HEARTSTREAMFR2 Aed with M3848A and M3849A, Models M3860A, M3861, M3840A, M3841A

K014157510(k) SpecialCleared Jan 17, 200229-day reviewClass III21 CFR 870.5310Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Jan 17, 2002 after a 29-day review. The FDA summary names its predicate without a number. 1 of the 10 devices in its family, counting itself, has a Class I recall. 9 later devices descend from it. It has been recalled 4 times.

Decided
Jan 17, 2002
Review
29 days
Class
Class III
Product code
MKJ
Recalls
4
Predicates
0

9 later devices descend from it, 4 citing it directly. 1 of the 10 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2004: 1 descendant cleared2011: 2 descendants cleared2005: 1 descendant cleared2013: 1 descendant cleared2012: 2 descendants cleared, 1 with a Class I recall12014: 1 descendant cleared2015: 1 descendant cleared200020052010201520202025
  1. This deviceK014157 HEARTSTREAMFR2 Aed with M3848A and M3849A, Models M3860A, M3861…Philips Medical Systems · 2002
  2. Descendant · cites it directly · 3 of its ownK040637 Powerheart Aed G3 ProCardiac Science, Inc. · 2004 · Recalled
  3. Descendant · cites it directly · 2 of its ownK111693 Philips Heartstart FR3 Automated External DefibrillatorPhilips Medical Systems · 2011 · Recalled
  4. Descendant · 2 generations on · 2 of its ownK102496 Powerheart Aed G3 Semi-Automatic, Powerheart Aed G3 Automatic…Cardiac Science Corporation · 2011 · Recalled
  5. Descendant · 3 generations on · 1 of its ownK122758 Powerheart G5 AedCardiac Science Corporation · 2014
  6. Descendant · cites it directly · 1 of its ownK042404 Lifepak, Model 1000Medtronic Emergency Response Systems, Inc. · 2005 · Recalled

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2004: 12005: 12011: 22012: 22013: 12014: 12015: 1
2004high 2 (2011) · low 0 (2006)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 9 · See all
  • K040637Powerheart Aed G3 ProCardiac Science, Inc. · 2004 · cites it directly3descendants
  • K111693Philips Heartstart FR3 Automated External DefibrillatorPhilips Medical Systems · 2011 · cites it directly2descendants
  • K102496Powerheart Aed G3 Semi-Automatic, Powerheart Aed G3 Automatic, Powerheart Aed G3 Pro…Cardiac Science Corporation · 2011 · 2 generations on2descendants
  • K122758Powerheart G5 AedCardiac Science Corporation · 2014 · 3 generations on1descendant
  • K042404Lifepak, Model 1000Medtronic Emergency Response Systems, Inc. · 2005 · cites it directly1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K014157HEARTSTREAMFR2 Aed with M3848A and M3849A, Models M3860A, M3861, M3840A, M3841APhilips Medical Systems2002—
+1K040637Powerheart Aed G3 ProCardiac Science, Inc.20048
+1K111693Philips Heartstart FR3 Automated External DefibrillatorPhilips Medical Systems20111
+2K102496Powerheart Aed G3 Semi-Automatic, Powerheart Aed G3 Automatic, Powerheart Aed…Cardiac Science Corporation20112
+3K122758Powerheart G5 AedCardiac Science Corporation2014—
+1K042404Lifepak, Model 1000Medtronic Emergency Response Systems, Inc.20052

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200510
201120
201221
201310
201410
201510
FIG. 1 — Lineage of K014157, 2002–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Heartstream FR2 AED
Named without a submission number
—
*—Medtronic/Physio-Control LifePak 500
Named without a submission number
—

What descends from it? (9)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K040637Powerheart Aed G3 ProCardiac Science, Inc.20043
K111693Philips Heartstart FR3 Automated External DefibrillatorPhilips Medical Systems20112
K042404Lifepak, Model 1000Medtronic Emergency Response Systems, Inc.20051
K130684Philips Heartstart FR3 Automated External DefibrillatorPhilips Medical Systems20130

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 28, 2010Not classifiedZ-0470-2011Nonconforming Material/ComponentFailure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable.
May 28, 2010Not classifiedZ-0472-2011Nonconforming Material/ComponentFailure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable.
May 28, 2010Not classifiedZ-0473-2011Nonconforming Material/ComponentFailure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable.
May 28, 2010Not classifiedZ-0471-2011Nonconforming Material/ComponentFailure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K122600Lifepak 1000 DefibrillatorPhysio-Control, Inc. · 2012 · 2 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K040637 (Cardiac Science, Inc., 2004) has gone furthest, with 3 descendants of its own.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

The Heartstream FR2 is indicated for use on persons experiencing the symptoms of sudden cardiac arrest:

Lack of responsiveness

Lack of breathing

When the patient is under 8 years or weighs less than 55 pounds (25 kg), the FR2 should be used with attenuated pediatric defibrillation pads. DO NOT DELAY TREATMENT TO DETERMINE THE CHILD’S EXACT AGE/WEIGHT.

The Heartstream FR2 is intended for use by personnel who have been trained in its operation. The user should be qualified by training in basic life support, advanced life support, or other physician-authorized emergency medical response.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K014157: HEARTSTREAMFR2 Aed with M3848A and M3849A, Models M3860A, M3861, M3840A, M3841A.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K014157

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026