510(k) Report

510(k) · Cleared

Recalled · 2008

Ventilator, Continuous (Respirator), Model LTV 1000

K002881510(k) SpecialCleared Oct 11, 200026-day reviewClass II21 CFR 868.5895Recalls 12

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pulmonetic Systems, Inc. · CBK Ventilator, Continuous, Facility Use

Cleared on Oct 11, 2000 after a 26-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 2 of the 11 devices in its family, counting itself, have a Class I recall. 10 later devices descend from it. It has been recalled 12 times.

Decided
Oct 11, 2000
Review
26 days
Class
Class II
Product code
CBK
Recalls
12
Predicates
0

10 later devices descend from it, 2 citing it directly. 2 of the 11 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2009: 1 descendant cleared, 1 with a Class I recall12008: 1 descendant cleared2010: 1 descendant cleared2012: 2 descendants cleared2011: 2 descendants cleared, 1 with a Class I recall12014: 1 descendant cleared2016: 1 descendant cleared2024: 1 descendant cleared200020052010201520202025
  1. This deviceK002881 Ventilator, Continuous (Respirator), Model LTV 1000Pulmonetic Systems, Inc. · 2000
  2. Descendant · cites it directly · 9 of its ownK082724 Newport HTS50 Ventilator with Dual Pac Battery SystemNewport Medical Instruments, Inc. · 2008 · Recalled
  3. Descendant · 2 generations on · 5 of its ownK100753 Flight 60 Ventilator, Model F-60Flight Medical, Ltd. · 2010
  4. Descendant · 3 generations on · 4 of its ownK111683 Flight 60 VentilatorFlight Medical Innovations, Ltd. · 2011
  5. Descendant · 4 generations on · 3 of its ownK120726 Flight 60 VentilatorFlight Medical Innovations, Ltd. · 2012
  6. Descendant · 5 generations on · 2 of its ownK130171 Flight 60 VentilatorFlight Medical Innovations, Ltd. · 2014

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2008: 12009: 12010: 12011: 22012: 22014: 12016: 12024: 1
2008high 2 (2011) · low 0 (2013)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 10 · See all
  • K082724Newport HTS50 Ventilator with Dual Pac Battery SystemNewport Medical Instruments, Inc. · 2008 · cites it directly9descendants
  • K100753Flight 60 Ventilator, Model F-60Flight Medical, Ltd. · 2010 · 2 generations on5descendants
  • K111683Flight 60 VentilatorFlight Medical Innovations, Ltd. · 2011 · 3 generations on4descendants
  • K120726Flight 60 VentilatorFlight Medical Innovations, Ltd. · 2012 · 4 generations on3descendants
  • K130171Flight 60 VentilatorFlight Medical Innovations, Ltd. · 2014 · 5 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K002881Ventilator, Continuous (Respirator), Model LTV 1000Pulmonetic Systems, Inc.2000—
+1K082724Newport HTS50 Ventilator with Dual Pac Battery SystemNewport Medical Instruments, Inc.20082
+2K100753Flight 60 Ventilator, Model F-60Flight Medical, Ltd.2010—
+3K111683Flight 60 VentilatorFlight Medical Innovations, Ltd.2011—
+4K120726Flight 60 VentilatorFlight Medical Innovations, Ltd.2012—
+5K130171Flight 60 VentilatorFlight Medical Innovations, Ltd.2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
200911
201010
201121
201220
201410
201610
202410
FIG. 1 — Lineage of K002881, 2000–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (10)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K082724Newport HTS50 Ventilator with Dual Pac Battery SystemNewport Medical Instruments, Inc.20089
K090888Newport HT70 Family of Ventilators Class INewport Medical Instruments, Inc.20092

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 22, 2008Not classifiedZ-1463-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1469-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1468-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1467-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1464-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1470-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1466-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1465-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Nov 9, 2005Not classifiedZ-0184-06OtherPulmonetic Systems has identified the potential of a printed circuit board malfunction, which can result in a ventilator malfunctio (e.g. Vent INOP, HW Fault, Xducer Fault) and possibly resulting in failure of the ventilator to breathe…
Nov 9, 2005Not classifiedZ-0186-06OtherPulmonetic Systems has identified the potential of a printed circuit board malfunction, which can result in a ventilator malfunctio (e.g. Vent INOP, HW Fault, Xducer Fault) and possibly resulting in failure of the ventilator to breathe…
Nov 9, 2005Not classifiedZ-0185-06OtherPulmonetic Systems has identified the potential of a printed circuit board malfunction, which can result in a ventilator malfunctio (e.g. Vent INOP, HW Fault, Xducer Fault) and possibly resulting in failure of the ventilator to breathe…
Sep 8, 2004Not classifiedZ-1485-04OtherVentilators have not consistently switched to internal battery operation when the external DC power source has become inadequate to supply stable power. This situation results in loss of ventilation to the patient.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K090888Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc. · 2009 · Cites it directly · ◆ in the 2009 descendants bar of FIG. 13 Class I recalls
  • K111146Newport HT70 VentilatorNewport Medical Instruments, Inc. · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 14 Class I recalls

What does this lineage say?

  • Of its descendants, K082724 (Newport Medical Instruments, Inc., 2008) has gone furthest, with 9 descendants of its own.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—
Cite this page

510(k) Report. “K002881: Ventilator, Continuous (Respirator), Model LTV 1000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K002881

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026