510(k) Report

510(k) · Cleared

Recalled · 2004

Roche Cobas Taqman Analyzer

K012966510(k) TraditionalCleared Feb 1, 2002150-day reviewClass I21 CFR 862.2170Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Molecular Systems, Inc. · JJF Analyzer, Chemistry, Micro, For Clinical Use

Cleared on Feb 1, 2002 as substantially equivalent to K964506 (Roche Molecular Systems, Inc., 1997) after a 150-day review. Its family tree reaches back 2 generations to 1993. 14 later devices descend from it. It has been recalled 3 times.

Decided
Feb 1, 2002
Review
150 days
Class
Class I
Product code
JJF
Recalls
3
Predicates
1

Its longest line of descent runs back 2 generations to K922906 (1993). 4 ancestors in all. 14 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor1994: 1 other ancestor2003: 1 descendant cleared2010: 2 descendants cleared2005: 1 descendant cleared2014: 2 descendants cleared2007: 2 descendants cleared2012: 1 descendant cleared2013: 2 descendants cleared2011: 2 descendants cleared2019: 1 descendant clearedK922906 · Amplicor Chlamydia Trachomatis Test · Roche Molecular Systems, Inc. · 1993K964506 · Roche Cobas Amplicor Analyzer · Roche Molecular Systems, Inc. · 1997 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors 1 generation forward: 2 descendants2 generations forward: 5 descendants3 generations forward: 6 descendants4 generations forward: 1 descendant K964506 · Roche Cobas Amplicor Analyzer · Roche Molecular Systems, Inc. · 1997 · recalledK922906 · Amplicor Chlamydia Trachomatis Test · Roche Molecular Systems, Inc. · 19930
  1. Root · 2 generations backK922906 Amplicor Chlamydia Trachomatis TestRoche Molecular Systems, Inc. · 1993
  2. 1 generation backK964506 Roche Cobas Amplicor AnalyzerRoche Molecular Systems, Inc. · 1997 · Recalled
  3. This deviceK012966 Roche Cobas Taqman AnalyzerRoche Molecular Systems, Inc. · 2002
  4. Descendant · cites it directly · 10 of its ownK033734 Lightcycler Instrument Version 1.2Roche Diagnostics Corp. · 2003 · Recalled
  5. Descendant · 2 generations on · 4 of its ownK051713 Jbaids Anthrac Detection SystemIdaho Technology, Inc. · 2005
  6. Descendant · 2 generations on · 3 of its ownK070597 Verigene System, F2 Nucleic Acid Test, F5 Nucleic Acid Test, and…Nanosphere, Inc. · 2007
  7. Descendant · 3 generations on · 3 of its ownK071188 Modification to Jbaids Anthrax Detection SystemIdaho Technology, Inc. · 2007
  8. Descendant · cites it directly · 2 of its ownK092705 Abbott M2000SP and Abbott M2000RT, Models 9K14-01 (G-Series)…Abbott Molecular, Inc. · 2010 · Recalled

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

2003: 12005: 12007: 22010: 22011: 22012: 12013: 22014: 22019: 1
2003high 2 (2007) · low 0 (2004)2026

