510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Stockert Heater-Cooler System 3T

K052601510(k) TraditionalCleared Jun 6, 2006258-day reviewClass II21 CFR 870.4250Recalls 18

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Sorin Group Deutschland GmbH · DWC Controller, Temperature, Cardiopulmonary Bypass

Cleared on Jun 6, 2006 as substantially equivalent to K811742 (Cincinnati Sub-Zero, 1981) and 2 other predicates after a 258-day review. Its family tree reaches back 2 generations to 1988. 4 later devices descend from it. It has been recalled 18 times.

Decided
Jun 6, 2006
Review
258 days
Class
Class II
Product code
DWC
Recalls
18
Predicates
3

Its longest line of descent runs back 2 generations to K883603 (1988). 4 ancestors in all, the oldest K811742 (1981). 4 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor1981: 1 other ancestor2020: 1 descendant cleared2014: 1 descendant cleared2022: 2 descendants clearedK883603 · Sarns Temp. Control and Monitor System (TCM II) · 3M Health Care, Sarns · 1988 · recalledK031544 · Heater Cooler Unit, Model Hcu 30 · Jostra AB · 2003 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 3 descendants2 generations forward: 1 descendant K031544 · Heater Cooler Unit, Model Hcu 30 · Jostra AB · 2003 · recalledK883603 · Sarns Temp. Control and Monitor System (TCM II) · 3M Health Care, Sarns · 1988 · recalled0
  1. Root · 2 generations backK883603 Sarns Temp. Control and Monitor System (TCM II)3M Health Care, Sarns · 1988 · Recalled
  2. 1 generation backK031544 Heater Cooler Unit, Model Hcu 30Jostra AB · 2003 · Recalled
  3. This deviceK052601 Stockert Heater-Cooler System 3TSorin Group Deutschland GmbH · 2006
  4. Descendant · cites it directly · 1 of its ownK191402 Heater-Cooler System 3TLivanova Deutschland, GmbH · 2020 · Recalled
  5. Descendant · 2 generations on · 0 of its ownK220635 Heater-Cooler System 3TLivanova Deutschland, GmbH · 2022
  6. Descendant · cites it directly · 0 of its ownK212657 Quantum Heater-CoolerSpectrum Medical, Ltd. · 2022
  7. Descendant · cites it directly · 0 of its ownK140012 Sorin FlexthermSorin Group Deutschland GmbH · 2014

Most influential descendants

Showing 4 of 4 · See all
  • K191402Heater-Cooler System 3TLivanova Deutschland, GmbH · 2020 · cites it directly1descendant
  • K220635Heater-Cooler System 3TLivanova Deutschland, GmbH · 2022 · 2 generations on0descendants
  • K212657Quantum Heater-CoolerSpectrum Medical, Ltd. · 2022 · cites it directly0descendants
  • K140012Sorin FlexthermSorin Group Deutschland GmbH · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K883603Sarns Temp. Control and Monitor System (TCM II)3M Health Care, Sarns19884
−1K031544Heater Cooler Unit, Model Hcu 30Jostra AB20034
0 (this device)K052601Stockert Heater-Cooler System 3TSorin Group Deutschland GmbH2006—
+1K191402Heater-Cooler System 3TLivanova Deutschland, GmbH20201
+2K220635Heater-Cooler System 3TLivanova Deutschland, GmbH2022—
+1K212657Quantum Heater-CoolerSpectrum Medical, Ltd.2022—
+1K140012Sorin FlexthermSorin Group Deutschland GmbH2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
202010
202220
FIG. 1 — Lineage of K052601, 1988–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DWC device

  1. #1 K220635 Heater-Cooler System 3Tscore 5.3 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K811742Hemotherm Model 400
Summary page 1
…oler Classification Name: Cardiopulmonary bypass temperature controller 3. PREDICATE DEVICE * Cincinnati Subzero Hemotherm (CSZ Hemotherm) (K811742) • Alpha Omega, Inc. Dual 2 Cooler-Heater (K001520) * Jostra AB Heater-Cooler Unit 30 (K031544) 4.…
Cincinnati Sub-Zero19812
*K001520Alpha Omega Dual Cooler-Heater
Summary page 1
…rature controller 3. PREDICATE DEVICE * Cincinnati Subzero Hemotherm (CSZ Hemotherm) (K811742) • Alpha Omega, Inc. Dual 2 Cooler-Heater (K001520) * Jostra AB Heater-Cooler Unit 30 (K031544) 4. DEVICE DESCRIPTION The Sorin Group Deutschland GmbH…
Alpha Omega, Inc.2000—
*K031544Heater Cooler Unit, Model Hcu 30
Summary page 1
…* Cincinnati Subzero Hemotherm (CSZ Hemotherm) (K811742) • Alpha Omega, Inc. Dual 2 Cooler-Heater (K001520) * Jostra AB Heater-Cooler Unit 30 (K031544) 4. DEVICE DESCRIPTION The Sorin Group Deutschland GmbH St6ckert Heater-Cooler System 3T consists of s…
Jostra AB20034

