510(k) Report

510(k) · Cleared

Cleared

Pinnacle Duofix HA Acetabular Cup Prosthesis

K192919510(k) SpecialCleared Nov 26, 201942-day reviewClass II21 CFR 888.3358

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

DePuy Orthopaedics, Inc. · LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Cleared on Nov 26, 2019 as substantially equivalent to K031495 (DePuy Orthopaedics, Inc., 2003) and 1 other predicate after a 42-day review. Its family tree reaches back 5 generations to 1994. 6 later devices descend from it.

Decided
Nov 26, 2019
Review
42 days
Class
Class II
Product code
LPH
Recalls
None
Predicates
2

Its longest line of descent runs back 5 generations to K941847 (1994). 19 ancestors in all, the oldest K840643 (1984). 6 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor1998: 2 other ancestors1997: 1 other ancestor1996: 2 other ancestors1995: 3 other ancestors1994: 2 other ancestors1993: 1 other ancestor1984: 1 other ancestor2024: 2 descendants cleared2025: 3 descendants cleared2026: 1 descendant clearedK941847 · 5/8 Porocoat Aml Hip Porocoat Porous Coating · Depuy, Inc. · 1994K964650 · TCP-Coated Porocoat Aml Femoral Hip Prosthesis · Depuy, Inc. · 1997K980711 · Harris/Galante Porous (HGP) Hip Prosthesis with Calcicoat Ceramic Coating · Zimmer, Inc. · 1998 · recalledK001277 · Triflange Acetabular Cup System · DePuy Orthopaedics, Inc. · 2000K031495 · Pinnacle Duofix HA Acetabular Cup Prosthesis · DePuy Orthopaedics, Inc. · 200319801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 9 other ancestors4 generations back: 2 other ancestors 1 generation forward: 1 descendant2 generations forward: 3 descendants3 generations forward: 2 descendants K031495 · Pinnacle Duofix HA Acetabular Cup Prosthesis · DePuy Orthopaedics, Inc. · 2003K001277 · Triflange Acetabular Cup System · DePuy Orthopaedics, Inc. · 2000K980711 · Harris/Galante Porous (HGP) Hip Prosthesis with Calcicoat Ceramic Coating · Zimmer, Inc. · 1998 · recalledK964650 · TCP-Coated Porocoat Aml Femoral Hip Prosthesis · Depuy, Inc. · 1997K941847 · 5/8 Porocoat Aml Hip Porocoat Porous Coating · Depuy, Inc. · 1994−50
  1. Root · 5 generations backK941847 5/8 Porocoat Aml Hip Porocoat Porous CoatingDepuy, Inc. · 1994
  2. 4 generations backK964650 TCP-Coated Porocoat Aml Femoral Hip ProsthesisDepuy, Inc. · 1997
  3. 3 generations backK980711 Harris/Galante Porous (HGP) Hip Prosthesis with Calcicoat Ceramic…Zimmer, Inc. · 1998 · Recalled
  4. 2 generations backK001277 Triflange Acetabular Cup SystemDePuy Orthopaedics, Inc. · 2000
  5. 1 generation backK031495 Pinnacle Duofix HA Acetabular Cup ProsthesisDePuy Orthopaedics, Inc. · 2003
  6. This deviceK192919 Pinnacle Duofix HA Acetabular Cup ProsthesisDePuy Orthopaedics, Inc. · 2019
  7. Descendant · cites it directly · 5 of its ownK231873 DePuy SUMMIT Porocoat Hip Prosthesis – MR Conditional, DePuy SUMMIT…Depuy Ireland UC · 2024
  8. Descendant · 2 generations on · 1 of its ownK251292 RECLAIM Monobloc Revision Femoral StemDepuy Ireland UC · 2025
  9. Descendant · 2 generations on · 1 of its ownK243977 EMPHASYS Acetabular SystemDepuy Ireland UC · 2025
  10. Descendant · 3 generations on · 0 of its ownK260173 M-Vizion Femoral Revision System ExtensionMedacta International S.A. · 2026
  11. Descendant · 3 generations on · 0 of its ownK252067 Velora Acetabular SystemRestor3D, Inc. · 2025

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2024: 22025: 32026: 1
2024high 3 (2025) · low 1 (2026)2026

