510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

Guider Softip Guiding Catheter XF 5F

K010853510(k) SpecialCleared Apr 18, 200128-day reviewClass II21 CFR 870.1250Recalls 17

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boston Scientific, Target · DQY Catheter, Percutaneous

Cleared on Apr 18, 2001 after a 28-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 4 of the 109 devices in its family, counting itself, have a Class I recall. 108 later devices descend from it. It has been recalled 17 times.

Decided
Apr 18, 2001
Review
28 days
Class
Class II
Product code
DQY
Recalls
17
Predicates
0

108 later devices descend from it, 3 citing it directly. 4 of the 109 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2013: 3 descendants cleared2011: 4 descendants cleared2007: 1 descendant cleared2008: 3 descendants cleared2010: 2 descendants cleared2024: 14 descendants cleared, 1 with a Class I recall12015: 6 descendants cleared2023: 11 descendants cleared2016: 3 descendants cleared2012: 1 descendant cleared2022: 9 descendants cleared, 1 with a Class I recall12020: 10 descendants cleared, 2 with a Class I recall22018: 3 descendants cleared2019: 5 descendants cleared2026: 14 descendants cleared2014: 1 descendant cleared2017: 2 descendants cleared2021: 6 descendants cleared2025: 10 descendants cleared200020052010201520202025
  1. This deviceK010853 Guider Softip Guiding Catheter XF 5FBoston Scientific, Target · 2001
  2. Descendant · cites it directly · 105 of its ownK070970 Neuron Support Catheter System, Model 5F/6FPenumbra, Inc. · 2007 · Recalled
  3. Descendant · 2 generations on · 88 of its ownK082290 Neuron Delivery Catheter 070Penumbra, Inc. · 2008 · Recalled
  4. Descendant · 2 generations on · 86 of its ownK083125 Neuron Select Catheter Size 070Penumbra, Inc. · 2008 · Recalled
  5. Descendant · 2 generations on · 61 of its ownK111380 Neuron Max SystemPenumbra, Inc. · 2011
  6. Descendant · 3 generations on · 27 of its ownK180169 EagleRay Long Sheath, 0.088 ID, 80cm, 90cm, 100cm and 110cm Lengths…Imperative Care, Inc. · 2018

Showing 5 of 108 descendants. See all descendants

Descendants cleared per year

2007: 12008: 32010: 22011: 42012: 12013: 32014: 12015: 62016: 32017: 22018: 32019: 52020: 102021: 62022: 92023: 112024: 142025: 102026: 14
2007high 14 (2024) · low 0 (2009)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 108 · See all
  • K070970Neuron Support Catheter System, Model 5F/6FPenumbra, Inc. · 2007 · cites it directly105descendants
  • K082290Neuron Delivery Catheter 070Penumbra, Inc. · 2008 · 2 generations on88descendants
  • K083125Neuron Select Catheter Size 070Penumbra, Inc. · 2008 · 2 generations on86descendants
  • K111380Neuron Max SystemPenumbra, Inc. · 2011 · 2 generations on61descendants
  • K180169EagleRay Long Sheath, 0.088 ID, 80cm, 90cm, 100cm and 110cm Lengths, EagleRay Access…Imperative Care, Inc. · 2018 · 3 generations on27descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K010853Guider Softip Guiding Catheter XF 5FBoston Scientific, Target2001—
+1K070970Neuron Support Catheter System, Model 5F/6FPenumbra, Inc.20071
+2K082290Neuron Delivery Catheter 070Penumbra, Inc.20081
+2K083125Neuron Select Catheter Size 070Penumbra, Inc.20081
+2K111380Neuron Max SystemPenumbra, Inc.2011—
+3K180169EagleRay Long Sheath, 0.088 ID, 80cm, 90cm, 100cm and 110cm Lengths, EagleRay…Imperative Care, Inc.2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
200830
201020
201140
201210
201330
201410
201560
201630
201720
201830
201950
2020102
202160
202291
2023110
2024141
2025100
2026140
FIG. 1 — Lineage of K010853, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQY device

  1. #1 K234115 APRO 55 Catheterscore 17.5 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (108)

