510(k) · Cleared
Recalled · 2007Lightcycler Instrument Version 1.2
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Roche Diagnostics Corp. · JJF Analyzer, Chemistry, Micro, For Clinical Use
Cleared on Dec 17, 2003 as substantially equivalent to K012966 (Roche Molecular Systems, Inc., 2002) after a 19-day review. Its family tree reaches back 3 generations to 1993. 10 later devices descend from it. It has been recalled once.
- Decided
- Dec 17, 2003
- Review
- 19 days
- Class
- Class I
- Product code
- JJF
- Recalls
- 1
- Predicates
- 1
Its longest line of descent runs back 3 generations to K922906 (1993). 5 ancestors in all. 10 later devices descend from it, 3 citing it directly.
- Root · 3 generations backK922906 Amplicor Chlamydia Trachomatis TestRoche Molecular Systems, Inc. · 1993
- 2 generations backK964506 Roche Cobas Amplicor AnalyzerRoche Molecular Systems, Inc. · 1997 · Recalled
- 1 generation backK012966 Roche Cobas Taqman AnalyzerRoche Molecular Systems, Inc. · 2002 · Recalled
- This deviceK033734 Lightcycler Instrument Version 1.2Roche Diagnostics Corp. · 2003
- Descendant · cites it directly · 4 of its ownK051713 Jbaids Anthrac Detection SystemIdaho Technology, Inc. · 2005
- Descendant · cites it directly · 3 of its ownK070597 Verigene System, F2 Nucleic Acid Test, F5 Nucleic Acid Test, and…Nanosphere, Inc. · 2007
- Descendant · 2 generations on · 3 of its ownK071188 Modification to Jbaids Anthrax Detection SystemIdaho Technology, Inc. · 2007
- Descendant · 2 generations on · 1 of its ownK140426 Anthracis Real-Time PCR AssayCenters For Disease Control and Prevention (CDC) · 2014
- Descendant · 3 generations on · 0 of its ownK192871 B. anthracis Real-time PCR AssayCenters for Disease Control and Prevention · 2019
Showing 5 of 10 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 10 · See all- K051713Jbaids Anthrac Detection SystemIdaho Technology, Inc. · 2005 · cites it directly4descendants
- K070597Verigene System, F2 Nucleic Acid Test, F5 Nucleic Acid Test, and Mthfr Nucleic Acid TestNanosphere, Inc. · 2007 · cites it directly3descendants
- K071188Modification to Jbaids Anthrax Detection SystemIdaho Technology, Inc. · 2007 · 2 generations on3descendants
- K140426Anthracis Real-Time PCR AssayCenters For Disease Control and Prevention (CDC) · 2014 · 2 generations on1descendant
- K192871B. anthracis Real-time PCR AssayCenters for Disease Control and Prevention · 2019 · 3 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K922906 | Amplicor Chlamydia Trachomatis Test | Roche Molecular Systems, Inc. | 1993 | — |
| −2 | K964506 | Roche Cobas Amplicor Analyzer | Roche Molecular Systems, Inc. | 1997 | 1 |
| −1 | K012966 | Roche Cobas Taqman Analyzer | Roche Molecular Systems, Inc. | 2002 | 3 |
| 0 (this device) | K033734 | Lightcycler Instrument Version 1.2 | Roche Diagnostics Corp. | 2003 | — |
| +1 | K051713 | Jbaids Anthrac Detection System | Idaho Technology, Inc. | 2005 | — |
| +1 | K070597 | Verigene System, F2 Nucleic Acid Test, F5 Nucleic Acid Test, and Mthfr Nucleic… | Nanosphere, Inc. | 2007 | — |
| +2 | K071188 | Modification to Jbaids Anthrax Detection System | Idaho Technology, Inc. | 2007 | — |
| +2 | K140426 | Anthracis Real-Time PCR Assay | Centers For Disease Control and Prevention (CDC) | 2014 | — |
| +3 | K192871 | B. anthracis Real-time PCR Assay | Centers for Disease Control and Prevention | 2019 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1994 | 1 | 0 |
| 1995 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2005 | 1 | 0 |
| 2007 | 2 | 0 |
| 2010 | 1 | 0 |
| 2011 | 2 | 0 |
| 2013 | 1 | 0 |
| 2014 | 2 | 0 |
| 2019 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JJF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K012966 | Roche Cobas Taqman AnalyzerSummary page 2…ial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed COBAS TaqMan Analyzer K012966. The LightCycler Instrument is a flexible benchtop analyzer that automates the amplification a… | 2002 |
What descends from it? (10)
3 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 30, 2007 | Not classified | Z-0459-2008 | Process design | False Negative Results: Air bubbles in the lens or the area surrounding the lens of the LightCycler 100 uL capillaries may cause false negative results. The affected capillaries showed up to a 40% reduction in signal |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 10 years and passes through 2 companies; a new generation was cleared every 3.9 years on median.
- 3 of the 5 ancestors have a recall of any class.
- Of its descendants, K051713 (Idaho Technology, Inc., 2005) has gone furthest, with 4 descendants of its own.
- 1 other device also cites its predicate K012966.
How does its category compare?
All JJF devicesJJF Analyzer, Chemistry, Micro, For Clinical Use: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JJF
FDA has decided no JJF device since 2007.
Median 510(k) review, days · JJF
Too few JJF 510(k)s in any year since 2007 to give a median.
Recalls per year · JJF
2in 2018
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JJF
2in 2025
MAUDE reports naming product code JJF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 2 | 10 |
| 2008 | 0 | — | 0 | 2 |
| 2009 | 0 | — | 0 | 3 |
| 2010 | 0 | — | 0 | 21 |
| 2011 | 0 | — | 0 | 19 |
| 2012 | 0 | — | 3 | 8 |
| 2013 | 0 | — | 0 | 1 |
| 2014 | 0 | — | 0 | 3 |
| 2015 | 0 | — | 0 | 0 |
| 2016 | 0 | — | 0 | 0 |
| 2017 | 0 | — | 0 | 0 |
| 2018 | 0 | — | 2 | 0 |
| 2019 | 0 | — | 0 | 0 |
| 2020 | 0 | — | 0 | 0 |
| 2021 | 0 | — | 0 | 0 |
| 2022 | 0 | — | 0 | 1 |
| 2023 | 0 | — | 0 | 1 |
| 2024 | 0 | — | 0 | 0 |
| 2025 | 0 | — | 0 | 2 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The LightCycler Instrument is a fully automated amplification and detection system for
nucleic acids using fluorescence detection.
The LightCycler is intended to be used by laboratory professionals trained in laboratory
techniques and on the use of the analyzer.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K033734 · printed Nov 6, 2026
Cite this page
510(k) Report. “K033734: Lightcycler Instrument Version 1.2.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K033734
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026