510(k) · Cleared
Recalled · 20086F MACH1/34356-XXX; 5F MACH1/35845-XXX; 7F MACH1/34357-XXX; 8F MACH1/34358-XXX
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Boston Scientific · DQY Catheter, Percutaneous
Cleared on Jan 22, 2002 as substantially equivalent to K010874 (Boston Scientific Corp, 2001) after an 18-day review. Its family tree reaches back 1 generation to 2001. 132 later devices descend from it. It has been recalled 11 times.
- Decided
- Jan 22, 2002
- Review
- 18 days
- Class
- Class II
- Product code
- DQY
- Recalls
- 11
- Predicates
- 1
Its longest line of descent runs back 1 generation to K010874 (2001). 132 later devices descend from it, 3 citing it directly.
- Root · 1 generation backK010874 6F Mach 1, Model 34356-XXXBoston Scientific Corp · 2001 · Recalled
- This deviceK020028 6F MACH1/34356-XXX; 5F MACH1/35845-XXX; 7F MACH1/34357-XXX; 8F…Boston Scientific · 2002
- Descendant · cites it directly · 109 of its ownK061601 Asahi Precious Guide CatheterAsahi Intecc Co., Ltd. · 2006
- Descendant · 2 generations on · 105 of its ownK093268 Intellivue Patient MonitorPhilips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2010 · Recalled
- Descendant · 3 generations on · 103 of its ownK102562 Intellivue Patient Monitor, Models MP2, X2, MP5, MP5T, MP20, MP30…Philips Medizin Systeme Boblingen GmbH · 2010
- Descendant · 3 generations on · 90 of its ownK110622 Intellivue Patient MonitorPhilips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2011 · Recalled
- Descendant · 4 generations on · 84 of its ownK110474 Philips Intellivue TCG10Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2011 · Recalled
Showing 5 of 132 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 132 · See all- K061601Asahi Precious Guide CatheterAsahi Intecc Co., Ltd. · 2006 · cites it directly109descendants
- K093268Intellivue Patient MonitorPhilips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2010 · 2 generations on105descendants
- K102562Intellivue Patient Monitor, Models MP2, X2, MP5, MP5T, MP20, MP30, MP40, MP50, MP60…Philips Medizin Systeme Boblingen GmbH · 2010 · 3 generations on103descendants
- K110622Intellivue Patient MonitorPhilips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2011 · 3 generations on90descendants
- K110474Philips Intellivue TCG10Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2011 · 4 generations on84descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K010874 | 6F Mach 1, Model 34356-XXX | Boston Scientific Corp | 2001 | 11 |
| 0 (this device) | K020028 | 6F MACH1/34356-XXX; 5F MACH1/35845-XXX; 7F MACH1/34357-XXX; 8F MACH1/34358-XXX | Boston Scientific | 2002 | — |
| +1 | K061601 | Asahi Precious Guide Catheter | Asahi Intecc Co., Ltd. | 2006 | — |
| +2 | K093268 | Intellivue Patient Monitor | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2010 | 2 |
| +3 | K102562 | Intellivue Patient Monitor, Models MP2, X2, MP5, MP5T, MP20, MP30, MP40, MP50… | Philips Medizin Systeme Boblingen GmbH | 2010 | — |
| +3 | K110622 | Intellivue Patient Monitor | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2011 | 2 |
| +4 | K110474 | Philips Intellivue TCG10 | Philips Medizinsysteme Boeblingen GmbH, Cardiac AN | 2011 | 1 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2006 | 1 | 0 |
| 2007 | 2 | 0 |
| 2009 | 1 | 0 |
| 2010 | 4 | 0 |
| 2011 | 4 | 0 |
| 2012 | 4 | 0 |
| 2013 | 6 | 0 |
| 2014 | 6 | 0 |
| 2015 | 6 | 0 |
| 2016 | 7 | 0 |
| 2017 | 11 | 0 |
| 2018 | 10 | 0 |
| 2019 | 6 | 0 |
| 2020 | 6 | 0 |
| 2021 | 5 | 0 |
| 2022 | 12 | 0 |
| 2023 | 12 | 0 |
| 2024 | 8 | 0 |
| 2025 | 14 | 0 |
| 2026 | 7 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DQY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DQY device
- #1 K234115 APRO 55 Catheterscore 17.5 · 6 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K010874 | 6F Mach 1, Model 34356-XXXSummary page 2…tended for use in the cerebral vasculature. 6.5 Summary of Technological Characteristics Same as currently marketed 6F Mach1 guide catheters, cleared for market under K010874, June 21, 2001. 6.6 Non-clinical Test Summary Functional testing consisted of shaft and distal segment tensile and elongation,… | 2001 |
