510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Bioscanner 2000

K013068510(k) TraditionalCleared Nov 20, 2001145-day reviewClass II21 CFR 862.1345Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Polymer Technology Systems, Inc. · CGA Glucose Oxidase, Glucose

Cleared on Nov 20, 2001 as substantially equivalent to K982002 (Boehringer Mannheim Corp., 1998) after a 145-day review. Its family tree reaches back 2 generations to 1998. 5 later devices descend from it. It has been recalled once.

Decided
Nov 20, 2001
Review
145 days
Class
Class II
Product code
CGA
Recalls
1
Predicates
1

Its longest line of descent runs back 2 generations to K980731 (1998). 5 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2004: 1 descendant cleared2007: 2 descendants cleared2016: 1 descendant cleared2014: 1 descendant cleared×2: 2 generations cleared close together, K980731 (1998), K982002 (1998) 1995200020052010201520202025
Descendants, by generation 1 generation forward: 3 descendants2 generations forward: 2 descendants K982002 · Accu-Chek Comfort Curve Test Strip · Boehringer Mannheim Corp. · 1998 · recalledK980731 · Accu-Chek Comfort Curve Test Strips · Boehringer Mannheim Corp. · 19980
  1. Root · 2 generations backK980731 Accu-Chek Comfort Curve Test StripsBoehringer Mannheim Corp. · 1998
  2. 1 generation backK982002 Accu-Chek Comfort Curve Test StripBoehringer Mannheim Corp. · 1998 · Recalled
  3. This deviceK013068 Bioscanner 2000Polymer Technology Systems, Inc. · 2001
  4. Descendant · cites it directly · 2 of its ownK071507 PTS Panels Chol+HDL+Glu Panel Test StripsPolymer Technology Systems, Inc. · 2007 · Recalled
  5. Descendant · cites it directly · 1 of its ownK070017 PTS Panels Metabolic Chemistry Panel Test StripsPolymer Technology Systems, Inc. · 2007
  6. Descendant · cites it directly · 1 of its ownK041750 PTS Panels Chol+Glu Test PanelPolymer Technology Systems, Inc. · 2004 · Recalled
  7. Descendant · 2 generations on · 0 of its ownK162282 CardioChek Plus Test System, CardioChek Home Test SystemPolymer Technology Systems, Inc. D/B/A Pts Diagnostics · 2016 · Recalled
  8. Descendant · 2 generations on · 0 of its ownK142302 BioScanner Plus Glucose Test System, PTS PANELS Chol+Glu Test Panel…Polymer Technology Systems, Inc. · 2014

Most influential descendants

Showing 5 of 5 · See all
  • K071507PTS Panels Chol+HDL+Glu Panel Test StripsPolymer Technology Systems, Inc. · 2007 · cites it directly2descendants
  • K070017PTS Panels Metabolic Chemistry Panel Test StripsPolymer Technology Systems, Inc. · 2007 · cites it directly1descendant
  • K041750PTS Panels Chol+Glu Test PanelPolymer Technology Systems, Inc. · 2004 · cites it directly1descendant
  • K162282CardioChek Plus Test System, CardioChek Home Test SystemPolymer Technology Systems, Inc. D/B/A Pts Diagnostics · 2016 · 2 generations on0descendants
  • K142302BioScanner Plus Glucose Test System, PTS PANELS Chol+Glu Test Panel System, PTS PANELS…Polymer Technology Systems, Inc. · 2014 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K980731Accu-Chek Comfort Curve Test StripsBoehringer Mannheim Corp.1998—
−1K982002Accu-Chek Comfort Curve Test StripBoehringer Mannheim Corp.19981
0 (this device)K013068Bioscanner 2000Polymer Technology Systems, Inc.2001—
+1K071507PTS Panels Chol+HDL+Glu Panel Test StripsPolymer Technology Systems, Inc.20074
+1K070017PTS Panels Metabolic Chemistry Panel Test StripsPolymer Technology Systems, Inc.2007—
+1K041750PTS Panels Chol+Glu Test PanelPolymer Technology Systems, Inc.20041
+2K162282CardioChek Plus Test System, CardioChek Home Test SystemPolymer Technology Systems, Inc. D/B/A Pts Diagnostics20162
+2K142302BioScanner Plus Glucose Test System, PTS PANELS Chol+Glu Test Panel System…Polymer Technology Systems, Inc.2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200720
201410
201610
FIG. 1 — Lineage of K013068, 1998–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CGA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CGA device

  1. #1 K252488 ABL90 FLEX PLUS Systemscore 12.8 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K982002Accu-Chek Comfort Curve Test Strip
Summary page 2
…Test Strips are substantially equivalent to the predicate device noted below. Name: Accu-Chek Comfort Curve Test Strips Device Company: Roche Diagnostics 510(k) Number: K 982002 Similarities and Differences (Predicate and BioScanner Glucose) Similarities Both systems measure Glucose concentrations in bl…
Boehringer Mannheim Corp.19981

Also cleared against K982002: 2 other devices →

What descends from it? (5)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K071507PTS Panels Chol+HDL+Glu Panel Test StripsPolymer Technology Systems, Inc.20072
K070017PTS Panels Metabolic Chemistry Panel Test StripsPolymer Technology Systems, Inc.20071
K041750PTS Panels Chol+Glu Test PanelPolymer Technology Systems, Inc.20041

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 4, 2013Class IIZ-1403-2013Under Investigation by firmInternal investigation demonstrated under-recovery for glucose when tested against a reference method. This could result in a low to moderate health risk.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 3 years and passes through 2 companies; a new generation was cleared every 3.2 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K071507 (Polymer Technology Systems, Inc., 2007) has gone furthest, with 2 descendants of its own.
  • 2 other devices also cite its predicate K982002.

How does its category compare?

All CGA devices

CGA Glucose Oxidase, Glucose: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CGA

2in 2024

2007: 72008: 62009: 102010: 82011: 92012: 102013: 32014: 52015: 42016: 42017: 22018: 22019: 22020: 22021: 12022: 12023: 42024: 22026 (partial year): 1
2007high 10 (2009) · low 0 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CGA

249in 2023

2007: 122 days median2008: 56 days median2009: 155 days median2010: 122 days median2011: 217 days median2012: 123 days median2013: 441 days median2014: 265 days median2015: 150 days median2016: 157 days median2023: 249 days median
2007high 441 (2013) · low 56 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CGA

3in 2023

2007: 22008: 22009: 42010: 62011: 32013: 32014: 22015: 6, 2 Class I2016: 22019: 12023: 3
2007high 6 (2010) · low 0 (2012)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · CGA

59in 2025

2007: 292008: 372009: 752010: 1722011: 4232012: 2332013: 1722014: 1222015: 1552016: 362017: 922018: 692019: 1162020: 682021: 482022: 2722023: 1,1412024: 2662025: 59
2007high 1,141 (2023) · low 29 (2007)2025

MAUDE reports naming product code CGA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077122229
2008656237
200910155475
201081226172
201192173423
2012101230233
201334413172
201452652122
201541506 (2)155
20164157236
20172—092
20182—069
20192—1116
20202—068
20211—048
20221—0272
2023424931,141
20242—0266
20250—059
20261—0—

What is it indicated for?

The BioScanner Glucose Test Strips are intended to be used by healthcare professionals to measure glucose in whole blood and by individuals with diabetes to measure glucose in fingerstick whole blood at home. Glucose measurements are used in the management of carbohydrate metabolism disorders.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K013068: Bioscanner 2000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K013068

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026