510(k) Report

510(k) · Cleared

Recalled · 2003

Accu-Chek Comfort Curve Test Strip

K982002510(k) TraditionalCleared Sep 2, 199886-day reviewClass II21 CFR 862.1345Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boehringer Mannheim Corp. · LFR Glucose Dehydrogenase, Glucose

Cleared on Sep 2, 1998 as substantially equivalent to K980731 (Boehringer Mannheim Corp., 1998) after an 86-day review. Its family tree reaches back 1 generation to 1998. 3 of the 51 devices in its family, counting itself, have a Class I recall. 49 later devices descend from it. It has been recalled once.

Decided
Sep 2, 1998
Review
86 days
Class
Class II
Product code
LFR
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K980731 (1998). 49 later devices descend from it, 3 citing it directly. 3 of the 51 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2001: 5 descendants cleared2000: 1 descendant cleared2004: 1 descendant cleared2007: 5 descendants cleared2002: 4 descendants cleared2006: 2 descendants cleared2003: 1 descendant cleared2010: 3 descendants cleared, 2 with a Class I recall22005: 2 descendants cleared2008: 2 descendants cleared2015: 5 descendants cleared2012: 4 descendants cleared, 1 with a Class I recall12016: 3 descendants cleared2014: 1 descendant cleared2017: 2 descendants cleared2009: 1 descendant cleared2013: 3 descendants cleared2018: 1 descendant cleared2022: 2 descendants cleared2011: 1 descendant clearedK980731 · Accu-Chek Comfort Curve Test Strips · Boehringer Mannheim Corp. · 19981995200020052010201520202025
Descendants, by generation 1 generation forward: 3 descendants2 generations forward: 12 descendants, 1 with a Class I recall13 generations forward: 14 descendants4 generations forward: 13 descendants, 2 with a Class I recall25 generations forward: 7 descendants K980731 · Accu-Chek Comfort Curve Test Strips · Boehringer Mannheim Corp. · 19980+5
  1. Root · 1 generation backK980731 Accu-Chek Comfort Curve Test StripsBoehringer Mannheim Corp. · 1998
  2. This deviceK982002 Accu-Chek Comfort Curve Test StripBoehringer Mannheim Corp. · 1998
  3. Descendant · cites it directly · 33 of its ownK992684 Freestyle Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2000 · Recalled
  4. Descendant · 2 generations on · 23 of its ownK012014 Modification to Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2001
  5. Descendant · 2 generations on · 22 of its ownK020866 Freestyle Tracker Diabetes Management SystemAbbott Diabetes Care, Inc. · 2002
  6. Descendant · 2 generations on · 18 of its ownK031260 Freestyle Blood Glucose Monitoring System, Freestyle Tracker…Abbott Diabetes Care, Inc. · 2003
  7. Descendant · 2 generations on · 12 of its ownK050500 Freestyle 600 Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2005