What descends from it? (132)
See all 1323 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 7, 2008 | Not classified | Z-1301-2008 | Labeling mix-ups | Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length. |
| Feb 7, 2008 | Not classified | Z-1305-2008 | Labeling mix-ups | Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length. |
| Feb 7, 2008 | Not classified | Z-1304-2008 | Labeling mix-ups | Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length. |
| Feb 7, 2008 | Not classified | Z-1303-2008 | Labeling mix-ups | Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length. |
| Feb 7, 2008 | Not classified | Z-1300-2008 | Labeling mix-ups | Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length. |
| Feb 7, 2008 | Not classified | Z-1302-2008 | Labeling mix-ups | Mislabeled: Seven lots/ batches of Boston Scientific Mach 1 Guide Catheter may be mislabeled with the incorrect device length. |
| Sep 11, 2007 | Not classified | Z-0104-2008 | Process control | Missing holes: Catheters are missing the catheter's side holes. |
| Dec 13, 2006 | Not classified | Z-0384-2007 | Other | Excess strands of resin may exist in the lumen near the hub of certain batches/lots of the Mach 1 Guide Catheters. If this excess resin is present in the catheter, there is the potential of an embolization resulting from the excess resin… |
| Dec 13, 2006 | Not classified | Z-0386-2007 | Other | Excess strands of resin may exist in the lumen near the hub of certain batches/lots of the Mach 1 Guide Catheters. If this excess resin is present in the catheter, there is the potential of an embolization resulting from the excess resin… |
| Dec 13, 2006 | Not classified | Z-0385-2007 | Other | Excess strands of resin may exist in the lumen near the hub of certain batches/lots of the Mach 1 Guide Catheters. If this excess resin is present in the catheter, there is the potential of an embolization resulting from the excess resin… |
| Dec 13, 2006 | Not classified | Z-0387-2007 | Other | Excess strands of resin may exist in the lumen near the hub of certain batches/lots of the Mach 1 Guide Catheters. If this excess resin is present in the catheter, there is the potential of an embolization resulting from the excess resin… |
The 5 latest of 11 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All DQY devicesDQY Catheter, Percutaneous: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DQY
21in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DQY
115in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DQY
1in 2025
Class I (10 since 2007) Class II/III not classified
Adverse event reports per year · DQY
3,056in 2025
MAUDE reports naming product code DQY, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 16 | 85 | 6 | 313 |
| 2008 | 31 | 66 | 12 | 496 |
| 2009 | 37 | 73 | 17 | 819 |
| 2010 | 22 | 74 | 32 | 1,293 |
| 2011 | 36 | 71 | 6 | 1,211 |
| 2012 | 30 | 87 | 2 | 1,191 |
| 2013 | 32 | 92 | 36 (1) | 1,352 |
| 2014 | 19 | 127 | 3 | 1,497 |
| 2015 | 27 | 118 | 3 | 2,337 |
| 2016 | 22 | 117 | 22 (7) | 2,252 |
| 2017 | 26 | 113 | 8 | 2,489 |
| 2018 | 25 | 106 | 2 | 2,855 |
| 2019 | 30 | 111 | 1 | 2,902 |
| 2020 | 20 | 120 | 2 | 2,889 |
| 2021 | 18 | 81 | 2 | 3,006 |
| 2022 | 18 | 74 | 0 | 2,643 |
| 2023 | 18 | 114 | 17 | 3,049 |
| 2024 | 16 | 87 | 1 | 3,190 |
| 2025 | 21 | 115 | 1 (1) | 3,056 |
| 2026 | 11 | 30 | 2 (1) | — |
What is it indicated for?
Mach1 Guide Catheters are intended for use in general intravascular and coronary applications. They provide a pathway through which medical instruments, such as balloon dilatation catheters, guide wires or other therapeutic devices may be introduced. These devices are not intended for use in the cerebral vasculature.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K020028 · printed Nov 6, 2026
Cite this page
510(k) Report. “K020028: 6F MACH1/34356-XXX; 5F MACH1/35845-XXX; 7F MACH1/34357-XXX; 8F MACH1/34358-XXX.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K020028