Most influential descendants

Showing 5 of 14 · See all
  • K033734Lightcycler Instrument Version 1.2Roche Diagnostics Corp. · 2003 · cites it directly10descendants
  • K051713Jbaids Anthrac Detection SystemIdaho Technology, Inc. · 2005 · 2 generations on4descendants
  • K070597Verigene System, F2 Nucleic Acid Test, F5 Nucleic Acid Test, and Mthfr Nucleic Acid TestNanosphere, Inc. · 2007 · 2 generations on3descendants
  • K071188Modification to Jbaids Anthrax Detection SystemIdaho Technology, Inc. · 2007 · 3 generations on3descendants
  • K092705Abbott M2000SP and Abbott M2000RT, Models 9K14-01 (G-Series), 9K14-02 (E-Series), 9K1501Abbott Molecular, Inc. · 2010 · cites it directly2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K922906Amplicor Chlamydia Trachomatis TestRoche Molecular Systems, Inc.1993—
−1K964506Roche Cobas Amplicor AnalyzerRoche Molecular Systems, Inc.19971
0 (this device)K012966Roche Cobas Taqman AnalyzerRoche Molecular Systems, Inc.2002—
+1K033734Lightcycler Instrument Version 1.2Roche Diagnostics Corp.20031
+2K051713Jbaids Anthrac Detection SystemIdaho Technology, Inc.2005—
+2K070597Verigene System, F2 Nucleic Acid Test, F5 Nucleic Acid Test, and Mthfr Nucleic…Nanosphere, Inc.2007—
+3K071188Modification to Jbaids Anthrax Detection SystemIdaho Technology, Inc.2007—
+1K092705Abbott M2000SP and Abbott M2000RT, Models 9K14-01 (G-Series), 9K14-02…Abbott Molecular, Inc.20107

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200510
200720
201020
201120
201210
201320
201420
201910
FIG. 1 — Lineage of K012966, 1993–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JJF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K964506Roche Cobas Amplicor Analyzer
Summary page 2
…AS TagMan Analyzer Predicate Regulatory Classification Predicate Predicate Device Class Number Product Name 510(k) Number COBAS AMPLICOR Class 1 862.2170 COBAS AMPLICOR K964506 Analyzer Analyzer 510(k) Premarket Notification August 31, 2001 CONFIDENTIAL AND PROPRIETARY 807.87(h), Attachment 1 Page 2…
Roche Molecular Systems, Inc.19971

What descends from it? (14)

See all 14

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K033734Lightcycler Instrument Version 1.2Roche Diagnostics Corp.200310
K092705Abbott M2000SP and Abbott M2000RT, Models 9K14-01 (G-Series), 9K14-02 (E-Series), 9K1501Abbott Molecular, Inc.20102

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 24, 2004Not classifiedZ-1119-04OtherImproper seating of a fiber optic cable which may cause pixel cross-talk leading to falsely elevated test results or cases of misalignment to sample-to-result mismatch.
May 24, 2004Not classifiedZ-1120-04OtherImproper seating of a fiber optic cable which may cause pixel cross-talk leading to falsely elevated test results or cases of misalignment to sample-to-result mismatch.
Mar 6, 2003Not classifiedZ-0917-03OtherA false positive result may be reported.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 1 company; a new generation was cleared every 4.7 years on median.
  • 2 of the 4 ancestors have a recall of any class.
  • Of its descendants, K033734 (Roche Diagnostics Corp., 2003) has gone furthest, with 10 descendants of its own.

How does its category compare?

All JJF devices

JJF Analyzer, Chemistry, Micro, For Clinical Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JJF

FDA has decided no JJF device since 2007.

Median 510(k) review, days · JJF

Too few JJF 510(k)s in any year since 2007 to give a median.

Recalls per year · JJF

2in 2018

2007: 22012: 32018: 2
2007high 3 (2012) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JJF

2in 2025

2007: 102008: 22009: 32010: 212011: 192012: 82013: 12014: 32022: 12023: 12025: 2
2007high 21 (2010) · low 0 (2015)2025

MAUDE reports naming product code JJF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—210
20080—02
20090—03
20100—021
20110—019
20120—38
20130—01
20140—03
20150—00
20160—00
20170—00
20180—20
20190—00
20200—00
20210—00
20220—01
20230—01
20240—00
20250—02
20260—0—

What is it indicated for?

The COBAS TagMan Analyzer is a fully automated amplification and detection system for

nucleic acids using 5° nuclease technology. The COBAS TagMan Analyzer is intended to be

used by laboratory professionals trained in laboratory techniques and on the use of the

Analyzer.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K012966: Roche Cobas Taqman Analyzer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K012966

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026