510(k) · Cleared
Recalled · 2011Varis 1.4G
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Varian Medical Systems, Inc. · IYE Accelerator, Linear, Medical
Cleared on Jun 23, 2000 after a 24-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 183 later devices descend from it. It has been recalled 8 times.
- Decided
- Jun 23, 2000
- Review
- 24 days
- Class
- Class II
- Product code
- IYE
- Recalls
- 8
- Predicates
- 0
183 later devices descend from it, 2 citing it directly.
- This deviceK001643 Varis 1.4GVarian Medical Systems, Inc. · 2000
- Descendant · cites it directly · 121 of its ownK042956 VisionVarian Medical Systems, Inc. · 2005
- Descendant · 2 generations on · 99 of its ownK112930 Mobile MimMim Software, Inc. · 2011
- Descendant · cites it directly · 72 of its ownK050479 4D Integrated Treatment ConsoleVarian Medical Systems · 2005 · Recalled
- Descendant · 2 generations on · 69 of its ownK081036 4D Integrated Treatment Console with Patient Accessory Verification…Varian Medical Systems, Inc. · 2008 · Recalled
- Descendant · 3 generations on · 68 of its ownK151913 MIM-Thin Client (mobile)Mim Software, Inc. · 2016
Showing 5 of 183 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 183 · See all- K042956VisionVarian Medical Systems, Inc. · 2005 · cites it directly121descendants
- K112930Mobile MimMim Software, Inc. · 2011 · 2 generations on99descendants
- K0504794D Integrated Treatment ConsoleVarian Medical Systems · 2005 · cites it directly72descendants
- K0810364D Integrated Treatment Console with Patient Accessory Verification System (Pavs)Varian Medical Systems, Inc. · 2008 · 2 generations on69descendants
- K151913MIM-Thin Client (mobile)Mim Software, Inc. · 2016 · 3 generations on68descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K001643 | Varis 1.4G | Varian Medical Systems, Inc. | 2000 | — |
| +1 | K042956 | Vision | Varian Medical Systems, Inc. | 2005 | — |
| +2 | K112930 | Mobile Mim | Mim Software, Inc. | 2011 | — |
| +1 | K050479 | 4D Integrated Treatment Console | Varian Medical Systems | 2005 | 2 |
| +2 | K081036 | 4D Integrated Treatment Console with Patient Accessory Verification System… | Varian Medical Systems, Inc. | 2008 | 3 |
| +3 | K151913 | MIM-Thin Client (mobile) | Mim Software, Inc. | 2016 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2005 | 3 | 0 |
| 2008 | 1 | 0 |
| 2009 | 3 | 0 |
| 2010 | 1 | 0 |
| 2011 | 4 | 0 |
| 2012 | 6 | 0 |
| 2013 | 2 | 0 |
| 2014 | 8 | 0 |
| 2015 | 6 | 0 |
| 2016 | 5 | 0 |
| 2017 | 9 | 0 |
| 2018 | 13 | 0 |
| 2019 | 12 | 0 |
| 2020 | 9 | 0 |
| 2021 | 20 | 0 |
| 2022 | 15 | 0 |
| 2023 | 20 | 0 |
| 2024 | 10 | 0 |
| 2025 | 17 | 0 |
| 2026 | 19 | 0 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new IYE device
- #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it? (183)
See all 1832 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 7, 2011 | Not classified | Z-1957-2011 | Software design | An issue with the Varis Vision software where treatment can occur without the dynamic Multileaf Collimator component and may deliver a higher than expected dose of radiation. |
| Sep 15, 2009 | Not classified | Z-0133-2010 | Software design | Software issue: Under certain circumstances, when a therapy treatment field is transferred to VARiS Vision 6.2, 6.1 and 6.0 via LINK procedures, the rotation direction will be reversed. |
| May 8, 2009 | Not classified | Z-1809-2009 | Software design | Potential mistreatment-- due to software issues, retired plans may be used and treated repeatedly. |
| Jun 2, 2008 | Not classified | Z-0164-2009 | Software design | A malfunction within the Varis software might result in a misadministration (underdose). The software does not correctly interpret the number, causing field parameters having decimal values to be incorrect. |
| May 21, 2008 | Not classified | Z-2194-2008 | Software design | Dose delivery may be altered after import from a third party planning system during radiation therapy. |
| Oct 2, 2007 | Not classified | Z-0720-2008 | Software design | Incorrect Software Validation- The ARC treatment plans containing a Dose Dynamic MLC for IMRT delivery that will not be recognized properly by the system. Under these circumstances, an arc plan can be administered to a patient whereby the… |
| Sep 21, 2006 | Not classified | Z-0304-2007 | Software design | Selecting an empty space next to the ''Override'' or ''Acquire Actual'' buttons may result in mistreatment to the patient. |
| Jan 31, 2003 | Not classified | Z-0633-03 | Software design | The product has a software problem. This could result in a treatment without a motorized wedge where one had been planned. |
The 5 latest of 8 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K042956 (Varian Medical Systems, Inc., 2005) has gone furthest, with 121 descendants of its own.
How does its category compare?
All IYE devicesIYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · IYE
12in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · IYE
170in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · IYE
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · IYE
101in 2025
MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 23 | 44 | 78 | 155 |
| 2008 | 14 | 69 | 49 | 114 |
| 2009 | 20 | 63 | 22 | 174 |
| 2010 | 23 | 74 | 27 | 246 |
| 2011 | 11 | 94 | 73 | 280 |
| 2012 | 28 | 106 | 20 | 128 |
| 2013 | 18 | 88 | 11 | 96 |
| 2014 | 20 | 100 | 14 | 124 |
| 2015 | 20 | 108 | 8 | 100 |
| 2016 | 21 | 97 | 7 | 97 |
| 2017 | 23 | 94 | 7 | 125 |
| 2018 | 21 | 58 | 7 | 85 |
| 2019 | 19 | 83 | 2 | 91 |
| 2020 | 20 | 67 | 4 | 74 |
| 2021 | 19 | 86 | 5 | 68 |
| 2022 | 12 | 172 | 3 | 66 |
| 2023 | 18 | 129 | 7 | 128 |
| 2024 | 9 | 239 | 1 | 81 |
| 2025 | 12 | 170 | 1 | 101 |
| 2026 | 14 | 120 | 4 | — |
https://510kreport.com/device/K001643 · printed Nov 6, 2026
Cite this page
510(k) Report. “K001643: Varis 1.4G.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K001643
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026