See all 108

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K070970Neuron Support Catheter System, Model 5F/6FPenumbra, Inc.2007105
K132542Transgate Guide CatheterStryker Neurovascular20130
K094048Percupro Messenger Balloon CatheterCardiosolutions, Inc.20110

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 9, 2022Class IIZ-1582-2022Process controlThe impacted products were distributed with an MPXF tip curve shape instead of the 40XF tip curve shape for the Guider/40XF/8FR/90CM guide catheter.
Dec 11, 2017Class IIZ-2735-2018Process change controlPotential polymer material degradation.
Oct 5, 2016Class IIZ-0338-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Oct 5, 2016Class IIZ-0340-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Oct 5, 2016Class IIZ-0341-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Oct 5, 2016Class IIZ-0342-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Oct 5, 2016Class IIZ-0344-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Oct 5, 2016Class IIZ-0337-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Oct 5, 2016Class IIZ-0343-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Oct 5, 2016Class IIZ-0333-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Oct 5, 2016Class IIZ-0335-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Oct 5, 2016Class IIZ-0346-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Oct 5, 2016Class IIZ-0347-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Oct 5, 2016Class IIZ-0334-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Oct 5, 2016Class IIZ-0336-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Oct 5, 2016Class IIZ-0339-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.
Oct 5, 2016Class IIZ-0345-2017Employee errorFirm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K192804CEREBASE DA Guide Sheath, 95cm, CEREBASE DA Guide Sheath, 90cm, CEREBASE DA Guide Sheath, 80cm, CEREBASE DA…Medos International SARL · 2020 · 3 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall
  • K242376Next Generation Access PlatformBalt USA, LLC · 2024 · 4 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall
  • K203043Route 92 Medical 070 Access SystemRoute 92 Medical, Inc. · 2020 · 6 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall
  • K222743Route 92 Medical Full Length 070 Access SystemRoute 92 Medical, Inc. · 2022 · 7 generations on · ◆ in the 2022 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K070970 (Penumbra, Inc., 2007) has gone furthest, with 105 descendants of its own.

How does its category compare?

All DQY devices

DQY Catheter, Percutaneous: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DQY

21in 2025

2007: 162008: 312009: 372010: 222011: 362012: 302013: 322014: 192015: 272016: 222017: 262018: 252019: 302020: 202021: 182022: 182023: 182024: 162025: 212026 (partial year): 11
2007high 37 (2009) · low 16 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQY

115in 2025

2007: 85 days median2008: 66 days median2009: 73 days median2010: 74 days median2011: 71 days median2012: 87 days median2013: 92 days median2014: 127 days median2015: 118 days median2016: 117 days median2017: 113 days median2018: 106 days median2019: 111 days median2020: 120 days median2021: 81 days median2022: 74 days median2023: 114 days median2024: 87 days median2025: 115 days median2026 (partial year): 30 days median
2007high 127 (2014) · low 66 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQY

1in 2025

2007: 62008: 122009: 172010: 322011: 62012: 22013: 36, 1 Class I2014: 32015: 32016: 22, 7 Class I2017: 82018: 22019: 12020: 22021: 22023: 172024: 12025: 1, 1 Class I2026 (partial year): 2, 1 Class I
2007high 36 (2013) · low 0 (2022)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · DQY

3,056in 2025

2007: 3132008: 4962009: 8192010: 1,2932011: 1,2112012: 1,1912013: 1,3522014: 1,4972015: 2,3372016: 2,2522017: 2,4892018: 2,8552019: 2,9022020: 2,8892021: 3,0062022: 2,6432023: 3,0492024: 3,1902025: 3,056
2007high 3,190 (2024) · low 313 (2007)2025

MAUDE reports naming product code DQY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200716856313
2008316612496
2009377317819
20102274321,293
2011367161,211
2012308721,191
2013329236 (1)1,352
20141912731,497
20152711832,337
20162211722 (7)2,252
20172611382,489
20182510622,855
20193011112,902
20202012022,889
2021188123,006
2022187402,643
202318114173,049
2024168713,190
2025211151 (1)3,056
202611302 (1)—
Cite this page

510(k) Report. “K010853: Guider Softip Guiding Catheter XF 5F.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K010853

Report an extraction error