Roche Diagnostics Omni S Analyzer

K032311510(k) TraditionalCleared Oct 17, 200381-day reviewClass II21 CFR 862.1600Recalls 9

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diagnostics Corp. · CEM Electrode, Ion Specific, Potassium

Cleared on Oct 17, 2003 as substantially equivalent to K990092 (Avl Scientific Corp., 1999) and 1 other predicate after an 81-day review. Its family tree reaches back 3 generations to 1994. 1 later device descends from it. It has been recalled 9 times.

Decided
Oct 17, 2003
Review
81 days
Class
Class II
Product code
CEM
Recalls
9
Predicates
2

Its longest line of descent runs back 3 generations to K940918 (1994). 5 ancestors in all. 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor1995: 1 other ancestor2012: 1 descendant clearedK940918 · I-Stat 200 Portable Clinical Analyzer · I-Stat Corp. · 1994 · recalledK990092 · Avl Omni Modular Analyzer · Avl Scientific Corp. · 1999 · recalledK013373 · Omni C Analyzer · Roche Diagnostics Corp. · 2001 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors 1 generation forward: 1 descendant K013373 · Omni C Analyzer · Roche Diagnostics Corp. · 2001 · recalledK990092 · Avl Omni Modular Analyzer · Avl Scientific Corp. · 1999 · recalledK940918 · I-Stat 200 Portable Clinical Analyzer · I-Stat Corp. · 1994 · recalled0
  1. Root · 3 generations backK940918 I-Stat 200 Portable Clinical AnalyzerI-Stat Corp. · 1994 · Recalled
  2. 2 generations backK990092 Avl Omni Modular AnalyzerAvl Scientific Corp. · 1999 · Recalled
  3. 1 generation backK013373 Omni C AnalyzerRoche Diagnostics Corp. · 2001 · Recalled
  4. This deviceK032311 Roche Diagnostics Omni S AnalyzerRoche Diagnostics Corp. · 2003
  5. Descendant · cites it directly · 0 of its ownK111188 Cobas B 123 Poc System, Cobas B 123 Autoqc Pack Tri-Level, Cobas B…Roche Diagnostics · 2012 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K111188Cobas B 123 Poc System, Cobas B 123 Autoqc Pack Tri-Level, Cobas B 123 Autocvc Pack…Roche Diagnostics · 2012 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K940918I-Stat 200 Portable Clinical AnalyzerI-Stat Corp.199419
−2K990092Avl Omni Modular AnalyzerAvl Scientific Corp.19994
−1K013373Omni C AnalyzerRoche Diagnostics Corp.20011
0 (this device)K032311Roche Diagnostics Omni S AnalyzerRoche Diagnostics Corp.2003—
+1K111188Cobas B 123 Poc System, Cobas B 123 Autoqc Pack Tri-Level, Cobas B 123 Autocvc…Roche Diagnostics20124

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
FIG. 1 — Lineage of K032311, 1994–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CEM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CEM device

  1. #1 K253491 ISE indirect K for Gen.2; ISE indirect Na for…score 11.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K990092Avl Omni Modular Analyzer
Summary page 1
…lobin, carboxyhemoglobin and methemoglobin. 3) Predicate We claim substantial equivalence to the current legally marketed Omni device Modular Analyzer (K990092) and the Omni C Analyzer (K013373). 4) Device The Omni S analyzer represents a combined pH, blood gas, electrolyte, Des…
Avl Scientific Corp.19994
*K013373Omni C Analyzer
Summary page 1
…moglobin. 3) Predicate We claim substantial equivalence to the current legally marketed Omni device Modular Analyzer (K990092) and the Omni C Analyzer (K013373). 4) Device The Omni S analyzer represents a combined pH, blood gas, electrolyte, Description glucose, lactate, u…
Roche Diagnostics Corp.20011

Also cleared against K990092: 3 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111188Cobas B 123 Poc System, Cobas B 123 Autoqc Pack Tri-Level, Cobas B 123 Autocvc Pack…Roche Diagnostics20120

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 22, 2018Class IIZ-1611-2018Software designThe software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
Feb 22, 2018Class IIZ-1613-2018Software designThe software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
Feb 22, 2018Class IIZ-1610-2018Software designThe software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
Feb 22, 2018Class IIZ-1612-2018Software designThe software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.
Oct 31, 2016Class IIZ-0673-2017Software designIncorrect Urea results may be generated sporadically on cobas b 221<6>Roche OMNI S6 system and cannot be detected by QC. QC measurement results can be affected as well. This issue is only detectable if comparison result of a second…
Jun 24, 2015Class IIZ-2330-2015Process controlThe affected S2 fluid packs may generate a transponder error when the fluid pack is loaded on to the cobas b 221 system. The error prevents the fluid from being recognized by the system and the instrument cannot start measurements until…
Dec 18, 2014Class IIZ-1041-2015Material/Component ContaminationChemical component of the F3 Fluid pack and environmental bacterial contamination causes a positive bias up to 39% at 4.1 mmol/l for the glucose measurement and 30% at 4.12 mmol/l for lactate. Urea readings are unaffected.
Sep 12, 2014Class IIIZ-0064-2015Labeling Change ControlThe default references for normal values are inconsistent between cobas b 221 and other blood gas instruments. In addition, the reference ranges are different when comparing with corresponding ranges listed in the Instructions for Use for…
Feb 28, 2014Class IIZ-1315-2014Device DesignReference electrode used beyond the guaranteed in-use 52 week lifetime,may leak and potentially cause erroneous pH and/or sodium results. Other Ion parameters are not affected.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 3 companies; a new generation was cleared every 2.8 years on median.
  • 4 of the 5 ancestors have a recall of any class.
  • Of its descendants, K111188 (Roche Diagnostics, 2012) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K990092.

How does its category compare?

All CEM devices

CEM Electrode, Ion Specific, Potassium: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CEM

1in 2025

2007: 12009: 12010: 32011: 32012: 42020: 12023: 22025: 12026 (partial year): 1
2007high 4 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CEM

180in 2012

2010: 195 days median2011: 74 days median2012: 180 days median
2007high 195 (2010) · low 74 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CEM

4in 2025

2008: 12010: 122011: 62016: 12019: 12020: 32021: 12023: 22025: 4
2007high 12 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CEM

29in 2025

2007: 662008: 2452009: 792010: 1132011: 1262012: 452013: 372014: 42015: 72016: 62017: 322018: 132019: 252020: 272021: 182022: 162023: 302024: 422025: 29
2007high 245 (2008) · low 4 (2014)2025

MAUDE reports naming product code CEM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—066
20080—1245
20091—079
2010319512113
20113746126
20124180045
20130—037
20140—04
20150—07
20160—16
20170—032
20180—013
20190—125
20201—327
20210—118
20220—016
20232—230
20240—042
20251—429
20261—0—

What is it indicated for?

The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to

be used for the measurement of pH, P02, PC02, sodium, potassium, ionized calcium, chloride,

hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin,

deoxyhemoglobin, carboxyhemoglobin and methemoglobin in samples of whole blood, serum,

plasma and aqueous solutions as appropriate.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K032311: Roche Diagnostics Omni S Analyzer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K032311

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026