Protector 21, Protector 50, Protector 14, Injector Luer Lock, Infusion Adapter

K001368510(k) SpecialCleared May 12, 200011-day reviewClass II21 CFR 880.5440Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Carmel Pharma Ab. · LHI Set, I.V. Fluid Transfer

Cleared on May 12, 2000 after an 11-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 2 of the 45 devices in its family, counting itself, have a Class I recall. 44 later devices descend from it. It has been recalled 4 times.

Decided
May 12, 2000
Review
11 days
Class
Class II
Product code
LHI
Recalls
4
Predicates
0

44 later devices descend from it, 5 citing it directly. 2 of the 45 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2002: 1 descendant cleared2009: 2 descendants cleared2012: 1 descendant cleared2007: 1 descendant cleared2011: 1 descendant cleared2008: 1 descendant cleared, 1 with a Class I recall12013: 3 descendants cleared, 1 with a Class I recall12018: 5 descendants cleared2017: 3 descendants cleared2024: 8 descendants cleared2020: 3 descendants cleared2014: 2 descendants cleared2015: 2 descendants cleared2021: 2 descendants cleared2025: 6 descendants cleared2023: 2 descendants cleared2022: 1 descendant cleared200020052010201520202025
  1. This deviceK001368 Protector 21, Protector 50, Protector 14, Injector Luer Lock…Carmel Pharma Ab. · 2000
  2. Descendant · cites it directly · 38 of its ownK023747 PhasealCarmel Pharma Ab. · 2002
  3. Descendant · 2 generations on · 37 of its ownK110023 Infusion Adapter C100Carmel Pharma Ab. · 2011
  4. Descendant · cites it directly · 36 of its ownK092782 Injector Luer, Model N34, Injector Luer Lock, Model N35, Injector…Carmel Pharma Ab. · 2009
  5. Descendant · cites it directly · 36 of its ownK090634 Phaseal Protector P14, P21, P28 and P50Carmel Pharma Ab. · 2009
  6. Descendant · cites it directly · 35 of its ownK120384 Protector, Injector, ConnectorBd Medical-Medical Surgical Systems · 2012

Showing 5 of 44 descendants. See all descendants

Descendants cleared per year

2002: 12007: 12008: 12009: 22011: 12012: 12013: 32014: 22015: 22017: 32018: 52020: 32021: 22022: 12023: 22024: 82025: 6
2002high 8 (2024) · low 0 (2003)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 44 · See all
  • K023747PhasealCarmel Pharma Ab. · 2002 · cites it directly38descendants
  • K110023Infusion Adapter C100Carmel Pharma Ab. · 2011 · 2 generations on37descendants
  • K092782Injector Luer, Model N34, Injector Luer Lock, Model N35, Injector Luer Lock, N35C…Carmel Pharma Ab. · 2009 · cites it directly36descendants
  • K090634Phaseal Protector P14, P21, P28 and P50Carmel Pharma Ab. · 2009 · cites it directly36descendants
  • K120384Protector, Injector, ConnectorBd Medical-Medical Surgical Systems · 2012 · cites it directly35descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K001368Protector 21, Protector 50, Protector 14, Injector Luer Lock, Infusion AdapterCarmel Pharma Ab.2000—
+1K023747PhasealCarmel Pharma Ab.2002—
+2K110023Infusion Adapter C100Carmel Pharma Ab.2011—
+1K092782Injector Luer, Model N34, Injector Luer Lock, Model N35, Injector Luer Lock…Carmel Pharma Ab.2009—
+1K090634Phaseal Protector P14, P21, P28 and P50Carmel Pharma Ab.2009—
+1K120384Protector, Injector, ConnectorBd Medical-Medical Surgical Systems2012—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200210
200710
200811
200920
201110
201210
201331
201420
201520
201730
201850
202030
202120
202210
202320
202480
202560
FIG. 1 — Lineage of K001368, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LHI device

  1. #1 K252987 ADRx™ Admixture Devicescore 12.4 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (44)

