510(k) Report

510(k) · Decided

Recalled · 2007

Coolgard 3000, Cooline, Start-UP, Models 3000, CL-2085B, CL-2295A, CG-500D

K014241510(k) TraditionalDecided Aug 1, 2003583-day reviewClass II21 CFR 870.5900Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Alsius Corp. · NCX System, Hypothermia, Intravenous, Cooling

Cleared on Aug 1, 2003 as substantially equivalent to K001520 (Alpha Omega, Inc., 2000) and 3 other predicates after a 583-day review. Its family tree reaches back 2 generations to 1982. 3 of the 20 devices in its family, counting itself, have a Class I recall. 11 later devices descend from it. It has been recalled 6 times.

Decided
Aug 1, 2003
Review
583 days
Class
Class II
Product code
NCX
Recalls
6
Predicates
4

Its longest line of descent runs back 2 generations to K820648 (1982). 8 ancestors in all. 11 later devices descend from it, 5 citing it directly. 3 of the 20 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2000: 2 other ancestors1997: 1 other ancestor1990: 1 other ancestor, 1 with a Class I recall11986: 1 other ancestor, 1 with a Class I recall11985: 1 other ancestor2010: 1 descendant cleared2016: 2 descendants cleared2004: 1 descendant cleared2005: 1 descendant cleared2015: 1 descendant cleared2008: 1 descendant cleared2014: 1 descendant cleared2021: 1 descendant cleared2023: 1 descendant cleared2022: 1 descendant clearedK820648 · Arrow M.I.S. Multi-Lumen Catheteriza · Arrow Intl., Inc. · 1982 · recalledK970864 · Arrow-Howes Large Bore Multi-Lumen Central Venous Catheters · Arrow Intl., Inc. · 1997 · Class I recall19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors2 generations back: 3 other ancestors, 2 with a Class I recall2 1 generation forward: 5 descendants2 generations forward: 3 descendants3 generations forward: 3 descendants K970864 · Arrow-Howes Large Bore Multi-Lumen Central Venous Catheters · Arrow Intl., Inc. · 1997 · Class I recallK820648 · Arrow M.I.S. Multi-Lumen Catheteriza · Arrow Intl., Inc. · 1982 · recalled0
  1. Root · 2 generations backK820648 Arrow M.I.S. Multi-Lumen CatheterizaArrow Intl., Inc. · 1982 · Recalled
  2. 1 generation backK970864 Arrow-Howes Large Bore Multi-Lumen Central Venous CathetersArrow Intl., Inc. · 1997 · Recalled (Class I)
  3. This deviceK014241 Coolgard 3000, Cooline, Start-UP, Models 3000, CL-2085B, CL-2295A…Alsius Corp. · 2003
  4. Descendant · cites it directly · 6 of its ownK101987 Cool Line Catheter, Model CL-2295A, Icy Catheter, Model IC-3893A and…Zoll Circulation · 2010
  5. Descendant · cites it directly · 4 of its ownK150046 Start-Up Kit (SUK), Cool Line Intravascular Heat Exchange Catheter…Zoll Circulation, Inc. · 2015 · Recalled
  6. Descendant · 2 generations on · 4 of its ownK141139 Solex Intravascular Heat Exchange CatheterZoll Circulation · 2014
  7. Descendant · cites it directly · 3 of its ownK153226 Solex 7 Intravascular Heat Exchange Catheter, ZOLL Start-Up KitZoll Circulation, Inc. · 2016 · Recalled
  8. Descendant · 2 generations on · 2 of its ownK213031 Solex 7 Intravascular Heat Exchange Catheter, Cool Line…Zoll Circulation, Inc. · 2021

Showing 5 of 11 descendants. See all descendants

Descendants cleared per year

2004: 12005: 12008: 12010: 12014: 12015: 12016: 22021: 12022: 12023: 1
2004high 2 (2016) · low 0 (2006)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 11 · See all
  • K101987Cool Line Catheter, Model CL-2295A, Icy Catheter, Model IC-3893A and Quattro Catheter…Zoll Circulation · 2010 · cites it directly6descendants
  • K150046Start-Up Kit (SUK), Cool Line Intravascular Heat Exchange Catheter, ICY Intravascular…Zoll Circulation, Inc. · 2015 · cites it directly4descendants
  • K141139Solex Intravascular Heat Exchange CatheterZoll Circulation · 2014 · 2 generations on4descendants
  • K153226Solex 7 Intravascular Heat Exchange Catheter, ZOLL Start-Up KitZoll Circulation, Inc. · 2016 · cites it directly3descendants
  • K213031Solex 7 Intravascular Heat Exchange Catheter, Cool Line Intravascular Heat Exchange…Zoll Circulation, Inc. · 2021 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K820648Arrow M.I.S. Multi-Lumen CatheterizaArrow Intl., Inc.19827
−1K970864Arrow-Howes Large Bore Multi-Lumen Central Venous CathetersArrow Intl., Inc.1997Class I
0 (this device)K014241Coolgard 3000, Cooline, Start-UP, Models 3000, CL-2085B, CL-2295A, CG-500DAlsius Corp.2003—
+1K101987Cool Line Catheter, Model CL-2295A, Icy Catheter, Model IC-3893A and Quattro…Zoll Circulation2010—
+1K150046Start-Up Kit (SUK), Cool Line Intravascular Heat Exchange Catheter, ICY…Zoll Circulation, Inc.201521
+2K141139Solex Intravascular Heat Exchange CatheterZoll Circulation2014—
+1K153226Solex 7 Intravascular Heat Exchange Catheter, ZOLL Start-Up KitZoll Circulation, Inc.20164
+2K213031Solex 7 Intravascular Heat Exchange Catheter, Cool Line Intravascular Heat…Zoll Circulation, Inc.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198510
198611
199011
199710
200020

