510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

Medtronic Mida Rex Legend System, Midas Rex Legend Pneumatic High Speed System or Legend System

K020069510(k) TraditionalCleared Mar 18, 200268-day reviewClass II21 CFR 882.4370Recalls 9

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Midas Rex · HBB Motor, Drill, Pneumatic

Cleared on Mar 18, 2002 after a 68-day review. The FDA summary names its predicate without a number. 43 later devices descend from it. It has been recalled 9 times.

Decided
Mar 18, 2002
Review
68 days
Class
Class II
Product code
HBB
Recalls
9
Predicates
0

43 later devices descend from it, 11 citing it directly.

Descendants, by year cleared 2009: 1 descendant cleared2002: 1 descendant cleared2017: 4 descendants cleared2007: 2 descendants cleared2005: 1 descendant cleared2010: 1 descendant cleared2021: 4 descendants cleared2003: 1 descendant cleared2004: 1 descendant cleared2006: 1 descendant cleared2015: 2 descendants cleared2024: 6 descendants cleared2016: 1 descendant cleared2019: 4 descendants cleared2026: 5 descendants cleared2022: 3 descendants cleared2020: 1 descendant cleared2023: 2 descendants cleared2025: 2 descendants cleared200020052010201520202025
  1. This deviceK020069 Medtronic Mida Rex Legend System, Midas Rex Legend Pneumatic High…Medtronic Midas Rex · 2002
  2. Descendant · cites it directly · 20 of its ownK090112 Midas Rex MR7 Pneumatic High Speed System, Models PM700, PM710…Medtronic Powered Surgical Solutions · 2009
  3. Descendant · cites it directly · 13 of its ownK072315 Midas Rex Curved BurMedtronic Powered Surgical Solutions · 2007
  4. Descendant · cites it directly · 11 of its ownK170312 IPC systems, Legend EHS handpieces and Legend Stylus Touch…Medtronic Powered Surgical Solutions · 2017
  5. Descendant · cites it directly · 10 of its ownK163182 Medtronic Legend Pneumatic; MR7 Pneumatic; Triton Pneumatic Drill…Medtronic Powered Surgical Solutions · 2017
  6. Descendant · 2 generations on · 9 of its ownK160713 Stealth-Midas SystemMedtronic Powered Surgical Solutions · 2016

Showing 5 of 43 descendants. See all descendants

Descendants cleared per year

2002: 12003: 12004: 12005: 12006: 12007: 22009: 12010: 12015: 22016: 12017: 42019: 42020: 12021: 42022: 32023: 22024: 62025: 22026: 5
2002high 6 (2024) · low 0 (2008)2026

Most influential descendants

Showing 5 of 43 · See all
  • K090112Midas Rex MR7 Pneumatic High Speed System, Models PM700, PM710, PC700, PC710, PC720, PA700Medtronic Powered Surgical Solutions · 2009 · cites it directly20descendants
  • K072315Midas Rex Curved BurMedtronic Powered Surgical Solutions · 2007 · cites it directly13descendants
  • K170312IPC systems, Legend EHS handpieces and Legend Stylus Touch handpiece, Microsaw…Medtronic Powered Surgical Solutions · 2017 · cites it directly11descendants
  • K163182Medtronic Legend Pneumatic; MR7 Pneumatic; Triton Pneumatic Drill System incorporating…Medtronic Powered Surgical Solutions · 2017 · cites it directly10descendants
  • K160713Stealth-Midas SystemMedtronic Powered Surgical Solutions · 2016 · 2 generations on9descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K020069Medtronic Mida Rex Legend System, Midas Rex Legend Pneumatic High Speed System…Medtronic Midas Rex2002—
+1K090112Midas Rex MR7 Pneumatic High Speed System, Models PM700, PM710, PC700, PC710…Medtronic Powered Surgical Solutions2009—
+1K072315Midas Rex Curved BurMedtronic Powered Surgical Solutions2007—
+1K170312IPC systems, Legend EHS handpieces and Legend Stylus Touch handpiece, Microsaw…Medtronic Powered Surgical Solutions2017—
+1K163182Medtronic Legend Pneumatic; MR7 Pneumatic; Triton Pneumatic Drill System…Medtronic Powered Surgical Solutions2017—
+2K160713Stealth-Midas SystemMedtronic Powered Surgical Solutions2016—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200210
200310
200410
200510
200610
200720
200910
201010
201520
201610
201740
201940
202010
202140
202230
202320
202460
202520
202650
FIG. 1 — Lineage of K020069, 2002–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HBB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HBB device

