Dimension Vista Vitamin B12 and Folate Flex Reagent Cartridges and Loci 4 Calibrator

K071224510(k) TraditionalCleared Nov 13, 2007195-day reviewClass II21 CFR 862.1810Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Dade Behring, Inc. · CDD Radioassay, Vitamin B12

Cleared on Nov 13, 2007 as substantially equivalent to K041133 (Bayer Healthcare, LLC, 2004) and 2 other predicates after a 195-day review. Its family tree reaches back 4 generations to 1990. 4 later devices descend from it. It has been recalled once.

Decided
Nov 13, 2007
Review
195 days
Class
Class II
Product code
CDD
Recalls
1
Predicates
3

Its longest line of descent runs back 4 generations to K902336 (1990). 7 ancestors in all. 4 later devices descend from it, 3 citing it directly.

  1. Root · 4 generations backK902336 Ciba Corning Acs:180 Automated ChemiluminescenceCiba Corning Diagnostics Corp. · 1990
  2. 3 generations backK971418 Chiron Diagnostics Automated Chemiluminescence System (Asc) Centaur…Chiron Diagnostics Corp. · 1997 · Recalled
  3. 2 generations backK032525 Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC · 2003
  4. 1 generation backK041133 Modification to Bayer Advia Centaur Automated Chemiluminescence…Bayer Healthcare, LLC · 2004 · Recalled
  5. This deviceK071224 Dimension Vista Vitamin B12 and Folate Flex Reagent Cartridges and…Dade Behring, Inc. · 2007
  6. Descendant · cites it directly · 1 of its ownK121994 Dimension Vista Vitamin B12 (VB12) Flex Reagent Cartridge, Dimension…Siemens Healthcare Diagnostics, Inc. · 2012
  7. Descendant · 2 generations on · 0 of its ownK133512 Loci Vitamin B12 (VB12) Flex Reagent Cartridge, Loci Anemia…Siemens Healthcare Diagnostics · 2014
  8. Descendant · cites it directly · 0 of its ownK093645 Exl Anem Cal, Model RC640Siemens Healthcare Diagnostics · 2010
  9. Descendant · cites it directly · 0 of its ownK093631 B12 Reagent Cartridge and Fol Flex Reagent Cartridge, Model RF642…Siemens Healthcare Diagnostics · 2010

Most influential descendants

Showing 4 of 4 · See all
  • K121994Dimension Vista Vitamin B12 (VB12) Flex Reagent Cartridge, Dimension Vista Loci 4…Siemens Healthcare Diagnostics, Inc. · 2012 · cites it directly1descendant
  • K133512Loci Vitamin B12 (VB12) Flex Reagent Cartridge, Loci Anemia Calibrator (Anem Cal)Siemens Healthcare Diagnostics · 2014 · 2 generations on0descendants
  • K093645Exl Anem Cal, Model RC640Siemens Healthcare Diagnostics · 2010 · cites it directly0descendants
  • K093631B12 Reagent Cartridge and Fol Flex Reagent Cartridge, Model RF642 and RF644Siemens Healthcare Diagnostics · 2010 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K902336Ciba Corning Acs:180 Automated ChemiluminescenceCiba Corning Diagnostics Corp.1990—
−3K971418Chiron Diagnostics Automated Chemiluminescence System (Asc) Centaur AnalyzerChiron Diagnostics Corp.19974
−2K032525Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC2003—
−1K041133Modification to Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC20048
0 (this device)K071224Dimension Vista Vitamin B12 and Folate Flex Reagent Cartridges and Loci 4…Dade Behring, Inc.2007—
+1K121994Dimension Vista Vitamin B12 (VB12) Flex Reagent Cartridge, Dimension Vista…Siemens Healthcare Diagnostics, Inc.2012—
+2K133512Loci Vitamin B12 (VB12) Flex Reagent Cartridge, Loci Anemia Calibrator (Anem…Siemens Healthcare Diagnostics2014—
+1K093645Exl Anem Cal, Model RC640Siemens Healthcare Diagnostics2010—
+1K093631B12 Reagent Cartridge and Fol Flex Reagent Cartridge, Model RF642 and RF644Siemens Healthcare Diagnostics2010—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199810
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201020
201210
201410
FIG. 1 — Lineage of K071224, 1990–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CDD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CDD device

