510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Advia Centaur and Acs: 180 Folate Immunoassay

K010050510(k) TraditionalCleared Feb 5, 200131-day reviewClass II21 CFR 862.1295Recalls 14

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bayer Diagnostics Corp. · CGN Acid, Folic, Radioimmunoassay

Cleared on Feb 5, 2001 after a 31-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 7 later devices descend from it. It has been recalled 14 times.

Decided
Feb 5, 2001
Review
31 days
Class
Class II
Product code
CGN
Recalls
14
Predicates
0

7 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2018: 1 descendant cleared2007: 1 descendant cleared2009: 1 descendant cleared2010: 2 descendants cleared2012: 1 descendant cleared2014: 1 descendant cleared200020052010201520202025
  1. This deviceK010050 Advia Centaur and Acs: 180 Folate ImmunoassayBayer Diagnostics Corp. · 2001
  2. Descendant · cites it directly · 4 of its ownK071224 Dimension Vista Vitamin B12 and Folate Flex Reagent Cartridges and…Dade Behring, Inc. · 2007 · Recalled
  3. Descendant · 2 generations on · 1 of its ownK121994 Dimension Vista Vitamin B12 (VB12) Flex Reagent Cartridge, Dimension…Siemens Healthcare Diagnostics, Inc. · 2012
  4. Descendant · cites it directly · 0 of its ownK172201 Atellica IM Folate AssaySiemens Healthcare Diagnostics, Inc. · 2018 · Recalled
  5. Descendant · 3 generations on · 0 of its ownK133512 Loci Vitamin B12 (VB12) Flex Reagent Cartridge, Loci Anemia…Siemens Healthcare Diagnostics · 2014
  6. Descendant · 2 generations on · 0 of its ownK093645 Exl Anem Cal, Model RC640Siemens Healthcare Diagnostics · 2010

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2007: 12009: 12010: 22012: 12014: 12018: 1
2007high 2 (2010) · low 0 (2008)2026

Most influential descendants

Showing 5 of 7 · See all
  • K071224Dimension Vista Vitamin B12 and Folate Flex Reagent Cartridges and Loci 4 CalibratorDade Behring, Inc. · 2007 · cites it directly4descendants
  • K121994Dimension Vista Vitamin B12 (VB12) Flex Reagent Cartridge, Dimension Vista Loci 4…Siemens Healthcare Diagnostics, Inc. · 2012 · 2 generations on1descendant
  • K172201Atellica IM Folate AssaySiemens Healthcare Diagnostics, Inc. · 2018 · cites it directly0descendants
  • K133512Loci Vitamin B12 (VB12) Flex Reagent Cartridge, Loci Anemia Calibrator (Anem Cal)Siemens Healthcare Diagnostics · 2014 · 3 generations on0descendants
  • K093645Exl Anem Cal, Model RC640Siemens Healthcare Diagnostics · 2010 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K010050Advia Centaur and Acs: 180 Folate ImmunoassayBayer Diagnostics Corp.2001—
+1K071224Dimension Vista Vitamin B12 and Folate Flex Reagent Cartridges and Loci 4…Dade Behring, Inc.20071
+2K121994Dimension Vista Vitamin B12 (VB12) Flex Reagent Cartridge, Dimension Vista…Siemens Healthcare Diagnostics, Inc.2012—
+1K172201Atellica IM Folate AssaySiemens Healthcare Diagnostics, Inc.20183
+3K133512Loci Vitamin B12 (VB12) Flex Reagent Cartridge, Loci Anemia Calibrator (Anem…Siemens Healthcare Diagnostics2014—
+2K093645Exl Anem Cal, Model RC640Siemens Healthcare Diagnostics2010—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
200910
201020
201210
201410
201810
FIG. 1 — Lineage of K010050, 2001–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CGN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CGN device

  1. #1 K233060 Elecsys Folate IIIscore 9.9 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (7)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K071224Dimension Vista Vitamin B12 and Folate Flex Reagent Cartridges and Loci 4 CalibratorDade Behring, Inc.20074
K172201Atellica IM Folate AssaySiemens Healthcare Diagnostics, Inc.20180
K091332Aia-Pack RBC FolateTosoh Bioscience, Inc.20090

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 9, 2023Class IIZ-1152-2023Under Investigation by firmNegative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.
Jan 9, 2023Class IIZ-1153-2023Under Investigation by firmNegative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.
Jan 9, 2023Class IIZ-1154-2023Under Investigation by firmNegative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.
Jul 22, 2019Class IIZ-0831-2020Device DesignHomocysteine Assay May Cause Elevated Results in the Folate Assay
Jul 22, 2019Class IIZ-0829-2020Device DesignHomocysteine Assay May Cause Elevated Results in the Folate Assay
Jul 22, 2019Class IIZ-0830-2020Device DesignHomocysteine Assay May Cause Elevated Results in the Folate Assay
Jan 2, 2018Class IIZ-1689-2018Error in labelingConcentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the…
Jul 27, 2015Class IIZ-2739-2015Process designLow red blood cell control recoveries and low results for red blood cell patient samples when using the ADVIA Centaur Systems Folate assay.
Mar 25, 2014Class IIZ-1564-2014Nonconforming Material/ComponentSerum controls may go out of range low with the ADVIA Centaur ¿Systems Folate assay and a shift in whole blood patient may result with kit lots ending in 219 and 223
Jul 25, 2013Class IIZ-2172-2013Material/Component ContaminationCalibration failures and/or significant negative shifts in quality control (QC) and patient results.
Jul 25, 2013Class IIZ-2170-2013Material/Component ContaminationCalibration failures and/or significant negative shifts in quality control (QC) and patient results.
Jul 25, 2013Class IIZ-2171-2013Material/Component ContaminationCalibration failures and/or significant negative shifts in quality control (QC) and patient results.
Jul 25, 2013Class IIZ-2169-2013Material/Component ContaminationCalibration failures and/or significant negative shifts in quality control (QC) and patient results.
Nov 14, 2012Class IIZ-0511-2013Device DesignOperating range of the ADVIA Centaur systems is 18C to 30C (64.4F to 86.0F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K071224 (Dade Behring, Inc., 2007) has gone furthest, with 4 descendants of its own.

How does its category compare?

All CGN devices

CGN Acid, Folic, Radioimmunoassay: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CGN

1in 2024

2009: 22010: 12011: 12014: 12016: 12018: 12020: 12023: 12024: 1
2007high 2 (2009) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CGN

Too few CGN 510(k)s in any year since 2007 to give a median.

Recalls per year · CGN

3in 2019

2007: 22008: 12009: 12010: 12012: 22013: 12018: 22019: 3
2007high 3 (2019) · low 0 (2011)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CGN

9in 2025

2007: 262009: 32011: 282012: 112013: 22014: 42015: 12016: 42017: 62018: 22019: 12020: 52021: 32022: 132023: 82024: 192025: 9
2007high 28 (2011) · low 0 (2008)2025

MAUDE reports naming product code CGN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—226
20080—10
20092—13
20101—10
20111—028
20120—211
20130—12
20141—04
20150—01
20161—04
20170—06
20181—22
20190—31
20201—05
20210—03
20220—013
20231—08
20241—019
20250—09
20260—0—
Cite this page

510(k) Report. “K010050: Advia Centaur and Acs: 180 Folate Immunoassay.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K010050

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026