510(k) Report

510(k) · Cleared

Recalled · 2008

Hitachi Altaire

K002420510(k) AbbreviatedCleared Oct 17, 200070-day reviewClass II21 CFR 892.1000Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hitachi Medical Systems America, Inc. · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Oct 17, 2000 after a 70-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 19 later devices descend from it. It has been recalled 8 times.

Decided
Oct 17, 2000
Review
70 days
Class
Class II
Product code
LNH
Recalls
8
Predicates
0

19 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2005: 2 descendants cleared2002: 1 descendant cleared2006: 1 descendant cleared2021: 1 descendant cleared2009: 1 descendant cleared2007: 1 descendant cleared2012: 1 descendant cleared2022: 1 descendant cleared2015: 2 descendants cleared2016: 3 descendants cleared2018: 1 descendant cleared2023: 1 descendant cleared2024: 2 descendants cleared2025: 1 descendant cleared200020052010201520202025
  1. This deviceK002420 Hitachi AltaireHitachi Medical Systems America, Inc. · 2000
  2. Descendant · cites it directly · 17 of its ownK050602 AltaireHitachi Medical Systems America, Inc. · 2005
  3. Descendant · 2 generations on · 16 of its ownK052172 HHF1 Magnetic Resonance Imaging SystemHitachi Medical Systems America, Inc. · 2005 · Recalled
  4. Descendant · 3 generations on · 11 of its ownK083533 Echelon C Mri SystemHitachi Medical Systems America, Inc. · 2009 · Recalled
  5. Descendant · 4 generations on · 10 of its ownK113145 Echelon Oval Mri SystemHitachi Medical Systems America, Inc. · 2012 · Recalled
  6. Descendant · 5 generations on · 6 of its ownK151015 ECHELON Oval V 5.0 MRI SystemHitachi Medical Systems America, Inc. · 2015 · Recalled

Showing 5 of 19 descendants. See all descendants

Descendants cleared per year

2002: 12005: 22006: 12007: 12009: 12012: 12015: 22016: 32018: 12021: 12022: 12023: 12024: 22025: 1
2002high 3 (2016) · low 0 (2003)2026

Most influential descendants

Showing 5 of 19 · See all
  • K050602AltaireHitachi Medical Systems America, Inc. · 2005 · cites it directly17descendants
  • K052172HHF1 Magnetic Resonance Imaging SystemHitachi Medical Systems America, Inc. · 2005 · 2 generations on16descendants
  • K083533Echelon C Mri SystemHitachi Medical Systems America, Inc. · 2009 · 3 generations on11descendants
  • K113145Echelon Oval Mri SystemHitachi Medical Systems America, Inc. · 2012 · 4 generations on10descendants
  • K151015ECHELON Oval V 5.0 MRI SystemHitachi Medical Systems America, Inc. · 2015 · 5 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K002420Hitachi AltaireHitachi Medical Systems America, Inc.2000—
+1K050602AltaireHitachi Medical Systems America, Inc.2005—
+2K052172HHF1 Magnetic Resonance Imaging SystemHitachi Medical Systems America, Inc.20053
+3K083533Echelon C Mri SystemHitachi Medical Systems America, Inc.20092
+4K113145Echelon Oval Mri SystemHitachi Medical Systems America, Inc.20123
+5K151015ECHELON Oval V 5.0 MRI SystemHitachi Medical Systems America, Inc.20151

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200210
200520
200610
200710
200910
201210
201520
201630
201810
202110
202210
202310
202420
202510
FIG. 1 — Lineage of K002420, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LNH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (19)

See all 19

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K050602AltaireHitachi Medical Systems America, Inc.200517
K021751Altaire Magnetic Resonance Imaging SystemHitachi Medical Systems America, Inc.20020

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 11, 2008Not classifiedZ-1542-2008Component design/selectionThe firm's Altaire MR Scanner's Emergency Rundown Switch Unit (ERDU), that is used to shut down the device in an emergency, was found to be defective. An ERDU failure could delay or prevent patient access leading to possible patient death…
Jul 30, 2007Not classifiedZ-1263-2007OtherOverheating; A portion (balun assembly) of the cable that connects the shoulder coil to the actual MRI unit may malfunction and result in overheating of the balun assembly. The overheating of the balun assembly could result in injury to…
Apr 28, 2006Not classifiedZ-1104-06OtherSoftware anomaly-A software defect can cause the slice line indicator to become mis-positioned during Multi-Planar Reconstructions (MPR) with the potential for patient mis-diagnosis.
Feb 21, 2006Not classifiedZ-0910-06OtherThe device has a software anomaly which causes the operating system to 'lock up' and data to become corrupted during service procedures.
Jan 27, 2006Not classifiedZ-1103-06OtherCircuit Failure-There is the potential risk of the Phase Loss Detector Safety Device (PHSEN PCB) malfunctioning in the event of a power loss. The circuit may fail to turn off the system main breaker when 2 of the 3 power phases are lost.
Jun 30, 2005Not classifiedZ-0888-05OtherImage reconstruction software problem that could potentially cause multiple patient images to be stored in one patient folder.
Apr 11, 2005Not classifiedZ-0797-05OtherThe electrical cables that supply gradient power to the magnet can become loose over time and create a potential fire hazard.
Nov 18, 2004Not classifiedZ-0846-05OtherThe absorber unit in the helium compressor which is located in equipment rooms, and is not accessed by patients or system operators, may fail under pressure, causing an explosion that damages the compressor.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K050602 (Hitachi Medical Systems America, Inc., 2005) has gone furthest, with 17 descendants of its own.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—
Cite this page

510(k) Report. “K002420: Hitachi Altaire.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K002420

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026