510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Chemolock, Chemolock Port

K131549510(k) TraditionalCleared Jul 26, 201358-day reviewClass II21 CFR 880.5440Recalls 60 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Icu Medical, Inc. · ONB Closed Antineoplastic And Hazardous Drug Reconstitution And Transfer System

Cleared on Jul 26, 2013 as substantially equivalent to K123213 (Becton, Dickinson & CO, 2013) and 2 other predicates after a 58-day review. Its family tree reaches back 7 generations to 1988. 3 of the 42 devices in its family, counting itself, have a Class I recall. 10 later devices descend from it. It has been recalled 60 times, including 2 Class I recalls.

Decided
Jul 26, 2013
Review
58 days
Class
Class II
Product code
ONB
Recalls
60 (2 Class I)
Predicates
3

Its longest line of descent runs back 7 generations to K883614 (1988). 31 ancestors in all. 10 later devices descend from it, 1 citing it directly. 3 of the 42 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2009: 3 other ancestors2008: 2 other ancestors, 1 with a Class I recall12007: 3 other ancestors2003: 2 other ancestors2001: 1 other ancestor2000: 2 other ancestors1998: 2 other ancestors1997: 2 other ancestors, 1 with a Class I recall11994: 4 other ancestors2006: 1 other ancestor1995: 2 other ancestors2015: 1 descendant cleared2017: 1 descendant cleared2020: 1 descendant cleared2018: 1 descendant cleared2021: 2 descendants cleared2023: 1 descendant cleared2025: 2 descendants cleared2024: 1 descendant clearedK883614 · Cytoguard (R) · Survival Technology, Inc. · 1988K972527 · Phaseal, System for Sealed Handling of Chemotherapeutic Agents · Quality System Consulting · 1997K980381 · Phaseal, Closed System Forhandling of Parenteral Drugs, Additional Administration Devices. C80 Infusion Adapter, A10 Pro · Carmel Pharma Ab. · 1998K023747 · Phaseal · Carmel Pharma Ab. · 2002K110023 · Infusion Adapter C100 · Carmel Pharma Ab. · 2011×2: 2 generations cleared close together, K120384 (2012), K123213 (2013) 1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors, 1 with a Class I recall12 generations back: 11 other ancestors, 1 with a Class I recall13 generations back: 8 other ancestors4 generations back: 3 other ancestors 1 generation forward: 1 descendant2 generations forward: 4 descendants3 generations forward: 5 descendants K123213 · BD Phaseal Closed System Transfer Device · Becton, Dickinson & CO · 2013 · recalledK120384 · Protector, Injector, Connector · Bd Medical-Medical Surgical Systems · 2012K110023 · Infusion Adapter C100 · Carmel Pharma Ab. · 2011K023747 · Phaseal · Carmel Pharma Ab. · 2002K980381 · Phaseal, Closed System Forhandling of Parenteral Drugs, Additional Administration Devices. C80 Infusion Adapter, A10 Pro · Carmel Pharma Ab. · 1998K972527 · Phaseal, System for Sealed Handling of Chemotherapeutic Agents · Quality System Consulting · 1997K883614 · Cytoguard (R) · Survival Technology, Inc. · 1988−50
  1. Root · 7 generations backK883614 Cytoguard (R)Survival Technology, Inc. · 1988
  2. 6 generations backK972527 Phaseal, System for Sealed Handling of Chemotherapeutic AgentsQuality System Consulting · 1997
  3. 5 generations backK980381 Phaseal, Closed System Forhandling of Parenteral Drugs, Additional…Carmel Pharma Ab. · 1998
  4. 4 generations backK023747 PhasealCarmel Pharma Ab. · 2002
  5. 3 generations backK110023 Infusion Adapter C100Carmel Pharma Ab. · 2011
  6. 2 generations backK120384 Protector, Injector, ConnectorBd Medical-Medical Surgical Systems · 2012
  7. 1 generation backK123213 BD Phaseal Closed System Transfer DeviceBecton, Dickinson & CO · 2013 · Recalled
  8. This deviceK131549 Chemolock, Chemolock PortIcu Medical, Inc. · 2013
  9. Descendant · cites it directly · 9 of its ownK141448 Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2015
  10. Descendant · 2 generations on · 3 of its ownK192866 Chemfort CSTDSimplivia Healthcare, Ltd. · 2020 · Recalled
  11. Descendant · 2 generations on · 1 of its ownK201422 Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC · 2021
  12. Descendant · 2 generations on · 1 of its ownK180489 TEVADAPTOR Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2018
  13. Descendant · 3 generations on · 0 of its ownK253033 Chemfort® 28-day 20 mm Vial Adaptor and 13 mm ConvertorSimplivia Healthcare, Ltd. · 2025

