510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

BCS System

K013114510(k) TraditionalCleared Nov 9, 200152-day reviewClass II21 CFR 864.5425Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Dade Behring, Inc. · JPA System, Multipurpose For In Vitro Coagulation Studies

Cleared on Nov 9, 2001 as substantially equivalent to K001145 (Dade Behring, Inc., 2000) after a 52-day review. Its family tree reaches back 1 generation to 2000. It has been recalled 2 times.

Decided
Nov 9, 2001
Review
52 days
Class
Class II
Product code
JPA
Recalls
2
Predicates
1

Its longest line of descent runs back 1 generation to K001145 (2000). No later device cites it as a predicate yet.

K001145 · Sysmex Automated Coagulation Analyzer, Model CA-6000 · Dade Behring, Inc. · 200020002002
  1. Root · 1 generation backK001145 Sysmex Automated Coagulation Analyzer, Model CA-6000Dade Behring, Inc. · 2000
  2. This deviceK013114 BCS SystemDade Behring, Inc. · 2001
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K001145Sysmex Automated Coagulation Analyzer, Model CA-6000Dade Behring, Inc.2000—
0 (this device)K013114BCS SystemDade Behring, Inc.2001—
FIG. 1 — Lineage of K013114, 2000–2001Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JPA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JPA device

  1. #1 K243858 TEG 6s Hemostasis System Citrated: K, KH, RT…score 12.0 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K001145Sysmex Automated Coagulation Analyzer, Model CA-6000
Summary page 1
…: Multipurpose system for in vitro Coagulation studies Classification Number: Class Il (864.5425) Identification of the Legally Marketed Device: Sysmex® CA-6000 System (K001145) Device Description: B The current BCS™ System was originally determined to be substantially equivalent as a fully automated ph…
Dade Behring, Inc.2000—

Also cleared against K001145: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 14, 2016Class IIZ-0824-2016Software designFalse short clotting times for PT on BCS and BCS XP with Dade Innovin can occur. This has also an effect for false low PT INR values. These false short clotting times were observed only with turbid, hemolytic and/or icteric samples from…
Dec 1, 2010Class IIZ-1867-2012Device DesignFirm has confirmed that if the Cleaner SCS vial becomes empty after processing a test for Antithrombin assay using the INNOVANCE(R) Antithrombin kit, and the test immediately following is an APTT based clotting assay several results may…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • 1 other device also cites its predicate K001145.

How does its category compare?

All JPA devices

JPA System, Multipurpose For In Vitro Coagulation Studies: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JPA

3in 2025

2007: 12008: 22009: 62010: 12011: 12013: 12015: 22016: 62017: 42018: 12019: 32021: 12022: 12023: 32024: 12025: 32026 (partial year): 1
2007high 6 (2009) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JPA

30in 2025

2009: 91 days median2016: 114 days median2017: 117 days median2019: 175 days median2023: 506 days median2025: 30 days median
2007high 506 (2023) · low 30 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JPA

1in 2020

2007: 12010: 62011: 42012: 12016: 32020: 12026 (partial year): 1
2007high 6 (2010) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JPA

11in 2025

2007: 1,7442008: 9182009: 1,3632010: 7272011: 1612012: 522013: 432014: 482015: 452016: 62017: 152018: 392019: 272020: 182021: 72022: 92023: 112024: 82025: 11
2007high 1,744 (2007) · low 6 (2016)2025

MAUDE reports naming product code JPA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—11,744
20082—0918
200969101,363
20101—6727
20111—4161
20120—152
20131—043
20140—048
20152—045
2016611436
20174117015
20181—039
20193175027
20200—118
20211—07
20221—09
20233506011
20241—08
2025330011
20261—1—

What is it indicated for?

The BCS™ System is an automated coagulation analyzer for in vitro diagnostic

use in clinical laboratories. The instrument performs the following parameters:

o Activated Partial Thromboplastin Time (APTT) e Heparin

e Antithrombin lila ¢ Lupus Anticoagulants

¢ Batroxobin ¢ Prothrombin Time (PT)

¢ D-dimer ¢ Plasminogen

¢ Deficient Plasmas e Protein C-clotting

e Derived Fibrinogen ¢ Protein C-chromogenic

¢ Factor V Leiden e Thrombin Time

¢ Fibrinogen ¢ von Willebrand factor

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K013114: BCS System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K013114

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026