Also cleared against K811742: 4 other devices →

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K191402Heater-Cooler System 3TLivanova Deutschland, GmbH20201
K212657Quantum Heater-CoolerSpectrum Medical, Ltd.20220
K140012Sorin FlexthermSorin Group Deutschland GmbH20140

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 24, 2020Class IIZ-1498-2020Device DesignThe firm is providing validated cleaning and disinfection instructions in their revised operating instructions to reduce the risk of NTM contamination.
Nov 1, 2018Class IIZ-0220-2020Device DesignThe firm has become aware that due to a chemical reaction, the disinfectant/water preservative used to disinfect the device to inhibit/limit the growth micro-organisms, may be rendered ineffective. In addition, the firm is releasing a…
Sep 14, 2018Class IIZ-1099-2019Material/Component ContaminationFor six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).
Apr 20, 2018Class IIZ-1901-2018Device DesignFirm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm.
Jun 30, 2017Class IIZ-2776-2017Device DesignPlastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
Jun 30, 2017Class IIZ-2775-2017Device DesignPlastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
Jun 30, 2017Class IIZ-2773-2017Device DesignPlastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
Jun 30, 2017Class IIZ-2772-2017Device DesignPlastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
Jun 30, 2017Class IIZ-2771-2017Device DesignPlastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
Jun 30, 2017Class IIZ-2774-2017Device DesignPlastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
Jun 15, 2015Class IIZ-2081-2015Error in labelingPotential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.
Jun 15, 2015Class IIZ-2079-2015Error in labelingPotential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.
Jun 15, 2015Class IIZ-2078-2015Error in labelingPotential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.
Jun 15, 2015Class IIZ-2080-2015Error in labelingPotential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.
Jun 15, 2015Class IIZ-2076-2015Error in labelingPotential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.
Jun 15, 2015Class IIZ-2077-2015Error in labelingPotential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.
Aug 1, 2014Class IIZ-0682-2017Material/Component ContaminationOn August 1, 2014 Sorin Group USA, Inc. informed consignees via letter to review the disinfection and monitoring practices and to strictly adhere to the instructions in the Instructions for Use (IFU).
Aug 15, 2012Class IIZ-1186-2016Device DesignOn August 15, 2012, Sorin informed consignees via letter of changes made to create Version 12 of the Heater-Cooler 3T IFU. The revised IFU included enhancements to achieve water quality that meets the standard for German drinking water.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 3 companies; a new generation was cleared every 14.8 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • Of its descendants, K191402 (Livanova Deutschland, GmbH, 2020) has gone furthest, with 1 descendant of its own.
  • 4 other devices also cite its predicate K811742.

How does its category compare?

All DWC devices

DWC Controller, Temperature, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DWC

2in 2022

2007: 12010: 12012: 12013: 12014: 12020: 12022: 2
2007high 2 (2022) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DWC

Too few DWC 510(k)s in any year since 2007 to give a median.

Recalls per year · DWC

5in 2023

2010: 12012: 22014: 12015: 12016: 12017: 12018: 32021: 52023: 5
2007high 5 (2021) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DWC

670in 2025

2007: 32009: 22010: 2432011: 2812012: 3512013: 2442014: 3782015: 5752016: 7872017: 7162018: 5642019: 6972020: 5482021: 6222022: 6052023: 9282024: 5552025: 670
2007high 928 (2023) · low 0 (2008)2025

MAUDE reports naming product code DWC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—03
20080—00
20090—02
20101—1243
20110—0281
20121—2351
20131—0244
20141—1378
20150—1575
20160—1787
20170—1716
20180—3564
20190—0697
20201—0548
20210—5622
20222—0605
20230—5928
20240—0555
20250—0670
20260—0—

What is it indicated for?

The Stbckert Heater-Cooler System 3T is used with a St6ckert S3 heart-lung machine and/or any other heart lung machine featuring a separate temperature control for extracorporeal perfusion of durations of up to 6 hours.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052601: Stockert Heater-Cooler System 3T.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K052601

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026