  1. #1 K233541 ARCHITECT Active-B12 (Holotranscobalamin)score 8.7 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041133Modification to Bayer Advia Centaur Automated Chemiluminescence Analyzer
Summary page 1
…Product Device number Class Code B 12 Flex® reagent cartridge for VB 12 for the ADVIA K041133 II 862.1810 CDD the Dimension Vista® system Centaur® System FOL Flex® reagent cartridge foJ FOL for the ADVIA…
Bayer Healthcare, LLC20048
*K010050Advia Centaur and Acs: 180 Folate Immunoassay
Summary page 1
…the ADVIA K041133 II 862.1810 CDD the Dimension Vista® system Centaur® System FOL Flex® reagent cartridge foJ FOL for the ADVIA K010050 Ii 862.1295 CGN the Dimension Vista® system Centaur® system LOCI 4 Calibrator for the FERR Calibrator fo…
Bayer Diagnostics Corp.200114
*K983548Ferritin Calibrator
Summary page 1
…e ADVIA K010050 Ii 862.1295 CGN the Dimension Vista® system Centaur® system LOCI 4 Calibrator for the FERR Calibrator for the K983548 II 862.1150 JDX Dimension Vista® system Dimension system Device Description: B12 The Dimension Vista®…
Dade Behring, Inc.1998—

Also cleared against K041133: 4 other devices →

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K121994Dimension Vista Vitamin B12 (VB12) Flex Reagent Cartridge, Dimension Vista Loci 4…Siemens Healthcare Diagnostics, Inc.20121
K093645Exl Anem Cal, Model RC640Siemens Healthcare Diagnostics20100
K093631B12 Reagent Cartridge and Fol Flex Reagent Cartridge, Model RF642 and RF644Siemens Healthcare Diagnostics20100

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 8, 2012Class IIZ-1869-2012Device DesignSiemens Healthcare Diagnostics has confirmed that the presence of Intrinsic Factor Blocking Antibody in some patients can result in a false and significant elevation in the reported concentration of Vitamin B12 on the Dimension Vista…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 4 companies; a new generation was cleared every 6.1 years on median.
  • 3 of the 7 ancestors have a recall of any class.
  • Of its descendants, K121994 (Siemens Healthcare Diagnostics, Inc., 2012) has gone furthest, with 1 descendant of its own.
  • 4 other devices also cite its predicate K041133.

How does its category compare?

All CDD devices

CDD Radioassay, Vitamin B12: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · CDD

1in 2024

2007: 12009: 12010: 12011: 32012: 32013: 12014: 22015: 12016: 22017: 12019: 22022: 12024: 1
2007high 3 (2011) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CDD

147in 2012

2011: 117 days median2012: 147 days median
2007high 147 (2012) · low 117 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CDD

1in 2021

2007: 12018: 22019: 12021: 1
2007high 2 (2018) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CDD

44in 2025

2010: 42011: 362012: 232013: 452014: 122015: 42016: 62017: 92018: 152019: 142020: 112021: 282022: 242023: 272024: 332025: 44
2007high 45 (2013) · low 0 (2007)2025

MAUDE reports naming product code CDD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—10
20080—00
20091—00
20101—04
20113117036
20123147023
20131—045
20142—012
20151—04
20162—06
20171—09
20180—215
20192—114
20200—011
20210—128
20221—024
20230—027
20241—033
20250—044
20260—0—

What is it indicated for?

The B12 Flex® reagent cartridge is an in vitro diagnostic test for the quantitative measurement of Vitamin B12 in human serum and plasma on the Dimension Vista® system. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption. The FOL Flex® reagent cartridge is an in vitro diagnostic test for the quantitative measurement of folate in human serum and plasma on the Dimension Vista® system.…

Read the full statement (116 words)

The B12 Flex® reagent cartridge is an in vitro diagnostic test for the quantitative measurement of Vitamin B12 in human serum and plasma on the Dimension Vista® system. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption.

The FOL Flex® reagent cartridge is an in vitro diagnostic test for the quantitative measurement of folate in human serum and plasma on the Dimension Vista® system. Measurements obtained by this device are used in the diagnosis and treatment of megaloblastic anemia.

The LOCI 4 Calibrator is an in vitro diagnostic product for the calibration of Ferritin (FERR), Vitamin B 12 (B12), and Folate (FOL) methods on the Dimension Vista® system.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K071224: Dimension Vista Vitamin B12 and Folate Flex Reagent Cartridges and Loci 4 Calibrator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K071224

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026