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2015: 12017: 12018: 12020: 12021: 22023: 12024: 12025: 2
2015high 2 (2021) · low 0 (2016)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 10 · See all
  • K141448Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2015 · cites it directly9descendants
  • K192866Chemfort CSTDSimplivia Healthcare, Ltd. · 2020 · 2 generations on3descendants
  • K201422Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC · 2021 · 2 generations on1descendant
  • K180489TEVADAPTOR Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2018 · 2 generations on1descendant
  • K253033Chemfort® 28-day 20 mm Vial Adaptor and 13 mm ConvertorSimplivia Healthcare, Ltd. · 2025 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K883614Cytoguard (R)Survival Technology, Inc.1988—
−6K972527Phaseal, System for Sealed Handling of Chemotherapeutic AgentsQuality System Consulting1997—
−5K980381Phaseal, Closed System Forhandling of Parenteral Drugs, Additional…Carmel Pharma Ab.1998—
−4K023747PhasealCarmel Pharma Ab.2002—
−3K110023Infusion Adapter C100Carmel Pharma Ab.2011—
−2K120384Protector, Injector, ConnectorBd Medical-Medical Surgical Systems2012—
−1K123213BD Phaseal Closed System Transfer DeviceBecton, Dickinson & CO20132
0 (this device)K131549Chemolock, Chemolock PortIcu Medical, Inc.2013—
+1K141448Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant2015—
+2K192866Chemfort CSTDSimplivia Healthcare, Ltd.20201
+2K201422Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC2021—
+2K180489TEVADAPTOR Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant2018—
+3K253033Chemfort® 28-day 20 mm Vial Adaptor and 13 mm ConvertorSimplivia Healthcare, Ltd.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199440
199520
199721
199820
200020
200110
200320
200610
200730
200821
200930

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201710
201810
202010
202120
202310
202410
202520
FIG. 1 — Lineage of K131549, 1988–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new ONB device

  1. #1 K240171 ProSeal™ Injector Plus (Model no. 421050)score 18.2 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K123213BD Phaseal Closed System Transfer Device
Summary page 1
…l Name; Closed System Drug Transfer Device Classification Names: 21 CFR 880.5440. Closed System Drug Transfer Device, Product Code: B80ONB and 80FPA. : Predicate Device: K123213 BD Phaseal Closed System Drug Transfer Device K081361 ChemoClave Cytotoxic Medication Preparation and Delivery System . K082806…
Becton, Dickinson & CO20132
*K081361Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXX
Summary page 1
…on Names: 21 CFR 880.5440. Closed System Drug Transfer Device, Product Code: B80ONB and 80FPA. : Predicate Device: K123213 BD Phaseal Closed System Drug Transfer Device K081361 ChemoClave Cytotoxic Medication Preparation and Delivery System . K082806 Spinning Spiros Device Description; The ChemoLock is a…
Icu Medical, Inc.2008Class I
*K082806Spinning Spiros, Model CH-2000
Summary page 1
…e: B80ONB and 80FPA. : Predicate Device: K123213 BD Phaseal Closed System Drug Transfer Device K081361 ChemoClave Cytotoxic Medication Preparation and Delivery System . K082806 Spinning Spiros Device Description; The ChemoLock is a needlefree, single-use, Closed System Drug Transfer Device (CSTD), The Ch…
Icu Medical, Inc.20095

