510(k) Report

510(k) · Cleared

Cleared

Citrated: K, KH, RTH, FFH

K232018510(k) TraditionalCleared Mar 29, 2024266-day reviewClass II21 CFR 864.5425

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Haemonetics Corporation · JPA System, Multipurpose For In Vitro Coagulation Studies

Cleared on Mar 29, 2024 as substantially equivalent to K150041 (Coramed Technologies, LLC, 2015) after a 266-day review. Its family tree reaches back 2 generations to 2000.

Decided
Mar 29, 2024
Review
266 days
Class
Class II
Product code
JPA
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K002177 (2000). No later device cites it as a predicate yet.

  1. Root · 2 generations backK002177 Thrombelastograph Coagulation Analyzer Teg-5000 SeriesHaemoscope Corp. · 2000 · Recalled
  2. 1 generation backK150041 CORA (Coagulation Resonance Analysis) SystemCoramed Technologies, LLC · 2015
  3. This deviceK232018 Citrated: K, KH, RTH, FFHHaemonetics Corporation · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K002177Thrombelastograph Coagulation Analyzer Teg-5000 SeriesHaemoscope Corp.20007
−1K150041CORA (Coagulation Resonance Analysis) SystemCoramed Technologies, LLC2015—
0 (this device)K232018Citrated: K, KH, RTH, FFHHaemonetics Corporation2024—
FIG. 1 — Lineage of K232018, 2000–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JPA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JPA device

  1. #1 K243858 TEG 6s Hemostasis System Citrated: K, KH, RT…score 12.0 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K150041CORA (Coagulation Resonance Analysis) System
Summary page 5
…FR 864.5425 Product Code: JPA 2. Legally Marketed Predicate Device Predicate # Predicate Trade Name Product Code K150041 TEG 6s with the Citrated Multichannel Cartridge JPA 3. Device Description Summary TEG® 6s System Description T…
Coramed Technologies, LLC2015—

Also cleared against K150041: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 3 companies; a new generation was cleared every 14.8 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • 2 other devices also cite its predicate K150041.

How does its category compare?

All JPA devices

JPA System, Multipurpose For In Vitro Coagulation Studies: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · JPA

3in 2025

2007: 12008: 22009: 62010: 12011: 12013: 12015: 22016: 62017: 42018: 12019: 32021: 12022: 12023: 32024: 12025: 32026 (partial year): 1
2007high 6 (2009) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JPA

30in 2025

2009: 91 days median2016: 114 days median2017: 117 days median2019: 175 days median2023: 506 days median2025: 30 days median
2007high 506 (2023) · low 30 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JPA

1in 2020

2007: 12010: 62011: 42012: 12016: 32020: 12026 (partial year): 1
2007high 6 (2010) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JPA

11in 2025

2007: 1,7442008: 9182009: 1,3632010: 7272011: 1612012: 522013: 432014: 482015: 452016: 62017: 152018: 392019: 272020: 182021: 72022: 92023: 112024: 82025: 11
2007high 1,744 (2007) · low 6 (2016)2025

MAUDE reports naming product code JPA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—11,744
20082—0918
200969101,363
20101—6727
20111—4161
20120—152
20131—043
20140—048
20152—045
2016611436
20174117015
20181—039
20193175027
20200—118
20211—07
20221—09
20233506011
20241—08
2025330011
20261—1—

What is it indicated for?

The TEG 6s Hemostasis System consists of the TEG 6s Hemostasis Analyzer and the Citrated: K, KH, RTH, FFH assay cartridge. The TEG 6s Hemostasis System is intended for in vitro diagnostic use with adult patients where an evaluation of their blood hemostasis properties is desired. The TEG 6s Hemostasis System records the kinetic changes in a sample of 3.2% citrated whole blood as the sample clots and provides semi-quantitative results. The TEG 6s…

Read the full statement (415 words)

The TEG 6s Hemostasis System consists of the TEG 6s Hemostasis Analyzer and the Citrated: K, KH, RTH, FFH assay cartridge. The TEG 6s Hemostasis System is intended for in vitro diagnostic use with adult patients where an evaluation of their blood hemostasis properties is desired. The TEG 6s Hemostasis System records the kinetic changes in a sample of 3.2% citrated whole blood as the sample clots and provides semi-quantitative results. The TEG 6s Hemostasis System can be used in the laboratory or at the point-of-care.

The Citrated: K, KH, RTH, FFH assay cartridge is intended to be used in patients where heparin/heparinoids may be present and who are at an increased risk of coagulopathy. Hemostasis evaluations are indicated to assess clinical conditions in cardiovascular surgery, cardiovascular procedures (e.g. minimally invasive valve replacement or repairs) and liver transplantation to assess hemorrhage or thrombosis conditions before, during and following the procedure.

The Citrated: K, KH, RTH, FFH assay cartridge contains four independent assays (CK, CKH, CRTH and CFFH) and the system output consists of a table of numerical values for parameters R, MA, and LY30.

The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. Clotting characteristics are described by the functional parameters R (clotting time) and MA (maximum clot strength).

The CKH assay monitors the effects of heparin in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. CKH is used in conjunction with CK, and heparin influence is determined by comparing Clotting Times (R) between the two tests. LY30 describes fibrinolysis 30 minutes after reaching maximum clot strength.

The CRTH assay monitors the hemostasis process after stimulation of both the intrinsic and extrinsic pathways in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System, neutralizing the effect of heparin in the sample. Clotting characteristics are described by the functional parameter MA (maximum clot strength with contributions of both platelets and fibrin).

The CFFH assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s Hemostasis System after blocking platelet contributions to clot strength, neutralizing the effect of heparin in the sample. Clotting characteristics are described by the functional parameter MA (fibrinogen contribution to maximum clot strength).

Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests.

For professional use only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K232018: Citrated: K, KH, RTH, FFH.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K232018

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026