Mahurkar Triple Lumen Catheter, 12 FR

K020089510(k) TraditionalCleared Apr 10, 200290-day reviewClass II21 CFR 876.5540Recalls 26 (25 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

The Kendall Company · NIE Catheter, Hemodialysis, Triple Lumen, Non-Implanted

Cleared on Apr 10, 2002 after a 90-day review. The FDA summary names its predicate without a number. 5 of the 16 devices in its family, counting itself, have a Class I recall. 15 later devices descend from it. It has been recalled 26 times, including 25 Class I recalls.

Decided
Apr 10, 2002
Review
90 days
Class
Class II
Product code
NIE
Recalls
26 (25 Class I)
Predicates
0

15 later devices descend from it, 4 citing it directly. 5 of the 16 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2009: 1 descendant cleared2004: 1 descendant cleared2010: 1 descendant cleared, 1 with a Class I recall12020: 2 descendants cleared2013: 1 descendant cleared2018: 1 descendant cleared2015: 2 descendants cleared, 2 with a Class I recall22014: 1 descendant cleared, 1 with a Class I recall12012: 1 descendant cleared2023: 2 descendants cleared2019: 1 descendant cleared2016: 1 descendant cleared200020052010201520202025
  1. This deviceK020089 Mahurkar Triple Lumen Catheter, 12 FRThe Kendall Company · 2002
  2. Descendant · cites it directly · 12 of its ownK083675 Power-Trialysis Triple Lumen Acute Dialysis CatheterC.R. Bard, Inc. · 2009
  3. Descendant · cites it directly · 7 of its ownK102605 Mahurkar Triple Lumen Dialysis CatheterCovidien, LLC, Formerly Registered AS Tyco Healthca · 2010 · Recalled (Class I)
  4. Descendant · cites it directly · 4 of its ownK033570 Medcomp T-3Medcomp · 2004
  5. Descendant · 2 generations on · 2 of its ownK133456 Power-Trialysis Short-Term Dialysis CatheterC.R. Bard, Inc. · 2014 · Recalled (Class I)
  6. Descendant · 2 generations on · 2 of its ownK123292 T3Medcomp · 2013

Showing 5 of 15 descendants. See all descendants

Descendants cleared per year

2004: 12009: 12010: 12012: 12013: 12014: 12015: 22016: 12018: 12019: 12020: 22023: 2
2004high 2 (2015) · low 0 (2005)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 15 · See all
  • K083675Power-Trialysis Triple Lumen Acute Dialysis CatheterC.R. Bard, Inc. · 2009 · cites it directly12descendants
  • K102605Mahurkar Triple Lumen Dialysis CatheterCovidien, LLC, Formerly Registered AS Tyco Healthca · 2010 · cites it directly7descendants
  • K033570Medcomp T-3Medcomp · 2004 · cites it directly4descendants
  • K133456Power-Trialysis Short-Term Dialysis CatheterC.R. Bard, Inc. · 2014 · 2 generations on2descendants
  • K123292T3Medcomp · 2013 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K020089Mahurkar Triple Lumen Catheter, 12 FRThe Kendall Company2002—
+1K083675Power-Trialysis Triple Lumen Acute Dialysis CatheterC.R. Bard, Inc.2009—
+1K102605Mahurkar Triple Lumen Dialysis CatheterCovidien, LLC, Formerly Registered AS Tyco Healthca2010Class I
+1K033570Medcomp T-3Medcomp2004—
+2K133456Power-Trialysis Short-Term Dialysis CatheterC.R. Bard, Inc.2014Class I
+2K123292T3Medcomp2013—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200910
201011
201210
201310
201411
201522
201610
201810
201910
202020
202320
FIG. 1 — Lineage of K020089, 2002–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NIE device

  1. #1 K232945 Trio-CT­® Triple Lumen Catheter w/ Curved…score 9.4 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—MAHURKAR® Triple Lumen Catheter for Infusion
Named without a submission number
—
*—MAHURKAR® Dual Lumen Catheter
Named without a submission number
—

What descends from it? (15)

See all 15

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K083675Power-Trialysis Triple Lumen Acute Dialysis CatheterC.R. Bard, Inc.200912
K102605Mahurkar Triple Lumen Dialysis Catheter Class ICovidien, LLC, Formerly Registered AS Tyco Healthca20107
K033570Medcomp T-3Medcomp20044
K192302Mahurkar Acute Single Lumen Catheter, Mahurkar Acute Dual Lumen Catheter, Mahurkar Acute…Covidien, LLC20200

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 28, 2023Class IZ-2377-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2367-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2382-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2386-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2381-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2372-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2370-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2389-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2383-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2369-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2373-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2385-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2378-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2374-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2375-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2376-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2387-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2368-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2391-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2390-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2380-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2388-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2371-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2384-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Jun 28, 2023Class IZ-2379-2023Process controlCatheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or…
Apr 29, 2019Class IIZ-1955-2019Error in labelingThe priming volume values printed on the MAHURKA and Argyle acute catheters and Instructions for Use (IFU) are higher than the volumes required to fill each lumen.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K102605Mahurkar Triple Lumen Dialysis CatheterCovidien, LLC, Formerly Registered AS Tyco Healthca · 2010 · Cites it directly · ◆ in the 2010 descendants bar of FIG. 125 Class I recalls
  • K143102Multi-Lumen Acute Hemodialysis Catheter for High Volume InfusionsArrow International, Inc. (Subsidiary of Teleflex, Inc.) · 2015 · 2 generations on · ◆ in the 2015 descendants bar of FIG. 11 Class I recall
  • K133456Power-Trialysis Short-Term Dialysis CatheterC.R. Bard, Inc. · 2014 · 2 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall
  • K141531Power-Trialysis Slim-Cath Short-Term Dialysis CatheterC.R. Bard, Inc. · 2015 · 3 generations on · ◆ in the 2015 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K083675 (C.R. Bard, Inc., 2009) has gone furthest, with 12 descendants of its own.

How does its category compare?

All NIE devices

NIE Catheter, Hemodialysis, Triple Lumen, Non-Implanted: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · NIE

2in 2023

2009: 12010: 12013: 12014: 12015: 22018: 12019: 12023: 2
2007high 2 (2015) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NIE

Too few NIE 510(k)s in any year since 2007 to give a median.

Recalls per year · NIE

6in 2025

2019: 12022: 1, 1 Class I2025: 62026 (partial year): 3, 1 Class I
2007high 6 (2025) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · NIE

72in 2025

2008: 22009: 32010: 52011: 122012: 92013: 72014: 132015: 72016: 132017: 382018: 882019: 612020: 512021: 612022: 1192023: 1032024: 1252025: 72
2007high 125 (2024) · low 0 (2007)2025

MAUDE reports naming product code NIE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—02
20091—03
20101—05
20110—012
20120—09
20131—07
20141—013
20152—07
20160—013
20170—038
20181—088
20191—161
20200—051
20210—061
20220—1 (1)119
20232—0103
20240—0125
20250—672
20260—3 (1)—

What is it indicated for?

Mahurkar® Triple Lumen Catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K020089: Mahurkar Triple Lumen Catheter, 12 FR.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K020089

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026