510(k) Report

510(k) · Cleared

Recalled · 2008

Dilating Tip Obturator

K012884510(k) TraditionalCleared Jan 17, 2002142-day reviewClass II21 CFR 876.1500Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Applied Medical Resources Corp. · GCJ Laparoscope, General & Plastic Surgery

Cleared on Jan 17, 2002 after a 142-day review. The FDA summary names its predicate without a number. 14 later devices descend from it. It has been recalled 7 times.

Decided
Jan 17, 2002
Review
142 days
Class
Class II
Product code
GCJ
Recalls
7
Predicates
0

14 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2007: 2 descendants cleared2011: 1 descendant cleared2003: 1 descendant cleared2006: 1 descendant cleared2004: 1 descendant cleared2021: 2 descendants cleared2008: 1 descendant cleared2017: 1 descendant cleared2020: 1 descendant cleared2023: 1 descendant cleared2026: 1 descendant cleared2024: 1 descendant cleared200020052010201520202025
  1. This deviceK012884 Dilating Tip ObturatorApplied Medical Resources Corp. · 2002
  2. Descendant · cites it directly · 13 of its ownK032889 Optical SeparatorApplied Medical Resources · 2003
  3. Descendant · 2 generations on · 10 of its ownK060096 Modular Trocar SystemApplied Medical Resources Corp. · 2006 · Recalled
  4. Descendant · 3 generations on · 4 of its ownK083638 Fixation TrocarsApplied Medical Resources Corp. · 2008 · Recalled
  5. Descendant · 4 generations on · 2 of its ownK211577 SoftFix Balloon TrocarUnimax Medical Systems, Inc. · 2021
  6. Descendant · 3 generations on · 2 of its ownK072674 Kii Trocar SystemApplied Medical Resources Corp. · 2007 · Recalled

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

2003: 12004: 12006: 12007: 22008: 12011: 12017: 12020: 12021: 22023: 12024: 12026: 1
2003high 2 (2007) · low 0 (2005)2026

Most influential descendants

Showing 5 of 14 · See all
  • K032889Optical SeparatorApplied Medical Resources · 2003 · cites it directly13descendants
  • K060096Modular Trocar SystemApplied Medical Resources Corp. · 2006 · 2 generations on10descendants
  • K083638Fixation TrocarsApplied Medical Resources Corp. · 2008 · 3 generations on4descendants
  • K211577SoftFix Balloon TrocarUnimax Medical Systems, Inc. · 2021 · 4 generations on2descendants
  • K072674Kii Trocar SystemApplied Medical Resources Corp. · 2007 · 3 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K012884Dilating Tip ObturatorApplied Medical Resources Corp.2002—
+1K032889Optical SeparatorApplied Medical Resources2003—
+2K060096Modular Trocar SystemApplied Medical Resources Corp.20067
+3K083638Fixation TrocarsApplied Medical Resources Corp.20081
+4K211577SoftFix Balloon TrocarUnimax Medical Systems, Inc.2021—
+3K072674Kii Trocar SystemApplied Medical Resources Corp.20072

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200410
200610
200720
200810
201110
201710
202010
202120
202310
202410
202610
FIG. 1 — Lineage of K012884, 2002–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GCJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GCJ device

  1. #1 K214046 780 nm 1688 4K Camera System, 780 nm L11 LED…score 21.4 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—EndoPath Optiview Optical Surgical Obturator
Named without a submission number
—

What descends from it? (14)

See all 14

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K032889Optical SeparatorApplied Medical Resources200313
K111002Reprocessed Endoscopic TrocarsSterilmed, Inc.20110
K063788Reprocessed Endoscopic Trocars and CannulasAscent Healthcare Solutions20070

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 25, 2008Not classifiedZ-0023-2009Process controlPotential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can stretch into a configuration that blocks the flow of insufflation gas (C02).
Jun 25, 2008Not classifiedZ-0025-2009Process controlPotential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can stretch into a configuration that blocks the flow of insufflation gas (C02).
Jun 25, 2008Not classifiedZ-0026-2009Process controlPotential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can stretch into a configuration that blocks the flow of insufflation gas (C02).
Jun 25, 2008Not classifiedZ-0027-2009Process controlPotential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can stretch into a configuration that blocks the flow of insufflation gas (C02).
Jun 25, 2008Not classifiedZ-0028-2009Process controlPotential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can stretch into a configuration that blocks the flow of insufflation gas (C02).
Jun 25, 2008Not classifiedZ-0029-2009Process controlPotential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can stretch into a configuration that blocks the flow of insufflation gas (C02).
Jun 25, 2008Not classifiedZ-0024-2009Process controlPotential inability to insufflate through the stopcock. Under certain circumstances, a flexible elastomeric component inside the 15mm trocar can stretch into a configuration that blocks the flow of insufflation gas (C02).

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K032889 (Applied Medical Resources, 2003) has gone furthest, with 13 descendants of its own.

How does its category compare?

All GCJ devices

GCJ Laparoscope, General & Plastic Surgery: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GCJ

21in 2025

2007: 182008: 192009: 212010: 242011: 152012: 172013: 152014: 192015: 262016: 272017: 202018: 212019: 152020: 282021: 182022: 202023: 312024: 292025: 212026 (partial year): 26
2007high 31 (2023) · low 15 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GCJ

108in 2025

2007: 42 days median2008: 73 days median2009: 85 days median2010: 122 days median2011: 113 days median2012: 88 days median2013: 93 days median2014: 90 days median2015: 76 days median2016: 99 days median2017: 77 days median2018: 88 days median2019: 60 days median2020: 61 days median2021: 116 days median2022: 103 days median2023: 109 days median2024: 90 days median2025: 108 days median2026 (partial year): 122 days median
2007high 122 (2010) · low 42 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GCJ

5in 2025

2007: 122008: 162009: 62010: 302011: 92012: 12013: 7, 1 Class I2014: 22016: 52017: 12018: 12022: 12023: 12025: 52026 (partial year): 1
2007high 30 (2010) · low 0 (2015)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · GCJ

5,860in 2025

2007: 1,6392008: 1,3192009: 1,9952010: 2,6572011: 2,1332012: 1,5262013: 1,5652014: 1,7142015: 1,7102016: 2,0802017: 3,4682018: 2,5622019: 2,5782020: 2,6182021: 3,4312022: 3,9452023: 6,0022024: 6,0472025: 5,860
2007high 6,047 (2024) · low 1,319 (2008)2025

MAUDE reports naming product code GCJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071842121,639
20081973161,319
2009218561,995
201024122302,657
20111511392,133
2012178811,526
201315937 (1)1,565
2014199021,714
2015267601,710
2016279952,080
2017207713,468
2018218812,562
2019156002,578
2020286102,618
20211811603,431
20222010313,945
20233110916,002
2024299006,047
20252110855,860
2026261221—

What is it indicated for?

The Dilating Tip Obturator is indicated for use in general, abdominal, gynecological and thoracic minimally invasive surgical procedures to establish a path of entry or to gain access through tissue planes and/or potential spaces for endoscopic instruments.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K012884: Dilating Tip Obturator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K012884

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.