See all 44

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K023747PhasealCarmel Pharma Ab.200238
K092782Injector Luer, Model N34, Injector Luer Lock, Model N35, Injector Luer Lock, N35C…Carmel Pharma Ab.200936
K090634Phaseal Protector P14, P21, P28 and P50Carmel Pharma Ab.200936
K120384Protector, Injector, ConnectorBd Medical-Medical Surgical Systems201235
K070633Genie, Model CH-200Icu Medical, Inc.200713

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 26, 2007Not classifiedZ-1181-2007OtherRisk of Leakage: When the PhaSeal Protector and Injector are used with drug vials with thick rubber stoppers which have a concave upper surface and a flat bottom surface there is a potential for leakage. These rubber stoppers are made of…
Jun 26, 2007Not classifiedZ-1180-2007OtherRisk of Leakage: When the PhaSeal Protector and Injector are used with drug vials with thick rubber stoppers which have a concave upper surface and a flat bottom surface there is a potential for leakage. These rubber stoppers are made of…
Jun 26, 2007Not classifiedZ-1179-2007OtherRisk of Leakage: When the PhaSeal Protector and Injector are used with drug vials with thick rubber stoppers which have a concave upper surface and a flat bottom surface there is a potential for leakage. These rubber stoppers are made of…
Jun 26, 2007Not classifiedZ-1178-2007OtherRisk of Leakage: When the PhaSeal Protector and Injector are used with drug vials with thick rubber stoppers which have a concave upper surface and a flat bottom surface there is a potential for leakage. These rubber stoppers are made of…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K081361Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXXIcu Medical, Inc. · 2008 · 2 generations on · ◆ in the 2008 descendants bar of FIG. 12 Class I recalls
  • K131549Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · 3 generations on · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Of its descendants, K023747 (Carmel Pharma Ab., 2002) has gone furthest, with 38 descendants of its own.

How does its category compare?

All LHI devices

LHI Set, I.V. Fluid Transfer: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LHI

6in 2025

2007: 32008: 42009: 82010: 52011: 32012: 52013: 42014: 62015: 32016: 32017: 52018: 12019: 42020: 32021: 32022: 62023: 52024: 92025: 62026 (partial year): 2
2007high 9 (2024) · low 1 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LHI

196in 2025

2007: 89 days median2008: 154 days median2009: 95 days median2010: 139 days median2011: 99 days median2012: 127 days median2013: 102 days median2014: 63 days median2015: 99 days median2016: 159 days median2017: 56 days median2019: 247 days median2020: 189 days median2021: 396 days median2022: 278 days median2023: 90 days median2024: 166 days median2025: 196 days median
2007high 396 (2021) · low 56 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LHI

1in 2024

2007: 12008: 32009: 12010: 32011: 122013: 32016: 1, 1 Class I2017: 12018: 3, 3 Class I2019: 12022: 22023: 2, 2 Class I2024: 12026 (partial year): 1
2007high 12 (2011) · low 0 (2012)2026

Class I (6 since 2007) Class II/III not classified

Adverse event reports per year · LHI

1,334in 2025

2007: 182008: 272009: 372010: 492011: 1552012: 5292013: 1542014: 1552015: 1282016: 2192017: 5362018: 1,4992019: 1,6342020: 1,1712021: 9332022: 1,8342023: 1,0472024: 1,2702025: 1,334
2007high 1,834 (2022) · low 18 (2007)2025

MAUDE reports naming product code LHI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007389118
20084154327
2009895137
20105139349
201139912155
201251270529
201341023154
20146630155
20153990128
201631591 (1)219
20175561536
20181—3 (3)1,499
2019424711,634
2020318901,171
202133960933
2022627821,834
20235902 (2)1,047
2024916611,270
2025619601,334
20262—1—
Cite this page

510(k) Report. “K001368: Protector 21, Protector 50, Protector 14, Injector Luer Lock, Infusion Adapter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K001368

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026