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200510
200810
201010
201410
201510
201620
202110
202210
202310
FIG. 1 — Lineage of K014241, 1982–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NCX device

  1. #1 K213031 Solex 7 Intravascular Heat Exchange Catheter…score 10.9 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K001520Alpha Omega Dual Cooler-HeaterAlpha Omega, Inc.2000—
*K001546ALLON 2001Mtre Advanced Technology, Ltd.2000—
*K962577Arrowgard Blue Quad-Lumen Central Venous Catheter
Summary page 2
…OOl 546; S.E. 06-08-2000 ARROWgard Blue Quad-Lumen Central Venous Catheter Arrow International, Inc., Reading PA K962577; S.E. 08-21-1997 Arrow-Howes Large Bore Multi Lumen central venous catheter Arrow Internati…
Arrow Intl., Inc.19971
*K970864Arrow-Howes Large Bore Multi-Lumen Central Venous Catheters
Summary page 2
…S.E. 08-21-1997 Arrow-Howes Large Bore Multi Lumen central venous catheter Arrow International, Inc., Reading PA K970864; S.E. 09-17-1997 indications for Use The Alsius CoolGard@3000 and Cool Line TM Catheter Thermal Regulation System is…
Arrow Intl., Inc.1997Class I

Also cleared against K001520: 1 other device →

What descends from it? (11)

See all 11

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K101987Cool Line Catheter, Model CL-2295A, Icy Catheter, Model IC-3893A and Quattro Catheter…Zoll Circulation20106
K150046Start-Up Kit (SUK), Cool Line Intravascular Heat Exchange Catheter, ICY Intravascular…Zoll Circulation, Inc.20154
K153226Solex 7 Intravascular Heat Exchange Catheter, ZOLL Start-Up KitZoll Circulation, Inc.20163
K033623Celsius Control SystemInnercool Therapies, Inc.20042
K051912Cool Line Catheter Kit, Models CL-2085B, CL-2295A; Icy Catheter Kit, Model IC-3585A…Alsius Corp.20050

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 1, 2007Not classifiedZ-0502-2007OtherTesting by Alsius has identified that 6-10% of catheters from the three affected lots may exhibit a significant reduction in their heat exchange capacity.
Jan 1, 2007Not classifiedZ-0503-2007OtherTesting by Alsius has identified that 6-10% of catheters from the three affected lots may exhibit a significant reduction in their heat exchange capacity.
Nov 9, 2005Not classifiedZ-0369-06OtherAlsius has found that fluid spills or ingress may result in an alarm condition. The potential health risk involved in this state can be the discontinuance of therapy.
Jul 25, 2005Not classifiedZ-1397-05OtherFirm wants users to be aware not to leave the units in standby mode for more than an hour in rooms in excess of 75 F due to risk of an alarm mode being generated due to overheating.
Nov 10, 2004Not classifiedZ-0258-05OtherProduct alarms that there may be a power supply interruption.
Jun 21, 2004Not classifiedZ-1159-04OtherHigher than expected rate of pump alarms occur under specific conditions.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K970864Arrow-Howes Large Bore Multi-Lumen Central Venous CathetersArrow Intl., Inc. · 1997 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K895417Arrow Large-Bore Dual Lumen Hemodialysis KitArrow Intl., Inc. · 1990 · 2 generations back · ◆ in the 1990 ancestors bar of FIG. 11 Class I recall
  • K862056[Central Venous][Multi-Lumen] Catheter W/Flex TipArrow Intl., Inc. · 1986 · 2 generations back · ◆ in the 1986 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 21 years and passes through 2 companies; a new generation was cleared every 15.5 years on median.
  • 5 of the 8 ancestors have a recall of any class; 3 have a Class I recall.
  • Of its descendants, K101987 (Zoll Circulation, 2010) has gone furthest, with 6 descendants of its own.
  • 1 other device also cites its predicate K001520.

How does its category compare?

All NCX devices

NCX System, Hypothermia, Intravenous, Cooling: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · NCX

1in 2023

2007: 62008: 32010: 12014: 12015: 12016: 32021: 12022: 12023: 1
2007high 6 (2007) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NCX

204in 2016

2007: 28 days median2008: 129 days median2016: 204 days median
2007high 204 (2016) · low 28 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NCX

28in 2021

2007: 22014: 12021: 28
2007high 28 (2021) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NCX

273in 2025

2007: 42008: 82009: 42010: 92011: 72012: 142013: 312014: 412015: 322016: 3112017: 4502018: 5092019: 4362020: 4152021: 4612022: 5112023: 3732024: 3432025: 273
2007high 511 (2022) · low 4 (2007)2025

MAUDE reports naming product code NCX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200762824
2008312908
20090—04
20101—09
20110—07
20120—014
20130—031
20141—141
20151—032
201632040311
20170—0450
20180—0509
20190—0436
20200—0415
20211—28461
20221—0511
20231—0373
20240—0343
20250—0273
20260—0—

What is it indicated for?

The Alsius CoolGardB 3000 and Cool Line Catheter Thermal Regulation System is indicated for use in fever reduction, as an adjunct to other antipyretic therapy, in patients with cerebral infarction and intracerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K014241: Coolgard 3000, Cooline, Start-UP, Models 3000, CL-2085B, CL-2295A, CG-500D.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K014241

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026