510(k) Report

510(k) · Cleared

Recalled · Class II · 2012

Pinnacle Acetabular System

K000306510(k) SpecialCleared Mar 23, 200051-day reviewClass II21 CFR 888.3358Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

DePuy Orthopaedics, Inc. · LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Cleared on Mar 23, 2000 as substantially equivalent to K983014 (Johnson & Johnson Professionals, Inc., 1998) after a 51-day review. Its family tree reaches back 2 generations to 1993. 403 later devices descend from it. It has been recalled once.

Decided
Mar 23, 2000
Review
51 days
Class
Class II
Product code
LPH
Recalls
1
Predicates
1

Its longest line of descent runs back 2 generations to K930712 (1993). 6 ancestors in all. 403 later devices descend from it, 18 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor1995: 2 other ancestors1994: 1 other ancestor2016: 19 descendants cleared2017: 23 descendants cleared2006: 7 descendants cleared2003: 2 descendants cleared2013: 26 descendants cleared2000: 1 descendant cleared2018: 23 descendants cleared2011: 17 descendants cleared2012: 30 descendants cleared2008: 10 descendants cleared2019: 25 descendants cleared2022: 26 descendants cleared2014: 16 descendants cleared2015: 14 descendants cleared2024: 18 descendants cleared2004: 2 descendants cleared2007: 2 descendants cleared2009: 8 descendants cleared2010: 15 descendants cleared2005: 1 descendant cleared2023: 26 descendants cleared2026: 19 descendants cleared2020: 21 descendants cleared2025: 29 descendants cleared2021: 23 descendants clearedK930712 · P.F.C. Total Hip System Porous Coated Modular · Johnson & Johnson Orthopaedics, Inc. · 1993K983014 · Summit Acetabular System · Johnson & Johnson Professionals, Inc. · 19981990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 4 other ancestors 1 generation forward: 18 descendants2 generations forward: 48 descendants3 generations forward: 90 descendants4 generations forward: 99 descendants5 generations forward: 52 descendants6 generations forward: 31 descendants7 generations forward: 24 descendants8 generations forward: 28 descendants9 generations forward: 7 descendants10 generations forward: 3 descendants11 generations forward: 3 descendants K983014 · Summit Acetabular System · Johnson & Johnson Professionals, Inc. · 1998K930712 · P.F.C. Total Hip System Porous Coated Modular · Johnson & Johnson Orthopaedics, Inc. · 19930+5+10
  1. Root · 2 generations backK930712 P.F.C. Total Hip System Porous Coated ModularJohnson & Johnson Orthopaedics, Inc. · 1993
  2. 1 generation backK983014 Summit Acetabular SystemJohnson & Johnson Professionals, Inc. · 1998
  3. This deviceK000306 Pinnacle Acetabular SystemDePuy Orthopaedics, Inc. · 2000
  4. Descendant · cites it directly · 267 of its ownK001534 Pinnacle Acetabular SystemDePuy Orthopaedics, Inc. · 2000
  5. Descendant · cites it directly · 217 of its ownK031110 Apex Modular Acetabular CupApex Surgical, LLC · 2003
  6. Descendant · cites it directly · 207 of its ownK031495 Pinnacle Duofix HA Acetabular Cup ProsthesisDePuy Orthopaedics, Inc. · 2003
  7. Descendant · 2 generations on · 198 of its ownK042344 Excia Total Hip SystemAesculap, Inc. · 2005 · Recalled
  8. Descendant · 2 generations on · 118 of its ownK040544 Duraloc Option Acetabular Cup System, Models 1599-01-044 to -066…Depuy, Inc. · 2004

Showing 5 of 403 descendants. See all descendants

Descendants cleared per year

2000: 12003: 22004: 22005: 12006: 72007: 22008: 102009: 82010: 152011: 172012: 302013: 262014: 162015: 142016: 192017: 232018: 232019: 252020: 212021: 232022: 262023: 262024: 182025: 292026: 19
2000high 30 (2012) · low 0 (2001)2026

Most influential descendants

Showing 5 of 403 · See all
  • K001534Pinnacle Acetabular SystemDePuy Orthopaedics, Inc. · 2000 · cites it directly267descendants
  • K031110Apex Modular Acetabular CupApex Surgical, LLC · 2003 · cites it directly217descendants
  • K031495Pinnacle Duofix HA Acetabular Cup ProsthesisDePuy Orthopaedics, Inc. · 2003 · cites it directly207descendants
  • K042344Excia Total Hip SystemAesculap, Inc. · 2005 · 2 generations on198descendants
  • K040544Duraloc Option Acetabular Cup System, Models 1599-01-044 to -066, 1599-11-044 to -062…Depuy, Inc. · 2004 · 2 generations on118descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K930712P.F.C. Total Hip System Porous Coated ModularJohnson & Johnson Orthopaedics, Inc.1993—
−1K983014Summit Acetabular SystemJohnson & Johnson Professionals, Inc.1998—
0 (this device)K000306Pinnacle Acetabular SystemDePuy Orthopaedics, Inc.2000—
+1K001534Pinnacle Acetabular SystemDePuy Orthopaedics, Inc.2000—
+1K031110Apex Modular Acetabular CupApex Surgical, LLC2003—
+1K031495Pinnacle Duofix HA Acetabular Cup ProsthesisDePuy Orthopaedics, Inc.2003—
+2K042344Excia Total Hip SystemAesculap, Inc.200510
+2K040544Duraloc Option Acetabular Cup System, Models 1599-01-044 to -066, 1599-11-044…Depuy, Inc.2004—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199520
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200010
200320
200420
200510
200670
200720
2008100
200980
2010150
2011170
2012300
2013260
2014160
2015140
2016190
2017230
2018230
2019250
2020210
2021230
2022260
2023260
2024180
2025290
2026190
FIG. 1 — Lineage of K000306, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LPH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LPH device

