510(k) Report

510(k) · Cleared

Recalled · Class I · 2025

Newport HT70 Ventilator

K111146510(k) TraditionalCleared Dec 1, 2011223-day reviewClass II21 CFR 868.5895Recalls 4 (4 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Newport Medical Instruments, Inc. · CBK Ventilator, Continuous, Facility Use

Cleared on Dec 1, 2011 as substantially equivalent to K090888 (Newport Medical Instruments, Inc., 2009) and 2 other predicates after a 223-day review. Its family tree reaches back 12 generations to 1991. 9 of the 85 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 4 times, including 4 Class I recalls.

Decided
Dec 1, 2011
Review
223 days
Class
Class II
Product code
CBK
Recalls
4 (4 Class I)
Predicates
3

Its longest line of descent runs back 12 generations to K902859 (1991). 83 ancestors in all, the oldest K841279 (1984). 1 later device descends from it, 1 citing it directly. 9 of the 85 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2008: 1 other ancestor2007: 1 other ancestor, 1 with a Class I recall12005: 2 other ancestors2004: 5 other ancestors, 1 with a Class I recall12003: 3 other ancestors, 1 with a Class I recall12002: 3 other ancestors2001: 6 other ancestors2000: 5 other ancestors, 1 with a Class I recall11999: 6 other ancestors1998: 11 other ancestors, 2 with a Class I recall21996: 3 other ancestors1995: 3 other ancestors1990: 6 other ancestors1997: 3 other ancestors1984: 1 other ancestor1994: 3 other ancestors1993: 1 other ancestor1992: 2 other ancestors1991: 4 other ancestors1988: 1 other ancestor2012: 1 descendant clearedK902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalledK960010 · Servo Ventilator 300 Computer Interface · Siemens Elema AB · 1996K970839 · Servo Ventilator 300A · Siemens Elema AB · 1997 · recalledK992788 · Bear Cub 750 PSV · Bear Medical Systems, Inc. · 1999K013642 · AVEA · Bird Products Corp. · 2002 · Class I recallK032451 · Bird Tbird Vela Ventilator · Bird Products Corp. · 2004 · recalled×3: 3 generations cleared close together, K052554 (2005), K053270 (2006), K061627 (2006) K063404 · Servo-I Ventilator System, Models 64, 87 and 800 · Maquet Critical Care AB · 2007K073694 · Ivent 2001 Mri Conditional · Versamed Medical Systems, Inc. · 2008K090888 · Newport HT70 Family of Ventilators · Newport Medical Instruments, Inc. · 2009 · Class I recall19801990200020102020
  1. Root · 12 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 8 generations between
  3. 3 generations backK063404 Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007
  4. 2 generations backK073694 Ivent 2001 Mri ConditionalVersamed Medical Systems, Inc. · 2008
  5. 1 generation backK090888 Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc. · 2009 · Recalled (Class I)
  6. This deviceK111146 Newport HT70 VentilatorNewport Medical Instruments, Inc. · 2011
  7. Descendant · cites it directly · 0 of its ownK121891 Newport Aura VentilatorNewport Medical Instruments, Inc. · 2012
Show all 12 generations
  1. Root · 12 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 11 generations backK960010 Servo Ventilator 300 Computer InterfaceSiemens Elema AB · 1996
  3. 10 generations backK970839 Servo Ventilator 300ASiemens Elema AB · 1997 · Recalled
  4. 9 generations backK992788 Bear Cub 750 PSVBear Medical Systems, Inc. · 1999
  5. 8 generations backK013642 AVEABird Products Corp. · 2002 · Recalled (Class I)
  6. 7 generations backK032451 Bird Tbird Vela VentilatorBird Products Corp. · 2004 · Recalled
  7. 6 generations backK052554 Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc. · 2005
  8. 5 generations backK053270 IVENT 201Versamed Medical Systems, Inc. · 2006
  9. 4 generations backK061627 Modification to I Vent 201Versamed Medical Systems, Inc. · 2006
  10. 3 generations backK063404 Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007
  11. 2 generations backK073694 Ivent 2001 Mri ConditionalVersamed Medical Systems, Inc. · 2008
  12. 1 generation backK090888 Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc. · 2009 · Recalled (Class I)
  13. This deviceK111146 Newport HT70 VentilatorNewport Medical Instruments, Inc. · 2011

