510(k) · Cleared
Recalled · 20064D Integrated Treatment Console
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Varian Medical Systems · IYE Accelerator, Linear, Medical
Cleared on Mar 24, 2005 as substantially equivalent to K001643 (Varian Medical Systems, Inc., 2000) after a 28-day review. Its family tree reaches back 1 generation to 2000. 72 later devices descend from it. It has been recalled 2 times.
- Decided
- Mar 24, 2005
- Review
- 28 days
- Class
- Class II
- Product code
- IYE
- Recalls
- 2
- Predicates
- 1
Its longest line of descent runs back 1 generation to K001643 (2000). 72 later devices descend from it, 3 citing it directly.
- Root · 1 generation backK001643 Varis 1.4GVarian Medical Systems, Inc. · 2000 · Recalled
- This deviceK050479 4D Integrated Treatment ConsoleVarian Medical Systems · 2005
- Descendant · cites it directly · 69 of its ownK081036 4D Integrated Treatment Console with Patient Accessory Verification…Varian Medical Systems, Inc. · 2008 · Recalled
- Descendant · 2 generations on · 61 of its ownK092871 Trilogy MXVarian Medical Systems, Inc. · 2009 · Recalled
- Descendant · 3 generations on · 43 of its ownK111106 TruebeamVarian Medical Systems, Inc. · 2011 · Recalled
- Descendant · 4 generations on · 37 of its ownK123291 TruebeamVarian Medical Systems, Inc. · 2012
- Descendant · 4 generations on · 36 of its ownK123808 AgilityElekta, Ltd. · 2013 · Recalled
Showing 5 of 72 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 72 · See all- K0810364D Integrated Treatment Console with Patient Accessory Verification System (Pavs)Varian Medical Systems, Inc. · 2008 · cites it directly69descendants
- K092871Trilogy MXVarian Medical Systems, Inc. · 2009 · 2 generations on61descendants
- K111106TruebeamVarian Medical Systems, Inc. · 2011 · 3 generations on43descendants
- K123291TruebeamVarian Medical Systems, Inc. · 2012 · 4 generations on37descendants
- K123808AgilityElekta, Ltd. · 2013 · 4 generations on36descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K001643 | Varis 1.4G | Varian Medical Systems, Inc. | 2000 | 8 |
| 0 (this device) | K050479 | 4D Integrated Treatment Console | Varian Medical Systems | 2005 | — |
| +1 | K081036 | 4D Integrated Treatment Console with Patient Accessory Verification System… | Varian Medical Systems, Inc. | 2008 | 3 |
| +2 | K092871 | Trilogy MX | Varian Medical Systems, Inc. | 2009 | 5 |
| +3 | K111106 | Truebeam | Varian Medical Systems, Inc. | 2011 | 5 |
| +4 | K123291 | Truebeam | Varian Medical Systems, Inc. | 2012 | — |
| +4 | K123808 | Agility | Elekta, Ltd. | 2013 | 2 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2008 | 1 | 0 |
| 2009 | 2 | 0 |
| 2010 | 1 | 0 |
| 2011 | 3 | 0 |
| 2012 | 3 | 0 |
| 2013 | 2 | 0 |
| 2014 | 4 | 0 |
| 2015 | 4 | 0 |
| 2016 | 2 | 0 |
| 2017 | 7 | 0 |
| 2018 | 5 | 0 |
| 2019 | 5 | 0 |
| 2020 | 2 | 0 |
| 2021 | 9 | 0 |
| 2022 | 4 | 0 |
| 2023 | 9 | 0 |
| 2024 | 2 | 0 |
| 2025 | 2 | 0 |
| 2026 | 5 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new IYE device
- #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K001643 | Varis 1.4GSummary page 1…I Product Code: 90 iYE 3. Predicate Devices to claim substantial equivalence: a. Varian Medical Systems' Varis l.4g (K001643) 4. Device Description: The 4D Integrated Treatment Console provides the user with: * a treatment verification… | 2000 |
What descends from it? (72)
See all 723 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 21, 2006 | Not classified | Z-0304-2007 | Software design | Selecting an empty space next to the ''Override'' or ''Acquire Actual'' buttons may result in mistreatment to the patient. |
| Sep 20, 2005 | Not classified | Z-1375-06 | Software design | A software anomaly may occur which can lead to patient treatment with the wrong field. The anomaly is reported only to occur when this software version (6.6.5022) is used with Elekta or Siemens linear accelerators . |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All IYE devicesIYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · IYE
12in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · IYE
170in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · IYE
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · IYE
101in 2025
MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 23 | 44 | 78 | 155 |
| 2008 | 14 | 69 | 49 | 114 |
| 2009 | 20 | 63 | 22 | 174 |
| 2010 | 23 | 74 | 27 | 246 |
| 2011 | 11 | 94 | 73 | 280 |
| 2012 | 28 | 106 | 20 | 128 |
| 2013 | 18 | 88 | 11 | 96 |
| 2014 | 20 | 100 | 14 | 124 |
| 2015 | 20 | 108 | 8 | 100 |
| 2016 | 21 | 97 | 7 | 97 |
| 2017 | 23 | 94 | 7 | 125 |
| 2018 | 21 | 58 | 7 | 85 |
| 2019 | 19 | 83 | 2 | 91 |
| 2020 | 20 | 67 | 4 | 74 |
| 2021 | 19 | 86 | 5 | 68 |
| 2022 | 12 | 172 | 3 | 66 |
| 2023 | 18 | 129 | 7 | 128 |
| 2024 | 9 | 239 | 1 | 81 |
| 2025 | 12 | 170 | 1 | 101 |
| 2026 | 14 | 120 | 4 | — |
What is it indicated for?
The 4D Integrated Treatment Console is designed to assist the operator of a radiation therapy device in providing accurate treatment set-ups for each patient by monitoring set up parameters and preventing the radiation therapy device from commencing irradiation while any parameter is out of conformance with the treatment plan.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K050479 · printed Nov 6, 2026
Cite this page
510(k) Report. “K050479: 4D Integrated Treatment Console.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K050479
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026