510(k) Report

510(k) · Cleared

Recalled · 2002

Omni C Analyzer

K013373510(k) SpecialCleared Nov 2, 200122-day reviewClass II21 CFR 862.1600Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diagnostics Corp. · CEM Electrode, Ion Specific, Potassium

Cleared on Nov 2, 2001 as substantially equivalent to K990092 (Avl Scientific Corp., 1999) after a 22-day review. Its family tree reaches back 2 generations to 1994. 2 later devices descend from it. It has been recalled once.

Decided
Nov 2, 2001
Review
22 days
Class
Class II
Product code
CEM
Recalls
1
Predicates
1

Its longest line of descent runs back 2 generations to K940918 (1994). 4 ancestors in all. 2 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor1995: 1 other ancestor2003: 1 descendant cleared2012: 1 descendant clearedK940918 · I-Stat 200 Portable Clinical Analyzer · I-Stat Corp. · 1994 · recalledK990092 · Avl Omni Modular Analyzer · Avl Scientific Corp. · 1999 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant K990092 · Avl Omni Modular Analyzer · Avl Scientific Corp. · 1999 · recalledK940918 · I-Stat 200 Portable Clinical Analyzer · I-Stat Corp. · 1994 · recalled0
  1. Root · 2 generations backK940918 I-Stat 200 Portable Clinical AnalyzerI-Stat Corp. · 1994 · Recalled
  2. 1 generation backK990092 Avl Omni Modular AnalyzerAvl Scientific Corp. · 1999 · Recalled
  3. This deviceK013373 Omni C AnalyzerRoche Diagnostics Corp. · 2001
  4. Descendant · cites it directly · 1 of its ownK032311 Roche Diagnostics Omni S AnalyzerRoche Diagnostics Corp. · 2003 · Recalled
  5. Descendant · 2 generations on · 0 of its ownK111188 Cobas B 123 Poc System, Cobas B 123 Autoqc Pack Tri-Level, Cobas B…Roche Diagnostics · 2012 · Recalled

Most influential descendants

Showing 2 of 2 · See all
  • K032311Roche Diagnostics Omni S AnalyzerRoche Diagnostics Corp. · 2003 · cites it directly1descendant
  • K111188Cobas B 123 Poc System, Cobas B 123 Autoqc Pack Tri-Level, Cobas B 123 Autocvc Pack…Roche Diagnostics · 2012 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K940918I-Stat 200 Portable Clinical AnalyzerI-Stat Corp.199419
−1K990092Avl Omni Modular AnalyzerAvl Scientific Corp.19994
0 (this device)K013373Omni C AnalyzerRoche Diagnostics Corp.2001—
+1K032311Roche Diagnostics Omni S AnalyzerRoche Diagnostics Corp.20039
+2K111188Cobas B 123 Poc System, Cobas B 123 Autoqc Pack Tri-Level, Cobas B 123 Autocvc…Roche Diagnostics20124

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
201210
FIG. 1 — Lineage of K013373, 1994–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CEM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CEM device

  1. #1 K253491 ISE indirect K for Gen.2; ISE indirect Na for…score 11.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K990092Avl Omni Modular Analyzer
Summary page 2
…pH, Blood Gas, Electrolyte, Hematocrit, Hemoglobin and Oxygen Saturation System We claim substantial equivalence to the current legally marketed Omni Modular Analyzer (K990092). The Omni C analyzer represents a combined blood gas, electrolyte, total hemoglobin, hematocrit and oxygen saturation test syst…
Avl Scientific Corp.19994

Also cleared against K990092: 3 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K032311Roche Diagnostics Omni S AnalyzerRoche Diagnostics Corp.20031

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 16, 2002Not classifiedZ-0438-03OtherAir bubbles trapped in the PO2 electrode will cause falsely elevated patient test results.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 7 years and passes through 3 companies; a new generation was cleared every 4.7 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • Of its descendants, K032311 (Roche Diagnostics Corp., 2003) has gone furthest, with 1 descendant of its own.
  • 3 other devices also cite its predicate K990092.

How does its category compare?

All CEM devices

CEM Electrode, Ion Specific, Potassium: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CEM

1in 2025

2007: 12009: 12010: 32011: 32012: 42020: 12023: 22025: 12026 (partial year): 1
2007high 4 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CEM

180in 2012

2010: 195 days median2011: 74 days median2012: 180 days median
2007high 195 (2010) · low 74 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CEM

4in 2025

2008: 12010: 122011: 62016: 12019: 12020: 32021: 12023: 22025: 4
2007high 12 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CEM

29in 2025

2007: 662008: 2452009: 792010: 1132011: 1262012: 452013: 372014: 42015: 72016: 62017: 322018: 132019: 252020: 272021: 182022: 162023: 302024: 422025: 29
2007high 245 (2008) · low 4 (2014)2025

MAUDE reports naming product code CEM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—066
20080—1245
20091—079
2010319512113
20113746126
20124180045
20130—037
20140—04
20150—07
20160—16
20170—032
20180—013
20190—125
20201—327
20210—118
20220—016
20232—230
20240—042
20251—429
20261—0—

What is it indicated for?

The Roche Diagnostics Omni C Analyzer is intended to be used for the measurement of pH, P02, PCO2, sodium, potassium, ionized calcium, chloride, hematocrit, total hemoglobin and oxygen saturation in samples of whole blood, serum, plasma and aqueous solutions as appropriate.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K013373: Omni C Analyzer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K013373

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026