Boston Scientific Corp · DQY Catheter, Percutaneous
Cleared on Jun 21, 2001 after a 90-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 160 later devices descend from it. It has been recalled 11 times.
- Decided
- Jun 21, 2001
- Review
- 90 days
- Class
- Class II
- Product code
- DQY
- Recalls
- 11
- Predicates
- 0
160 later devices descend from it, 3 citing it directly.
- This deviceK010874 6F Mach 1, Model 34356-XXXBoston Scientific Corp · 2001
- Descendant · cites it directly · 132 of its ownK020028 6F MACH1/34356-XXX; 5F MACH1/35845-XXX; 7F MACH1/34357-XXX; 8F…Boston Scientific · 2002 · Recalled
- Descendant · cites it directly · 109 of its ownK061601 Asahi Precious Guide CatheterAsahi Intecc Co., Ltd. · 2006
- Descendant · 2 generations on · 105 of its ownK093268 Intellivue Patient MonitorPhilips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2010 · Recalled
- Descendant · 3 generations on · 103 of its ownK102562 Intellivue Patient Monitor, Models MP2, X2, MP5, MP5T, MP20, MP30…Philips Medizin Systeme Boblingen GmbH · 2010
- Descendant · 3 generations on · 90 of its ownK110622 Intellivue Patient MonitorPhilips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2011 · Recalled
Showing 5 of 160 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 160 · See all- K0200286F MACH1/34356-XXX; 5F MACH1/35845-XXX; 7F MACH1/34357-XXX; 8F MACH1/34358-XXXBoston Scientific · 2002 · cites it directly132descendants
- K061601Asahi Precious Guide CatheterAsahi Intecc Co., Ltd. · 2006 · cites it directly109descendants
- K093268Intellivue Patient MonitorPhilips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2010 · 2 generations on105descendants
- K102562Intellivue Patient Monitor, Models MP2, X2, MP5, MP5T, MP20, MP30, MP40, MP50, MP60…Philips Medizin Systeme Boblingen GmbH · 2010 · 3 generations on103descendants
- K110622Intellivue Patient MonitorPhilips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2011 · 3 generations on90descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K010874 | 6F Mach 1, Model 34356-XXX | Boston Scientific Corp | 2001 | — |
| +1 | K020028 | 6F MACH1/34356-XXX; 5F MACH1/35845-XXX; 7F MACH1/34357-XXX; 8F MACH1/34358-XXX | Boston Scientific | 2002 | 11 |
| +1 | K061601 | Asahi Precious Guide Catheter | Asahi Intecc Co., Ltd. | 2006 | — |
| +2 | K093268 | Intellivue Patient Monitor | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2010 | 2 |
| +3 | K102562 | Intellivue Patient Monitor, Models MP2, X2, MP5, MP5T, MP20, MP30, MP40, MP50… | Philips Medizin Systeme Boblingen GmbH | 2010 | — |
| +3 | K110622 | Intellivue Patient Monitor | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2011 | 2 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2002 | 1 | 0 |
| 2003 | 1 | 0 |
| 2004 | 1 | 0 |
| 2006 | 1 | 0 |
| 2007 | 2 | 0 |
| 2009 | 2 | 0 |
| 2010 | 5 | 0 |
| 2011 | 6 | 0 |
| 2012 | 9 | 0 |
| 2013 | 6 | 0 |
| 2014 | 8 | 0 |
| 2015 | 8 | 0 |
| 2016 | 11 | 0 |
| 2017 | 13 | 0 |
| 2018 | 10 | 0 |
| 2019 | 7 | 0 |
| 2020 | 8 | 0 |
| 2021 | 5 | 0 |
| 2022 | 12 | 0 |
| 2023 | 14 | 0 |
| 2024 | 8 | 0 |
| 2025 | 14 | 0 |
| 2026 | 8 | 0 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DQY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DQY device
- #1 K234115 APRO 55 Catheterscore 17.5 · 6 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it? (160)
See all 1603 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 7, 2008 | Not classified | Z-1301-2008 | Labeling mix-ups | Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length. |
