (approximately 2 cubic feet).

3M™ Attest eBowie-Dick Auto-reader

The 3M™ Attest™ eBowie-Dick Auto-reader is designed to automatically interpret, store, and display the result of the 3M™ Attest™ eBowie-Dick Test Card.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “DEN230068: 3M™ Attest™ eBowie-Dick Test Card 10135 3M™ Attest™ eBowie-Dick Card Holder 10135CH 3M™ Attest™ eBowie-Dick Auto-reader 1190.” Lineage record from FDA De Novo data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/DEN230068

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026