  1. #1 K233498 Stryker Orthopaedics Hip Systems Labeling Updatescore 24.2 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K983014Summit Acetabular System
Summary page 1
…The PinnacleAcetabularSystem has thefollowingsimilarities totheacetabularcup liners thatwere clearedinK983014: same intendeduse; same material;same method of manufacture;same design;same sterilization…
Johnson & Johnson Professionals, Inc.1998—

Also cleared against K983014: 1 other device →

What descends from it? (403)

See all 403

18 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K001534Pinnacle Acetabular SystemDePuy Orthopaedics, Inc.2000267
K031110Apex Modular Acetabular CupApex Surgical, LLC2003217
K031495Pinnacle Duofix HA Acetabular Cup ProsthesisDePuy Orthopaedics, Inc.2003207
K062148Depuy Pinnacle Altrx Acetabular Cup LinerDePuy Orthopaedics, Inc.2006106
K121297Origin Hip Stem, Femoral Head, Logical PX-Series Acetabular Shell, Locigal G-Series…Signature Orthopaedics Pty, Ltd.201220
K120924Bencox Mirabo Cup Bencox Mirabo Insert (Standard& Elevated Tyes) Bencox Mirabo Cup Screw…Corentec Co., Ltd.20128
K192919Pinnacle Duofix HA Acetabular Cup ProsthesisDePuy Orthopaedics, Inc.20196
K162127Bencox M Stem Lateralized & Bencox Mirabo Cup SystemCorentec Co., Ltd.20166
K172857Latitud Hip Replacement SystemMeril Healthcare Pvt., Ltd.20185
K172806Bencox Mirabo Cup SystemCorentec Co., Ltd.20174

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 28, 2012Class IIZ-0569-2013Labeling Change ControlDePuy Orthopaedics, Inc. is issuing a voluntary recall of ten lots of the Pinnacle Cancellous Screw due to an incorrect description on the label. The label description reads Pinnacle Cannulated Screw and should read Pinnacle Cancellous…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 7 years and passes through 3 companies; a new generation was cleared every 5.1 years on median.
  • None of the 6 ancestors has a recall.
  • Of its descendants, K001534 (DePuy Orthopaedics, Inc., 2000) has gone furthest, with 267 descendants of its own.
  • 1 other device also cites its predicate K983014.

How does its category compare?

All LPH devices

LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LPH

12in 2025

2007: 102008: 102009: 32010: 152011: 162012: 252013: 112014: 202015: 182016: 102017: 122018: 92019: 222020: 72021: 82022: 142023: 102024: 62025: 122026 (partial year): 8
2007high 25 (2012) · low 3 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LPH

153in 2025

2007: 116 days median2008: 86 days median2009: 112 days median2010: 100 days median2011: 110 days median2012: 76 days median2013: 135 days median2014: 129 days median2015: 111 days median2016: 150 days median2017: 121 days median2018: 145 days median2019: 164 days median2020: 164 days median2021: 410 days median2022: 247 days median2023: 111 days median2024: 92 days median2025: 153 days median2026 (partial year): 256 days median
2007high 410 (2021) · low 76 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LPH

4in 2025

2007: 162008: 12009: 22010: 222011: 262012: 302013: 92014: 32015: 12016: 82017: 142018: 22019: 82020: 62021: 42022: 22023: 22024: 12025: 42026 (partial year): 2
2007high 30 (2012) · low 1 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LPH

6,163in 2025

2007: 5092008: 7062009: 1,1562010: 2,0072011: 3,3202012: 3,7852013: 5,0022014: 5,1632015: 5,8302016: 4,9002017: 5,3122018: 5,9542019: 7,1302020: 5,4292021: 5,9782022: 5,2232023: 5,3452024: 5,6262025: 6,163
2007high 7,130 (2019) · low 509 (2007)2025

MAUDE reports naming product code LPH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071011616509
200810861706
2009311221,156
201015100222,007
201116110263,320
20122576303,785
20131113595,002
20142012935,163
20151811115,830
20161015084,900
201712121145,312
2018914525,954
20192216487,130
2020716465,429
2021841045,978
20221424725,223
20231011125,345
202469215,626
20251215346,163
202682562—

What is it indicated for?

The Pinnacle Acetabular System is indicated for use in total hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, and nonunion of femoral fractures. Use of the prosthesis is also indicated for revision of the previous hip arthroplasty and for patients with congenital hip dysplasia…

Read the full statement (95 words)

The Pinnacle Acetabular System is indicated for use in total hip replacement procedures for

patients suffering severe pain and disability due to structural damage in the hip joint from

rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular

necrosis, and nonunion of femoral fractures. Use of the prosthesis is also indicated for revision

of the previous hip arthroplasty and for patients with congenital hip dysplasia, protrusio

acetabuli, slipped capital femoral epiphysis and disability due to previous fusion, where bone

stock is inadequate for other reconstruction techniques.

All Pinnacle porous-coated acetabular shells are indicated for cementless application.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K000306: Pinnacle Acetabular System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K000306

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026