510(k) Report

510(k) · Cleared

Recalled · Class I · 2022

Mahurkar Oplus Catheter, Model 13.5 FR

K030209510(k) TraditionalCleared May 13, 2003112-day reviewClass II21 CFR 876.5540Recalls 4 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

The Kendall Company · MPB Catheter, Hemodialysis, Non-Implanted

Cleared on May 13, 2003 after a 112-day review. The FDA summary names its predicate without a number. 1 of the 6 devices in its family, counting itself, has a Class I recall. 5 later devices descend from it. It has been recalled 4 times, including a Class I recall.

Decided
May 13, 2003
Review
112 days
Class
Class II
Product code
MPB
Recalls
4 (1 Class I)
Predicates
0

5 later devices descend from it, 3 citing it directly. 1 of the 6 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2012: 1 descendant cleared2020: 2 descendants cleared2007: 1 descendant cleared2011: 1 descendant cleared200020052010201520202025
  1. This deviceK030209 Mahurkar Oplus Catheter, Model 13.5 FRThe Kendall Company · 2003
  2. Descendant · cites it directly · 1 of its ownK120674 Mahurkar Elite Acute Dual Lumen Catheter, Elite Acute Triple Lumen…Covidien · 2012 · Recalled
  3. Descendant · cites it directly · 1 of its ownK063355 D-Line Catheter ST, Double Lumen ST; Extra FlowNovosci Corp. · 2007
  4. Descendant · 2 generations on · 0 of its ownK200426 Acute Dual Lumen Hemodialysis CatheterHealth Line International Corporation · 2020
  5. Descendant · cites it directly · 0 of its ownK192302 Mahurkar Acute Single Lumen Catheter, Mahurkar Acute Dual Lumen…Covidien, LLC · 2020 · Recalled
  6. Descendant · 2 generations on · 0 of its ownK102833 Disposable Hemodialysis Access Catheter SetsFoshan Nanhai Bai HE Medical Technology Co., Ltd. · 2011

Most influential descendants

Showing 5 of 5 · See all
  • K120674Mahurkar Elite Acute Dual Lumen Catheter, Elite Acute Triple Lumen CatheterCovidien · 2012 · cites it directly1descendant
  • K063355D-Line Catheter ST, Double Lumen ST; Extra FlowNovosci Corp. · 2007 · cites it directly1descendant
  • K200426Acute Dual Lumen Hemodialysis CatheterHealth Line International Corporation · 2020 · 2 generations on0descendants
  • K192302Mahurkar Acute Single Lumen Catheter, Mahurkar Acute Dual Lumen Catheter, Mahurkar Acute…Covidien, LLC · 2020 · cites it directly0descendants
  • K102833Disposable Hemodialysis Access Catheter SetsFoshan Nanhai Bai HE Medical Technology Co., Ltd. · 2011 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K030209Mahurkar Oplus Catheter, Model 13.5 FRThe Kendall Company2003—
+1K120674Mahurkar Elite Acute Dual Lumen Catheter, Elite Acute Triple Lumen CatheterCovidien20122
+1K063355D-Line Catheter ST, Double Lumen ST; Extra FlowNovosci Corp.2007—
+2K200426Acute Dual Lumen Hemodialysis CatheterHealth Line International Corporation2020—
+1K192302Mahurkar Acute Single Lumen Catheter, Mahurkar Acute Dual Lumen Catheter…Covidien, LLC20206
+2K102833Disposable Hemodialysis Access Catheter SetsFoshan Nanhai Bai HE Medical Technology Co., Ltd.2011—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
201110
201210
202020
FIG. 1 — Lineage of K030209, 2003–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MPB device

  1. #1 K250836 14F Duo-Flow® Side X Side Double Lumen Catheterscore 10.1 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—MAHURKARB Triple Lumen Catheter
Named without a submission number
—
*—MAHURKARB Dual Lumen Catheter
Named without a submission number
—

What descends from it? (5)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K120674Mahurkar Elite Acute Dual Lumen Catheter, Elite Acute Triple Lumen CatheterCovidien20121
K063355D-Line Catheter ST, Double Lumen ST; Extra FlowNovosci Corp.20071
K192302Mahurkar Acute Single Lumen Catheter, Mahurkar Acute Dual Lumen Catheter, Mahurkar Acute…Covidien, LLC20200

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 7, 2022Class IZ-0877-2023Process designThere is a potential leaking condition within the hub of affected devices, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
Apr 29, 2019Class IIZ-1955-2019Error in labelingThe priming volume values printed on the MAHURKA and Argyle acute catheters and Instructions for Use (IFU) are higher than the volumes required to fill each lumen.
Dec 2, 2003Not classifiedZ-0233-04OtherProduct is mislabeled to catheter length. Product labeled as a 13.5 Fr x 19.5 cm may contain a 13.5Fr x 16.0 cm
Dec 2, 2003Not classifiedZ-0234-04OtherProduct is mislabeled to catheter length. Product labeled as a 13.5 Fr x 19.5 cm may contain a 13.5 Fr x 16.0 cm

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K120674 (Covidien, 2012) has gone furthest, with 1 descendant of its own.

How does its category compare?

All MPB devices

MPB Catheter, Hemodialysis, Non-Implanted: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MPB

1in 2025

2007: 32009: 12010: 12011: 22012: 12013: 12016: 12017: 22020: 32025: 1
2007high 3 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MPB

205in 2020

2007: 136 days median2020: 205 days median
2007high 205 (2020) · low 136 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MPB

4so far in 2026, the most since 2007

2007: 12017: 22026 (partial year): 4, 3 Class I
2007high 4 (2026) · low 0 (2008)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · MPB

187in 2025

2007: 412008: 1472009: 1662010: 602011: 742012: 782013: 1242014: 822015: 512016: 702017: 862018: 1202019: 1192020: 1792021: 1922022: 1662023: 2032024: 1812025: 187
2007high 203 (2023) · low 41 (2007)2025

MAUDE reports naming product code MPB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20073136141
20080—0147
20091—0166
20101—060
20112—074
20121—078
20131—0124
20140—082
20150—051
20161—070
20172—286
20180—0120
20190—0119
202032050179
20210—0192
20220—0166
20230—0203
20240—0181
20251—0187
20260—4 (3)—

What is it indicated for?

MAHURKARm QPLUSB Catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K030209: Mahurkar Oplus Catheter, Model 13.5 FR.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K030209

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026