Lifepak, Model 1000

K042404510(k) TraditionalCleared Dec 22, 2005475-day reviewClass III21 CFR 870.5310Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Emergency Response Systems, Inc. · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Dec 22, 2005 as substantially equivalent to K983393 (Physio-Control Corp., 1999) and 3 other predicates after a 475-day review. Its family tree reaches back 6 generations to 1993. 1 of the 26 devices in its family, counting itself, has a Class I recall. 1 later device descends from it. It has been recalled 2 times.

Decided
Dec 22, 2005
Review
475 days
Class
Class III
Product code
MKJ
Recalls
2
Predicates
4

Its longest line of descent runs back 6 generations to K921669 (1993). 24 ancestors in all, the oldest K802288 (1980). 1 later device descends from it, 1 citing it directly. 1 of the 26 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2002: 2 other ancestors1999: 1 other ancestor2001: 3 other ancestors1996: 2 other ancestors1990: 1 other ancestor1984: 1 other ancestor1995: 4 other ancestors1980: 1 other ancestor1992: 1 other ancestor1991: 1 other ancestor1994: 1 other ancestor2012: 1 descendant cleared, 1 with a Class I recall1K921669 · Marquette SL Series Transport Remote Acquisition · Marquette Electronics, Inc. · 1993K964750 · Eagle 4000 Patient Monitor · Marquette Electronics, Inc. · 1997K973486 · Life Pak 12 · Physio-Control Corp. · 1998 · recalledK991910 · Lifepak 12 Biphasic · Physio-Control Corp. · 1999 · recalledK012274 · Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/Pacemaker · Medtronic Physio-Control Corp. · 2002 · recalledK033275 · Lifepak 12, 20, 500, CR Plus · Medtronic Physio-Control Corp. · 2003 · recalled19801990200020102020
  1. Root · 6 generations backK921669 Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc. · 1993
  2. 5 generations backK964750 Eagle 4000 Patient MonitorMarquette Electronics, Inc. · 1997
  3. 4 generations backK973486 Life Pak 12Physio-Control Corp. · 1998 · Recalled
  4. 3 generations backK991910 Lifepak 12 BiphasicPhysio-Control Corp. · 1999 · Recalled
  5. 2 generations backK012274 Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/PacemakerMedtronic Physio-Control Corp. · 2002 · Recalled
  6. 1 generation backK033275 Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp. · 2003 · Recalled
  7. This deviceK042404 Lifepak, Model 1000Medtronic Emergency Response Systems, Inc. · 2005
  8. Descendant · cites it directly · 0 of its ownK122600 Lifepak 1000 DefibrillatorPhysio-Control, Inc. · 2012 · Recalled (Class I)

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K122600Lifepak 1000 DefibrillatorPhysio-Control, Inc. · 2012 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K921669Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc.1993—
−5K964750Eagle 4000 Patient MonitorMarquette Electronics, Inc.1997—
−4K973486Life Pak 12Physio-Control Corp.19983
−3K991910Lifepak 12 BiphasicPhysio-Control Corp.19993
−2K012274Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/PacemakerMedtronic Physio-Control Corp.200217
−1K033275Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp.20031
0 (this device)K042404Lifepak, Model 1000Medtronic Emergency Response Systems, Inc.2005—
+1K122600Lifepak 1000 DefibrillatorPhysio-Control, Inc.2012Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198410
199010
199110
199210
199410
199540
199620
199910
200130
200220

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201211
FIG. 1 — Lineage of K042404, 1993–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K983393Lifepak 500 Biphasic
Summary page 1
…nce: The features and functions of the LIFEPAK 1000 defibrillator are substantially equivalent to the: * Medtronic LIFEPAK 500 AED, 510(k) K983393 (05/05/99) and K033275 (1 1/06/03), and * Philips HeartStart FR2+ Defibrillator with ECU Assessment Module…
Physio-Control Corp.19991
*K033275Lifepak 12, 20, 500, CR Plus
Summary page 1
…tures and functions of the LIFEPAK 1000 defibrillator are substantially equivalent to the: * Medtronic LIFEPAK 500 AED, 510(k) K983393 (05/05/99) and K033275 (1 1/06/03), and * Philips HeartStart FR2+ Defibrillator with ECU Assessment Module, 5 10(k)…
Medtronic Physio-Control Corp.20031
*K013425Heartstream/Heartstart FR2 Aed Ecg Cable, Models M3860A, M3861A, M3840A, M3841A
Summary page 1
…, 510(k) K983393 (05/05/99) and K033275 (1 1/06/03), and * Philips HeartStart FR2+ Defibrillator with ECU Assessment Module, 5 10(k) K0 13425 (01/14/02) and K0 14157 (01/17/02). Description: The LIFEPAK 1000 defibrillator is a semi-automatic defib…
Philips Medical Systems200210
*K014157HEARTSTREAMFR2 Aed with M3848A and M3849A, Models M3860A, M3861, M3840A, M3841A
Summary page 1
…99) and K033275 (1 1/06/03), and * Philips HeartStart FR2+ Defibrillator with ECU Assessment Module, 5 10(k) K0 13425 (01/14/02) and K0 14157 (01/17/02). Description: The LIFEPAK 1000 defibrillator is a semi-automatic defibrillator with manual…
Philips Medical Systems20024

Also cleared against K983393: 5 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K122600Lifepak 1000 Defibrillator Class IPhysio-Control, Inc.20120

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 9, 2014Class IIZ-1660-2014Software Manufacturing/Software DeploymentPhysio-Control has become aware of incidents where customers have attempted to use their LIFEPAK 1000 defibrillator and the device has shut down unexpectedly due to a very low battery. A defibrillator in this scenario has the potential to…
Feb 27, 2008Not classifiedZ-1656-2008Component change controlPotential for the display screen to dim and eventually go blank. The risk to patient is a delay in defibrillation therapy if the defibrillator was needed in a cardiac arrest situation. Death or permanent impairment can occur if the user…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K122600Lifepak 1000 DefibrillatorPhysio-Control, Inc. · 2012 · Cites it directly · ◆ in the 2012 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 12 years and passes through 4 companies; a new generation was cleared every 2.1 years on median.
  • 12 of the 24 ancestors have a recall of any class.
  • Of its descendants, K122600 (Physio-Control, Inc., 2012) has gone furthest, with 0 descendants of its own.
  • 5 other devices also cite its predicate K983393.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

Defibrillation: Defibrillation is a recognized means of terminating certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. The defibrillator is to be used in AED mode only on patients who are in cardiopulmonary arrest. The patient must be unresponsive, not breathing normally, and showing no signs of circulation. The defibrillator may be used with standard defibrillation pads only on adults and…

Read the full statement (134 words)

Defibrillation:

Defibrillation is a recognized means of terminating certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia.

The defibrillator is to be used in AED mode only on patients who are in cardiopulmonary arrest. The patient must be unresponsive, not breathing normally, and showing no signs of circulation.

The defibrillator may be used with standard defibrillation pads only on adults and children who are 8 years old or more or who weigh more than 25 kg (551bs). The defibrillator may be used on children who are less than 8 years old or weigh less than 25 kg (55 lbs) with Infant/Child Reduced Energy Defibrillation Electrodes.

ECG Monitoring:

ECG monitoring is for use on conscious and unconscious patients of all ages for the purpose of ECG rhythm recognition and heart rate monitoring.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042404: Lifepak, Model 1000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K042404

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026