510(k) Report

510(k) · Cleared

Recalled · 2006

Accu-Chek Comfort Curve Test Strip

K010362510(k) SpecialCleared Mar 7, 200129-day reviewClass II21 CFR 862.1345Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diagnostics Corp. · LFR Glucose Dehydrogenase, Glucose

Cleared on Mar 7, 2001 after a 29-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 1 of the 65 devices in its family, counting itself, has a Class I recall. 64 later devices descend from it. It has been recalled 6 times.

Decided
Mar 7, 2001
Review
29 days
Class
Class II
Product code
LFR
Recalls
6
Predicates
0

64 later devices descend from it, 1 citing it directly. 1 of the 65 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2005: 1 descendant cleared2011: 2 descendants cleared2009: 3 descendants cleared2016: 4 descendants cleared2013: 5 descendants cleared2006: 1 descendant cleared2008: 1 descendant cleared2012: 3 descendants cleared2007: 1 descendant cleared2015: 4 descendants cleared2017: 4 descendants cleared2014: 3 descendants cleared2018: 5 descendants cleared2023: 4 descendants cleared2022: 3 descendants cleared2020: 5 descendants cleared, 1 with a Class I recall12025: 5 descendants cleared2024: 3 descendants cleared2021: 2 descendants cleared2026: 5 descendants cleared200020052010201520202025
  1. This deviceK010362 Accu-Chek Comfort Curve Test StripRoche Diagnostics Corp. · 2001
  2. Descendant · cites it directly · 63 of its ownK043474 Accu-Chek Aviva SystemRoche Diagnostics Corp. · 2005 · Recalled
  3. Descendant · 2 generations on · 46 of its ownK060620 Accu-Chek Aviva Test StripsRoche Diagnostics Corp. · 2006 · Recalled
  4. Descendant · 2 generations on · 44 of its ownK101299 Accu-Chek Aviva Plus Blood Glucose Monitoring SystemRoche Diagnostics · 2011
  5. Descendant · 3 generations on · 37 of its ownK111353 Accu-Check Combo SystemRoche Diagnostics · 2012 · Recalled
  6. Descendant · 4 generations on · 33 of its ownK141929 Accu-Chek Connect Diabetes Management AppRoche Diagnostics Corporation · 2015 · Recalled

Showing 5 of 64 descendants. See all descendants

Descendants cleared per year

2005: 12006: 12007: 12008: 12009: 32011: 22012: 32013: 52014: 32015: 42016: 42017: 42018: 52020: 52021: 22022: 32023: 42024: 32025: 52026: 5
2005high 5 (2013) · low 0 (2010)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 64 · See all
  • K043474Accu-Chek Aviva SystemRoche Diagnostics Corp. · 2005 · cites it directly63descendants
  • K060620Accu-Chek Aviva Test StripsRoche Diagnostics Corp. · 2006 · 2 generations on46descendants
  • K101299Accu-Chek Aviva Plus Blood Glucose Monitoring SystemRoche Diagnostics · 2011 · 2 generations on44descendants
  • K111353Accu-Check Combo SystemRoche Diagnostics · 2012 · 3 generations on37descendants
  • K141929Accu-Chek Connect Diabetes Management AppRoche Diagnostics Corporation · 2015 · 4 generations on33descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K010362Accu-Chek Comfort Curve Test StripRoche Diagnostics Corp.2001—
+1K043474Accu-Chek Aviva SystemRoche Diagnostics Corp.200510
+2K060620Accu-Chek Aviva Test StripsRoche Diagnostics Corp.20061
+2K101299Accu-Chek Aviva Plus Blood Glucose Monitoring SystemRoche Diagnostics2011—
+3K111353Accu-Check Combo SystemRoche Diagnostics20123
+4K141929Accu-Chek Connect Diabetes Management AppRoche Diagnostics Corporation20151

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
200610
200710
200810
200930
201120
201230
201350
201430
201540
201640
201740
201850
202051
202120
202230
202340
202430
202550
202650
FIG. 1 — Lineage of K010362, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (64)

See all 64

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K043474Accu-Chek Aviva SystemRoche Diagnostics Corp.200563

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 27, 2006Not classifiedZ-1377-06OtherThe drying agent beads, located in the cap, may become loose and fall into the vial causing the test strips to potentially give incorrect blood glucose test results.
Feb 10, 2006Not classifiedZ-0571-06OtherThere may be a thin piece of plastic across the mouth of the vial, preventing the cap from sitting snugly onto the vial and, if present, this situation will cause erroneously low or high blood glucose readings.
Feb 10, 2006Not classifiedZ-0572-06OtherThere may be a thin piece of plastic across the mouth of the vial, preventing the cap from sitting snugly onto the vial and, if present, this situation will cause erroneously low or high blood glucose readings.
Feb 10, 2006Not classifiedZ-0573-06OtherThere may be a thin piece of plastic across the mouth of the vial, preventing the cap from sitting snugly onto the vial and, if present, this situation will cause erroneously low or high blood glucose readings..
Feb 10, 2006Not classifiedZ-0566-06OtherThere may be a thin piece of plastic across the mouth of the vial, preventing the cap from sitting snugly onto the vial and, if present, this situation will cause erroneously low or high blood glucose readings.
Feb 12, 2003Not classifiedZ-0715-03OtherCrack in the bottom of the test strip vial will cause erroneously low or high blood glucose readings.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K193483Basal-IQ TechnologyTandem Diabetes Care, Inc. · 2020 · 8 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K043474 (Roche Diagnostics Corp., 2005) has gone furthest, with 63 descendants of its own.

How does its category compare?

All LFR devices

LFR Glucose Dehydrogenase, Glucose: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LFR

1in 2015

2007: 12009: 12011: 62012: 132013: 32014: 32015: 1
2007high 13 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LFR

161in 2014

2011: 296 days median2012: 152 days median2013: 212 days median2014: 161 days median
2007high 296 (2011) · low 152 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LFR

1in 2012

2007: 32008: 12009: 42011: 12012: 1
2007high 4 (2009) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LFR

827in 2025

2007: 23,8192008: 6,5842009: 10,8232010: 4,8652011: 3,8022012: 3,3412013: 3,8212014: 3,9602015: 3,3662016: 2,1452017: 1032018: 1082019: 1082020: 1322021: 1642022: 1012023: 1032024: 1712025: 827
2007high 23,819 (2007) · low 101 (2022)2025

MAUDE reports naming product code LFR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—323,819
20080—16,584
20091—410,823
20100—04,865
2011629613,802
20121315213,341
2013321203,821
2014316103,960
20151—03,366
20160—02,145
20170—0103
20180—0108
20190—0108
20200—0132
20210—0164
20220—0101
20230—0103
20240—0171
20250—0827
20260—0—
Cite this page

510(k) Report. “K010362: Accu-Chek Comfort Curve Test Strip.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K010362

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026