510(k) · Cleared
Recalled · Class II · 2015Reprocessed Endoscopic Trocars and Cannulas
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ascent Healthcare Solutions · NLM Laparoscope, General & Plastic Surgery, Reprocessed
Cleared on Oct 15, 2007 as substantially equivalent to K041795 (Applied Medical Resources Corp., 2004) and 3 other predicates after a 298-day review. Its family tree reaches back 3 generations to 1999. It has been recalled 3 times.
- Decided
- Oct 15, 2007
- Review
- 298 days
- Class
- Class II
- Product code
- NLM
- Recalls
- 3
- Predicates
- 4
Its longest line of descent runs back 3 generations to K990028 (1999). 7 ancestors in all, the oldest K932995 (1993). No later device cites it as a predicate yet.
- Root · 3 generations backK990028 Endopath Optiview Optical Surgical Obturator and SleeveEthicon Endo-Surgery, Inc. · 1999
- 2 generations backK032889 Optical SeparatorApplied Medical Resources · 2003
- 1 generation backK041795 Modification to Optical Separator SystemApplied Medical Resources Corp. · 2004 · Recalled
- This deviceK063788 Reprocessed Endoscopic Trocars and CannulasAscent Healthcare Solutions · 2007
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K990028 | Endopath Optiview Optical Surgical Obturator and Sleeve | Ethicon Endo-Surgery, Inc. | 1999 | — |
| −2 | K032889 | Optical Separator | Applied Medical Resources | 2003 | — |
| −1 | K041795 | Modification to Optical Separator System | Applied Medical Resources Corp. | 2004 | 1 |
| 0 (this device) | K063788 | Reprocessed Endoscopic Trocars and Cannulas | Ascent Healthcare Solutions | 2007 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1993 | 1 | 0 |
| 1996 | 1 | 0 |
| 2001 | 1 | 0 |
| 2002 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in NLM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new NLM device
- #1 K201511 Reprocessed Endoscopic Trocars and Sleevesscore -1.9 · 2 generations · no Class I recall
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K041795 | Modification to Optical Separator SystemSummary page 1…lassification Name: Laparoscope, general & plastic surgery, reprocessed Predicate Device 510(k) Title Manufacturer K041795 Modification to Optical Separator Applied Medical K032889 Optical Separator A… | 2004 |
| * | K032889 | Optical SeparatorSummary page 1…reprocessed Predicate Device 510(k) Title Manufacturer K041795 Modification to Optical Separator Applied Medical K032889 Optical Separator Applied Medical K012884 Dilating Tip Obturator A… | 2003 |
| * | K012884 | Dilating Tip ObturatorSummary page 1…Manufacturer K041795 Modification to Optical Separator Applied Medical K032889 Optical Separator Applied Medical K012884 Dilating Tip Obturator Applied Medical K012968 Trocar Seal A… | 2002 |
| * | K012968 | Applied Access SealSummary page 1…Applied Medical K032889 Optical Separator Applied Medical K012884 Dilating Tip Obturator Applied Medical K012968 Trocar Seal Applied Medical Device description: Reprocessed Endoscopic Trocars and C… | 2001 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 1, 2015 | Class II | Z-1836-2015 | Under Investigation by firm | Stryker Sustainability Solutions is recalling Trocars because they have received reports indicating that some may fracture at the distal tip resulting in fragments at the surgical site. |
| Jun 1, 2015 | Class II | Z-1835-2015 | Under Investigation by firm | Stryker Sustainability Solutions is recalling Trocars because they have received reports indicating that some may fracture at the distal tip resulting in fragments at the surgical site. |
| Jun 1, 2015 | Class II | Z-1834-2015 | Under Investigation by firm | Stryker Sustainability Solutions is recalling Trocars because they have received reports indicating that some may fracture at the distal tip resulting in fragments at the surgical site. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 8 years and passes through 4 companies; a new generation was cleared every 3.1 years on median.
- 2 of the 7 ancestors have a recall of any class.
- 1 other device also cites its predicate K041795.
How does its category compare?
All NLM devicesNLM Laparoscope, General & Plastic Surgery, Reprocessed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.
FDA decisions per year · NLM
1in 2020
2026 to Oct 25, outlined.
Median 510(k) review, days · NLM
284in 2007
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · NLM
6so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · NLM
10in 2025
MAUDE reports naming product code NLM, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 4 | 284 | 0 | 2 |
| 2008 | 0 | — | 1 | 1 |
| 2009 | 0 | — | 0 | 2 |
| 2010 | 1 | — | 1 | 11 |
| 2011 | 1 | — | 0 | 7 |
| 2012 | 1 | — | 0 | 82 |
| 2013 | 1 | — | 0 | 102 |
| 2014 | 1 | — | 0 | 84 |
| 2015 | 0 | — | 0 | 27 |
| 2016 | 2 | — | 1 | 39 |
| 2017 | 2 | — | 0 | 15 |
| 2018 | 0 | — | 1 | 9 |
| 2019 | 1 | — | 0 | 6 |
| 2020 | 1 | — | 0 | 1 |
| 2021 | 0 | — | 0 | 5 |
| 2022 | 0 | — | 0 | 5 |
| 2023 | 0 | — | 0 | 5 |
| 2024 | 0 | — | 0 | 4 |
| 2025 | 0 | — | 0 | 10 |
| 2026 | 0 | — | 6 | — |
What is it indicated for?
Reprocessed Endoscopic Trocars are indicated for use to establish a port of entry for endoscopic instruments in patients requiring minimally invasive surgical procedures.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K063788 · printed Nov 5, 2026
Cite this page
510(k) Report. “K063788: Reprocessed Endoscopic Trocars and Cannulas.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K063788
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026