  1. #1 K253772 Velocity Alpha MR High Speed Surgical Drill…score 11.0 · 1 generation · no Class I recall
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Midas Rex Classic
Named without a submission number
—
*—Midas Rex III Instrumentation System
Named without a submission number
—
*—Mednext 1000 Bone Dissecting System
Named without a submission number
—

What descends from it? (43)

See all 43

11 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K090112Midas Rex MR7 Pneumatic High Speed System, Models PM700, PM710, PC700, PC710, PC720, PA700Medtronic Powered Surgical Solutions200920
K072315Midas Rex Curved BurMedtronic Powered Surgical Solutions200713
K170312IPC systems, Legend EHS handpieces and Legend Stylus Touch handpiece, Microsaw…Medtronic Powered Surgical Solutions201711
K163182Medtronic Legend Pneumatic; MR7 Pneumatic; Triton Pneumatic Drill System incorporating…Medtronic Powered Surgical Solutions201710
K041754Stryker Maestro SystemStryker Corp.20048
K163565MR8 Drill SystemMedtronic Powered Surgical Solutions20177
K032607Powerpro Pneumatic System, Models PRO6150, PRO6175, PRO6185Linvatec Corp.20036
K193630hekaDrillZethon, Ltd.20210
K093361Synthes Air Pen Drive (Apd) SystemSynthes (Usa)20100
K042974Medicon Vipair High Speed SystemMedicon EG20050

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 8, 2016Class IIZ-1602-2016Labeling mix-upsIncorrect color coding and attachment reference were included on the labeling of 10BA40D dissecting tools with lot number H5237248.
Oct 21, 2015Class IIZ-0321-2016Process controlThe wire pass hole is not present on the tools.
Nov 30, 2009Not classifiedZ-0546-2010Nonconforming Material/ComponentComponent of product attachment may be outside established tolerance with potential for footed portion to break during normal use.
May 30, 2008Not classifiedZ-2156-2008Process controlCutting flute geometry of surgical dissecting tools may be outside of established tolerance. This may lead to reduced cutting performance, inability to perform desired dissection, increased risk of overheating, and potential need for…
Sep 1, 2005Not classifiedZ-1574-05OtherPotential compromise of sterile packaging for disposable dissecting tools.
Jun 29, 2005Not classifiedZ-1067-05OtherComplaints of finger control component breaking off and falling into open surgical wound.
Nov 11, 2004Not classifiedZ-0278-05OtherSofter hose increases the propensity for outer exhaust hose to kink or become occluded as a result of clamping. Complaints of hose rupture/burst have been reported.
Nov 3, 2004Not classifiedZ-0213-05OtherReduced plastic thickness of lubricant/diffuser cartidge can crack under pressure causing leakage of lubricant as visible mist (appearance of smoke).
Nov 11, 2002Not classifiedZ-0419-03OtherLubricant leaking from motor after sterilization causing bleed-through sterilization wrap and/or excess lubricant in instrument case.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K090112 (Medtronic Powered Surgical Solutions, 2009) has gone furthest, with 20 descendants of its own.

How does its category compare?

All HBB devices

HBB Motor, Drill, Pneumatic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · HBB

1in 2018

2007: 12009: 12010: 22015: 12017: 12018: 12026 (partial year): 1
2007high 2 (2010) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HBB

Too few HBB 510(k)s in any year since 2007 to give a median.

Recalls per year · HBB

2in 2025

2007: 12008: 12009: 12013: 12015: 12017: 12025: 2
2007high 2 (2025) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HBB

1,016in 2025

2007: 822008: 1762009: 1962010: 1832011: 3992012: 3792013: 6852014: 5622015: 4402016: 2732017: 2032018: 2952019: 2842020: 2632021: 2602022: 1722023: 1802024: 7592025: 1,016
2007high 1,016 (2025) · low 82 (2007)2025

MAUDE reports naming product code HBB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—182
20080—1176
20091—1196
20102—0183
20110—0399
20120—0379
20130—1685
20140—0562
20151—1440
20160—0273
20171—1203
20181—0295
20190—0284
20200—0263
20210—0260
20220—0172
20230—0180
20240—0759
20250—21,016
20261—0—

What is it indicated for?

The Medtronic Midas Rex Legend System is a pneumatically operated surgical instrument system. The pneumatic motor provides power to operate removable rotating surgical cutting tools and their accessories intended for use in neurosurgery, including craniotomy and spinal surgery; as well as Ear Nose and Throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial and sternotomy surgeries.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K020069: Medtronic Mida Rex Legend System, Midas Rex Legend Pneumatic High Speed System or Legend System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K020069

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026