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K121891Newport Aura VentilatorNewport Medical Instruments, Inc. · 2012 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K902859Servo Ventilator 300Siemens Life Support Systems19913
−11K960010Servo Ventilator 300 Computer InterfaceSiemens Elema AB1996—
−10K970839Servo Ventilator 300ASiemens Elema AB19974
−9K992788Bear Cub 750 PSVBear Medical Systems, Inc.1999—
−8K013642AVEABird Products Corp.2002Class I
−7K032451Bird Tbird Vela VentilatorBird Products Corp.20041
−6K052554Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc.2005—
−5K053270IVENT 201Versamed Medical Systems, Inc.2006—
−4K061627Modification to I Vent 201Versamed Medical Systems, Inc.2006—
−3K063404Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB2007—
−2K073694Ivent 2001 Mri ConditionalVersamed Medical Systems, Inc.2008—
−1K090888Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc.2009Class I
0 (this device)K111146Newport HT70 VentilatorNewport Medical Instruments, Inc.2011—
+1K121891Newport Aura VentilatorNewport Medical Instruments, Inc.2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198410
198810
199060
199140
199220
199310
199430
199530
199630
199730
1998112
199960
200051
200160
200230
200331
200451
200520
200610
200711
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
FIG. 1 — Lineage of K111146, 1991–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

9 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K090888Newport HT70 Family of Ventilators
Summary page 1
…The HT7O is substantially equivalent to the following legally marketed predicate devices: * NEWPORT HT70® Ventilator cleared under K090888; * CareFusion LTV0 1200 cleared under K060647; * Versamed Medical Systems iVENTTM 201 cle…
Newport Medical Instruments, Inc.2009Class I
*K060647LTV 1200 Ventilator
Summary page 1
…y marketed predicate devices: * NEWPORT HT70® Ventilator cleared under K090888; * CareFusion LTV0 1200 cleared under K060647; * Versamed Medical Systems iVENTTM 201 cleared under K073694. IV. Device Description…
Pulmonetic Systems, Inc.2006—
*K073694Ivent 2001 Mri Conditional
Summary page 1
…tilator cleared under K090888; * CareFusion LTV0 1200 cleared under K060647; * Versamed Medical Systems iVENTTM 201 cleared under K073694. IV. Device Description The Newport HT70 Ventilator provides ventilatory support for neonate, infant,…
Versamed Medical Systems, Inc.2008—

Also cleared against K090888: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K121891Newport Aura VentilatorNewport Medical Instruments, Inc.20120

Has anything in its family been recalled? (9 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 21, 2025Class IZ-1915-2025Component design/selectionVentilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown…
Mar 30, 2017Class IZ-1874-2017Software in the Use EnvironmentMedtronic is issuing a voluntary field corrective action for all its Newport" HT70 and Newport" HT70 Plus ventilators because they may shutdown spontaneously during normal operation without an accompanying alarm.
Apr 5, 2013Class IZ-1171-2013Component design/selectionNewport Medical Instruments is conducting a voluntary recall on certain Newport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries due to customer reports of Newport HT70 and HT70 Plus ventilators alarming and going to…
Aug 3, 2012Class IZ-2250-2012Nonconforming Material/ComponentMay emit a continuous high priority alarm and the ventilator may stop ventilating, due to a component failure on the control board.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K090888Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc. · 2009 · Direct predicate · ◆ on the line of descent in FIG. 13 Class I recalls
  • K070594Palmtop Ventilator, Models PTV-8, PTV-10Pulmonetic Systems, Inc. · 2007 · 2 generations back · ◆ in the 2007 ancestors bar of FIG. 12 Class I recalls
  • K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · 3 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K013642AVEABird Products Corp. · 2002 · 3 generations back · ◆ on the line of descent in FIG. 113 Class I recalls
  • K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · 4 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K981072Esprit VentilatorRespironics, Inc. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K001646Puritan-Bennett 840 Ventilator System with Neomode OptionPuritan Bennett Corp. · 2000 · 6 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 20 years and passes through 7 companies; a new generation was cleared every 1.5 years on median.
  • 30 of the 83 ancestors have a recall of any class; 8 have a Class I recall.
  • Of its descendants, K121891 (Newport Medical Instruments, Inc., 2012) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K090888.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The Newport HT70 family of ventilators is intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces. Specifically, the Newport HT70 family of ventilators is applicable for infant, pediatric and adult patients greater than or equal to 5 kg (11 Ibs).

The Newport HT70 is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician.

The Newport HT70 is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K111146: Newport HT70 Ventilator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K111146

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026