Also cleared against K123213: 8 other devices →

What descends from it? (10)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K141448Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant20159

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 12, 2026Class IIZ-3159-2026Process controlIntravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid…
May 12, 2026Class IIZ-3161-2026Process controlIntravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration…
Mar 24, 2026Class IIZ-2210-2026Process controlOncology sets/closed system transfer devices with extension sets may leak resulting in therapy delayed if noticed and corrected during setup, but if the leak occurs during therapy this may result in therapy interruption with the potential…
Sep 8, 2025Class IIZ-0340-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0352-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0333-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0336-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0343-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0358-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0348-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0347-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0341-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0335-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0357-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0351-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0342-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0356-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0344-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0350-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0354-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0360-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0359-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0346-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0338-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0361-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0362-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0329-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0337-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0330-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0349-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0328-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0353-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0339-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0365-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0334-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0355-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0345-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0331-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0364-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0363-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Sep 8, 2025Class IIZ-0332-2026Under Investigation by firmPort weld of drug transfer device may separate or break during use and potentially result in a leak.
Nov 6, 2024Class IIZ-0652-2025Packaging process controlDue to finished goods incorrectly packaged and labeled.
Jun 27, 2022Class IIZ-1681-2022Component design/selectionPorts have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports…
Mar 16, 2021Class IIZ-1574-2021Component design/selectionDue to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Mar 16, 2021Class IIZ-1572-2021Component design/selectionDue to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Mar 16, 2021Class IIZ-1575-2021Component design/selectionDue to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Mar 16, 2021Class IIZ-1571-2021Component design/selectionDue to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Mar 16, 2021Class IIZ-1570-2021Component design/selectionDue to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Mar 16, 2021Class IIZ-1573-2021Component design/selectionDue to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Aug 31, 2020Class IIZ-0703-2021Process controlPotential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K081361Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXXIcu Medical, Inc. · 2008 · Direct predicate · ◆ in the 2008 ancestors bar of FIG. 12 Class I recalls
  • K964435Primary IV SetIcu Medical, Inc. · 1997 · 2 generations back · ◆ in the 1997 ancestors bar of FIG. 15 Class I recalls

What does this lineage say?

  • The line of descent spans 25 years and passes through 6 companies; a new generation was cleared every 1.4 years on median.
  • 9 of the 31 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K141448 (Teva Medical Ltd., Migada Plant, 2015) has gone furthest, with 9 descendants of its own.
  • 8 other devices also cite its predicate K123213.

How does its category compare?

All ONB devices

ONB Closed Antineoplastic And Hazardous Drug Reconstitution And Transfer System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · ONB

6in 2025

2013: 32014: 22015: 32017: 22018: 62019: 12020: 32021: 32022: 12023: 32024: 82025: 6
2007high 8 (2024) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · ONB

60in 2025

2013: 58 days median2015: 235 days median2018: 249 days median2020: 213 days median2021: 277 days median2023: 175 days median2024: 102 days median2025: 60 days median
2007high 277 (2021) · low 58 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · ONB

38in 2025

2016: 12019: 1, 1 Class I2020: 42022: 12023: 12024: 12025: 382026 (partial year): 3
2007high 38 (2025) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · ONB

934in 2025

2012: 12013: 22014: 32015: 42016: 132017: 662018: 2282019: 4642020: 4422021: 5482022: 4482023: 4562024: 6862025: 934
2007high 934 (2025) · low 0 (2007)2025

MAUDE reports naming product code ONB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—01
201335802
20142—03
2015323504
20160—113
20172—066
201862490228
20191—1 (1)464
202032134442
202132770548
20221—1448
202331751456
202481021686
202566038934
20260—3—

What is it indicated for?

The ChemoLock Closed System Drug Transfer Device prevents the transfer of environmental contaminants, including bacterial and airborne contaminants into the system, and the escape of drug or vapor concentrations outside the system. The ChemoLock is needlefree and cannot be deactivated, which will passively aid in preventing needlestick injuries and the exposure to cytotoxic medications for healthcare personnel.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K131549: Chemolock, Chemolock Port.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K131549

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026