Showing 5 of 49 descendants. See all descendants

Descendants cleared per year

2000: 12001: 52002: 42003: 12004: 12005: 22006: 22007: 52008: 22009: 12010: 32011: 12012: 42013: 32014: 12015: 52016: 32017: 22018: 12022: 2
2000high 5 (2001) · low 0 (2019)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 49 · See all
  • K992684Freestyle Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2000 · cites it directly33descendants
  • K012014Modification to Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2001 · 2 generations on23descendants
  • K020866Freestyle Tracker Diabetes Management SystemAbbott Diabetes Care, Inc. · 2002 · 2 generations on22descendants
  • K031260Freestyle Blood Glucose Monitoring System, Freestyle Tracker Diabetes Management System…Abbott Diabetes Care, Inc. · 2003 · 2 generations on18descendants
  • K050500Freestyle 600 Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2005 · 2 generations on12descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K980731Accu-Chek Comfort Curve Test StripsBoehringer Mannheim Corp.1998—
0 (this device)K982002Accu-Chek Comfort Curve Test StripBoehringer Mannheim Corp.1998—
+1K992684Freestyle Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc.20002
+2K012014Modification to Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc.2001—
+2K020866Freestyle Tracker Diabetes Management SystemAbbott Diabetes Care, Inc.2002—
+2K031260Freestyle Blood Glucose Monitoring System, Freestyle Tracker Diabetes…Abbott Diabetes Care, Inc.2003—
+2K050500Freestyle 600 Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc.2005—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200010
200150
200240
200310
200410
200520
200620
200750
200820
200910
201032
201110
201241
201330
201410
201550
201630
201720
201810
202220
FIG. 1 — Lineage of K982002, 1998–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K980731Accu-Chek Comfort Curve Test Strips
Summary page 2
…cts in commercial distribution intended for similar use. Most notably it is substantially equivalent to the previously 510(k) cleared Accu-Chek Comfort Curve Test Strip (k980731). Blood glucose concentrations can be expressed two ways, as whole blood glucose or as plasma (or serum) glucose. The concentrat…
Boehringer Mannheim Corp.1998—

Also cleared against K980731: 2 other devices →

What descends from it? (49)

See all 49

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K992684Freestyle Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc.200033
K013203Bioscanner Beyond Glucose Analyzer and Test StripsPolymer Technology Systems, Inc.20018
K013068Bioscanner 2000Polymer Technology Systems, Inc.20015

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 12, 2003Not classifiedZ-0715-03OtherCrack in the bottom of the test strip vial will cause erroneously low or high blood glucose readings.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K092638Freestyle Blood Glucose Test Strips, Models 100-Count: 12401-01, 12101-21, 12101-22, 12101-23Abbott Diabetes Care, Inc. · 2010 · 2 generations on · ◆ in the 2010 descendants bar of FIG. 12 Class I recalls
  • K092602FREESTYLE Lite Blood Glucose Monitoring System and Freestyle Freedom Lite Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2010 · 4 generations on · ◆ in the 2010 descendants bar of FIG. 12 Class I recalls
  • K120568Freestyle Insulinx Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2012 · 4 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K992684 (Abbott Diabetes Care, Inc., 2000) has gone furthest, with 33 descendants of its own.
  • 2 other devices also cite its predicate K980731.

How does its category compare?

All LFR devices

LFR Glucose Dehydrogenase, Glucose: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LFR

1in 2015

2007: 12009: 12011: 62012: 132013: 32014: 32015: 1
2007high 13 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LFR

161in 2014

2011: 296 days median2012: 152 days median2013: 212 days median2014: 161 days median
2007high 296 (2011) · low 152 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LFR

1in 2012

2007: 32008: 12009: 42011: 12012: 1
2007high 4 (2009) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LFR

827in 2025

2007: 23,8192008: 6,5842009: 10,8232010: 4,8652011: 3,8022012: 3,3412013: 3,8212014: 3,9602015: 3,3662016: 2,1452017: 1032018: 1082019: 1082020: 1322021: 1642022: 1012023: 1032024: 1712025: 827
2007high 23,819 (2007) · low 101 (2022)2025

MAUDE reports naming product code LFR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—323,819
20080—16,584
20091—410,823
20100—04,865
2011629613,802
20121315213,341
2013321203,821
2014316103,960
20151—03,366
20160—02,145
20170—0103
20180—0108
20190—0108
20200—0132
20210—0164
20220—0101
20230—0103
20240—0171
20250—0827
20260—0—

What is it indicated for?

The Accu-Chek Comfort Curve Test Strips are to be used with the Accu-Chek® Advantage® and Accu-Chek® Complete™ Monitors. The Accu-Chek Advantage and Accu-Chek Complete systems are designed for testing glucose in whole blood by persons with diabetes or by health care professionals in the home or in health care facilities.

Professionals may use the test strips to test capillary, venous, arterial and neonate (including cord) blood samples; lay use is limited to capillary blood testing.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K982002: Accu-Chek Comfort Curve Test Strip.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K982002

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026