| Feb 7, 2008 | Not classified | Z-1305-2008 | Labeling mix-ups | Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length. |
| Feb 7, 2008 | Not classified | Z-1304-2008 | Labeling mix-ups | Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length. |
| Feb 7, 2008 | Not classified | Z-1303-2008 | Labeling mix-ups | Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length. |
| Feb 7, 2008 | Not classified | Z-1300-2008 | Labeling mix-ups | Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length. |
| Feb 7, 2008 | Not classified | Z-1302-2008 | Labeling mix-ups | Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length. |
| Dec 13, 2006 | Not classified | Z-0383-2007 | Other | Excess strands of resin may exist in the lumen near the hub of certain batches/lots of the Mach 1 Guide Catheters. If this excess resin is present in the catheter, there is the potential of an embolization resulting from the excess resin… |
| Dec 13, 2006 | Not classified | Z-0382-2007 | Other | Excess strands of resin may exist in the lumen near the hub of certain batches/lots of the Mach 1 Guide Catheters. If this excess resin is present in the catheter, there is the potential of an embolization resulting from the excess resin… |
| Dec 13, 2006 | Not classified | Z-0388-2007 | Other | Excess strands of resin may exist in the lumen near the hub of certain batches/lots of the Mach 1 Guide Catheters. If this excess resin is present in the catheter, there is the potential of an embolization resulting from the excess resin… |
| May 24, 2004 | Not classified | Z-0432-06 | Other | Some Mach 1 6F Coronary Guide Catheter could be mislabeled in that a few units labeled as Mach 1 6F Kimny Curve style, could contain a Mach 1 6F JR 3.5 curve style and a few units labeled as Mach 1 6F JR 3.5 curve style, could contain a… |
| May 24, 2004 | Not classified | Z-0433-06 | Other | Some Mach 1 6F Coronary Guide Catheter could be mislabeled in that a few units labeled as Mach 1 6F Kimny Curve style, could contain a Mach 1 6F JR 3.5 curve style and a few units labeled as Mach 1 6F JR 3.5 curve style, could contain a… |
The 5 latest of 11 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K020028 (Boston Scientific, 2002) has gone furthest, with 132 descendants of its own.
How does its category compare?
All DQY devicesDQY Catheter, Percutaneous: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DQY
21in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DQY
115in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DQY
1in 2025
Class I (10 since 2007) Class II/III not classified
Adverse event reports per year · DQY
3,056in 2025
MAUDE reports naming product code DQY, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 16 | 85 | 6 | 313 |
| 2008 | 31 | 66 | 12 | 496 |
| 2009 | 37 | 73 | 17 | 819 |
| 2010 | 22 | 74 | 32 | 1,293 |
| 2011 | 36 | 71 | 6 | 1,211 |
| 2012 | 30 | 87 | 2 | 1,191 |
| 2013 | 32 | 92 | 36 (1) | 1,352 |
| 2014 | 19 | 127 | 3 | 1,497 |
| 2015 | 27 | 118 | 3 | 2,337 |
| 2016 | 22 | 117 | 22 (7) | 2,252 |
| 2017 | 26 | 113 | 8 | 2,489 |
| 2018 | 25 | 106 | 2 | 2,855 |
| 2019 | 30 | 111 | 1 | 2,902 |
| 2020 | 20 | 120 | 2 | 2,889 |
| 2021 | 18 | 81 | 2 | 3,006 |
| 2022 | 18 | 74 | 0 | 2,643 |
| 2023 | 18 | 114 | 17 | 3,049 |
| 2024 | 16 | 87 | 1 | 3,190 |
| 2025 | 21 | 115 | 1 (1) | 3,056 |
| 2026 | 11 | 30 | 2 (1) | — |
https://510kreport.com/device/K010874 · printed Nov 6, 2026