Most influential descendants

Showing 5 of 6 · See all
  • K231873DePuy SUMMIT Porocoat Hip Prosthesis – MR Conditional, DePuy SUMMIT DuoFix Hip…Depuy Ireland UC · 2024 · cites it directly5descendants
  • K251292RECLAIM Monobloc Revision Femoral StemDepuy Ireland UC · 2025 · 2 generations on1descendant
  • K243977EMPHASYS Acetabular SystemDepuy Ireland UC · 2025 · 2 generations on1descendant
  • K260173M-Vizion Femoral Revision System ExtensionMedacta International S.A. · 2026 · 3 generations on0descendants
  • K252067Velora Acetabular SystemRestor3D, Inc. · 2025 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K9418475/8 Porocoat Aml Hip Porocoat Porous CoatingDepuy, Inc.1994—
−4K964650TCP-Coated Porocoat Aml Femoral Hip ProsthesisDepuy, Inc.1997—
−3K980711Harris/Galante Porous (HGP) Hip Prosthesis with Calcicoat Ceramic CoatingZimmer, Inc.19981
−2K001277Triflange Acetabular Cup SystemDePuy Orthopaedics, Inc.2000—
−1K031495Pinnacle Duofix HA Acetabular Cup ProsthesisDePuy Orthopaedics, Inc.2003—
0 (this device)K192919Pinnacle Duofix HA Acetabular Cup ProsthesisDePuy Orthopaedics, Inc.2019—
+1K231873DePuy SUMMIT Porocoat Hip Prosthesis – MR Conditional, DePuy SUMMIT DuoFix Hip…Depuy Ireland UC2024—
+2K251292RECLAIM Monobloc Revision Femoral StemDepuy Ireland UC2025—
+2K243977EMPHASYS Acetabular SystemDepuy Ireland UC2025—
+3K260173M-Vizion Femoral Revision System ExtensionMedacta International S.A.2026—
+3K252067Velora Acetabular SystemRestor3D, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198410
199310
199420
199530
199620
199710
199820
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202420
202530
202610
FIG. 1 — Lineage of K192919, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LPH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LPH device

  1. #1 K233498 Stryker Orthopaedics Hip Systems Labeling Updatescore 24.2 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031495Pinnacle Duofix HA Acetabular Cup Prosthesis
Summary page 4
…ces Regulation 888.3358 Product Code(s) LPH Legally marketed device(s) to Pinnacle Duofix™ HA Acetabular Cup Prosthesis – K000306, K031495 which equivalence is claimed Reason for 510(k) submission The purpose of this submission is to support the manufacturing of th…
DePuy Orthopaedics, Inc.2003—
*K000306Pinnacle Acetabular System
Summary page 4
…tion Devices Regulation 888.3358 Product Code(s) LPH Legally marketed device(s) to Pinnacle Duofix™ HA Acetabular Cup Prosthesis – K000306, K031495 which equivalence is claimed Reason for 510(k) submission The purpose of this submission is to support the manufactur…
DePuy Orthopaedics, Inc.20001

Also cleared against K031495: 4 other devices →

What descends from it? (6)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K231873DePuy SUMMIT Porocoat Hip Prosthesis – MR Conditional, DePuy SUMMIT DuoFix Hip…Depuy Ireland UC20245

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 3 companies; a new generation was cleared every 2.9 years on median.
  • 4 of the 18 ancestors have a recall of any class.
  • Of its descendants, K231873 (Depuy Ireland UC, 2024) has gone furthest, with 5 descendants of its own.
  • 4 other devices also cite its predicate K031495.

How does its category compare?

All LPH devices

LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · LPH

12in 2025

2007: 102008: 102009: 32010: 152011: 162012: 252013: 112014: 202015: 182016: 102017: 122018: 92019: 222020: 72021: 82022: 142023: 102024: 62025: 122026 (partial year): 8
2007high 25 (2012) · low 3 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LPH

153in 2025

2007: 116 days median2008: 86 days median2009: 112 days median2010: 100 days median2011: 110 days median2012: 76 days median2013: 135 days median2014: 129 days median2015: 111 days median2016: 150 days median2017: 121 days median2018: 145 days median2019: 164 days median2020: 164 days median2021: 410 days median2022: 247 days median2023: 111 days median2024: 92 days median2025: 153 days median2026 (partial year): 256 days median
2007high 410 (2021) · low 76 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LPH

4in 2025

2007: 162008: 12009: 22010: 222011: 262012: 302013: 92014: 32015: 12016: 82017: 142018: 22019: 82020: 62021: 42022: 22023: 22024: 12025: 42026 (partial year): 2
2007high 30 (2012) · low 1 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LPH

6,163in 2025

2007: 5092008: 7062009: 1,1562010: 2,0072011: 3,3202012: 3,7852013: 5,0022014: 5,1632015: 5,8302016: 4,9002017: 5,3122018: 5,9542019: 7,1302020: 5,4292021: 5,9782022: 5,2232023: 5,3452024: 5,6262025: 6,163
2007high 7,130 (2019) · low 509 (2007)2025

MAUDE reports naming product code LPH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071011616509
200810861706
2009311221,156
201015100222,007
201116110263,320
20122576303,785
20131113595,002
20142012935,163
20151811115,830
20161015084,900
201712121145,312
2018914525,954
20192216487,130
2020716465,429
2021841045,978
20221424725,223
20231011125,345
202469215,626
20251215346,163
202682562—

What is it indicated for?

The Pinnacle Duofix™ HA Acetabular Cup Prosthesis is indicated for use in total hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, and nonunion of femoral fractures. Use of the prosthesis is also indicated for revision of the previous hip arthroplasty and for patients with…

Read the full statement (100 words)

The Pinnacle Duofix™ HA Acetabular Cup Prosthesis is indicated for use in total hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, and nonunion of femoral fractures. Use of the prosthesis is also indicated for revision of the previous hip arthroplasty and for patients with congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis and disability due to previous fusion, where bone stock is inadequate for other reconstruction techniques.

The Pinnacle Duofix HA Acetabular Cup Prosthesis is indicated for cementless application.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K192919: Pinnacle Duofix HA Acetabular Cup Prosthesis.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K192919

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026