510(k) · Cleared
Recalled · Class I · 2024CEREBASE DA Guide Sheath, 95cm, CEREBASE DA Guide Sheath, 90cm, CEREBASE DA Guide Sheath, 80cm, CEREBASE DA Guide Sheath, 70cm
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Medos International SARL · QJP Catheter, Percutaneous, Neurovasculature
Cleared on Feb 27, 2020 as substantially equivalent to K111380 (Penumbra, Inc., 2011) and 2 other predicates after a 149-day review. Its family tree reaches back 6 generations to 2001. 1 of the 13 devices in its family, counting itself, has a Class I recall. It has been recalled once, including a Class I recall.
- Decided
- Feb 27, 2020
- Review
- 149 days
- Class
- Class II
- Product code
- QJP
- Recalls
- 1 (1 Class I)
- Predicates
- 3
Its longest line of descent runs back 6 generations to K010954 (2001). 12 ancestors in all, the oldest K010853 (2001). No later device cites it as a predicate yet. 1 of the 13 devices in this family, counting this one, has a Class I recall.
- Root · 6 generations backK010954 Concentric Balloon Guide Catheter, Model 90031Concentric Medical, Inc. · 2001 · Recalled
- 5 generations backK021899 Concentric Balloon Guide Catheter, Model 90072Concentric Medical, Inc. · 2002 · Recalled
- 4 generations backK053491 Penumbra Balloon Guide Catheter, Model 7.5FPenumbra, Inc. · 2006
- 3 generations backK070970 Neuron Support Catheter System, Model 5F/6FPenumbra, Inc. · 2007 · Recalled
- 2 generations backK082290 Neuron Delivery Catheter 070Penumbra, Inc. · 2008 · Recalled
- 1 generation backK111380 Neuron Max SystemPenumbra, Inc. · 2011
- This deviceK192804 CEREBASE DA Guide Sheath, 95cm, CEREBASE DA Guide Sheath, 90cm…Medos International SARL · 2020
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K010954 | Concentric Balloon Guide Catheter, Model 90031 | Concentric Medical, Inc. | 2001 | 1 |
| −5 | K021899 | Concentric Balloon Guide Catheter, Model 90072 | Concentric Medical, Inc. | 2002 | 1 |
| −4 | K053491 | Penumbra Balloon Guide Catheter, Model 7.5F | Penumbra, Inc. | 2006 | — |
| −3 | K070970 | Neuron Support Catheter System, Model 5F/6F | Penumbra, Inc. | 2007 | 1 |
| −2 | K082290 | Neuron Delivery Catheter 070 | Penumbra, Inc. | 2008 | 1 |
| −1 | K111380 | Neuron Max System | Penumbra, Inc. | 2011 | — |
| 0 (this device) | K192804 | CEREBASE DA Guide Sheath, 95cm, CEREBASE DA Guide Sheath, 90cm, CEREBASE DA… | Medos International SARL | 2020 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2001 | 1 | 0 |
| 2008 | 1 | 0 |
| 2009 | 1 | 0 |
| 2012 | 1 | 0 |
| 2014 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new QJP device
- #1 K212838 Benchmark Intracranial Access Systemscore 16.9 · 6 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K111380 | Neuron Max SystemSummary page 4…Date Cleared Name Manufacturer Number K111380 07/19/2011 Neuron MAX System Penumbra, Inc. K140080 04/2… | 2011 |
| * | K140080 | Envoy Distal Access (DA) Guiding Catheter, 6F, 95CM & 105CM, Straight & XB, MPD & XB…Summary page 4…Number K111380 07/19/2011 Neuron MAX System Penumbra, Inc. K140080 04/24/2014 ENVOY DA Guiding Catheter Medos International… | 2014 |
| * | K140307 | Envoy Guiding CatheterSummary page 4…Medos International SARL K140307 04/21/2014 ENVOY Guiding Catheter Codman & Shurtleff,… | 2014 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 2, 2024 | Class I | Z-1284-2024 | Under Investigation by firm | Medos has received an increase in complaints for CEREBASE DA Guide Sheath with reports of fractures at the distal end, which may result in surgical procedural delay, vascular injury and in extreme rare occasions it may result in embolism. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 19 years and passes through 3 companies; a new generation was cleared every 2.7 years on median.
- 6 of the 12 ancestors have a recall of any class.
- 9 other devices also cite its predicate K111380.
How does its category compare?
All QJP devicesQJP Catheter, Percutaneous, Neurovasculature: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.
FDA decisions per year · QJP
22so far in 2026, the most since 2007
2026 to Oct 25, outlined.
Median 510(k) review, days · QJP
116in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · QJP
1in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · QJP
688in 2025
MAUDE reports naming product code QJP, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 0 | — | 0 | 0 |
| 2012 | 0 | — | 0 | 0 |
| 2013 | 0 | — | 0 | 0 |
| 2014 | 0 | — | 0 | 0 |
| 2015 | 0 | — | 0 | 0 |
| 2016 | 0 | — | 0 | 0 |
| 2017 | 0 | — | 0 | 0 |
| 2018 | 0 | — | 0 | 0 |
| 2019 | 0 | — | 0 | 0 |
| 2020 | 9 | 72 | 0 | 16 |
| 2021 | 12 | 139 | 0 | 29 |
| 2022 | 14 | 206 | 0 | 30 |
| 2023 | 14 | 184 | 0 | 75 |
| 2024 | 18 | 188 | 1 | 130 |
| 2025 | 14 | 116 | 0 | 688 |
| 2026 | 22 | 141 | 0 | — |
What is it indicated for?
The CEREBASE DA Guide Sheath is indicated for the introduction of interventional devices into the neuro vasculature.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K192804 · printed Nov 6, 2026
Cite this page
510(k) Report. “K192804: CEREBASE DA Guide Sheath, 95cm, CEREBASE DA Guide Sheath, 90cm, CEREBASE DA Guide Sheath, 80cm, CEREBASE DA Guide Sheath, 70cm.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K192804
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026