510(k) Report

510(k) · Cleared

Recalled · 2008

Deltec Insulin Infusion Pump and Accessories

K020655510(k) TraditionalCleared Aug 13, 2002165-day reviewClass II21 CFR 880.5725Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Deltec, Inc. · LZG Pump, Infusion, Insulin

Cleared on Aug 13, 2002 as substantially equivalent to K990801 (Medtronic Minimed, 1999) and 3 other predicates after a 165-day review. Its family tree reaches back 1 generation to 1998. 1 of the 5 devices in its family, counting itself, has a Class I recall. It has been recalled 6 times.

Decided
Aug 13, 2002
Review
165 days
Class
Class II
Product code
LZG
Recalls
6
Predicates
4

Its longest line of descent runs back 1 generation to K973917 (1998). 4 ancestors in all. No later device cites it as a predicate yet. 1 of the 5 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year cleared 2000: 1 other ancestor1999: 2 other ancestors, 1 with a Class I recall1K973917 · Cadd-Diplomat PC Communications System and the Cadd- Ambassador Pump Communications System · Sims Deltec, Inc. · 1998 · recalled19961998200020022004
Other ancestors, by generation 1 generation back: 3 other ancestors, 1 with a Class I recall1 K973917 · Cadd-Diplomat PC Communications System and the Cadd- Ambassador Pump Communications System · Sims Deltec, Inc. · 1998 · recalled0
  1. Root · 1 generation backK973917 Cadd-Diplomat PC Communications System and the Cadd- Ambassador Pump…Sims Deltec, Inc. · 1998 · Recalled
  2. This deviceK020655 Deltec Insulin Infusion Pump and AccessoriesDeltec, Inc. · 2002

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K973917Cadd-Diplomat PC Communications System and the Cadd- Ambassador Pump…Sims Deltec, Inc.19981
0 (this device)K020655Deltec Insulin Infusion Pump and AccessoriesDeltec, Inc.2002—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199921
200010
FIG. 1 — Lineage of K020655, 1998–2002Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZG device

  1. #1 K253534 Pivot Insulin Delivery Systemscore 16.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K990801Minimed 508 Insulin Pump
Summary page 2
…p. IV. DEVICE COMPARISON The Deltec CozmoTMInsulin Infbsion Pump is similar in design, fhnction, and intended use to the MiniMed Model 508 Insulin Pump (K990801). The pumps are small, ambulatory, battery-powered, microprocessor controlled, external insulin syringe pump delivery…
Medtronic Minimed1999Class I
*K9919363.0 ML Reservoir, Model 103/193
Summary page 2
…, IR windows, and 3-ml medication cartridge. The Deltec Cozmom 3-ml Cartridge is similar in design, hnction, and intended use to the MiniMed 3-ml Reservoir 6991936). Each device is a syringe style reservoir with a capacity of 3 ml. They are made of the same materials and both are provid…
Medtronic Minimed2000—
*K973917Cadd-Diplomat PC Communications System and the Cadd- Ambassador Pump Communications System
Summary page 2
…p. The Deltec CozmoTMTreatment AssistanP PC Communications System is similar in design, fbnction, and intended use to the Deltec CADD-Diplomat@System (K9739 17). The products allow communications between a personal computer (PC) and an inhsion pump. The software programs allow for d…
Sims Deltec, Inc.19981
*K993012Minimed Com-Station, Hardware-7301; Software-7311
Summary page 2
…mming of the pump via a PC. The Deltec Cornom Treatment Assistantm PC Communications System is also similar to the MiniMed Com-Stationm Communication System 6993012), which uses IR communication. 510(k) Summary of Safety and Effec…
Medtronic Minimed1999—

Also cleared against K990801: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 10, 2008Not classifiedZ-1622-2008Component change controlSmith Medical discovered an issue with a motor component in specific serial numbers of the Deltec Cozmo Insulin Pump, Model Numbers 1700 and 1800. This issue affects how the motor operates and may cause an over-delivery of insulin which…
Feb 22, 2008Not classifiedZ-1590-2008Device DesignSmith Medical has received reports of Deltec Cozmo Insulin Pumps powering down without an alarm. Investigation concluded that if the battery cap is not fully tightened, the Pump may power down and a brief "chirp" will sound. This may…
Feb 22, 2008Not classifiedZ-1591-2008Device DesignSmith Medical has received reports of Deltec Cozmo Insulin Pumps powering down without an alarm. Investigation concluded that if the battery cap is not fully tightened, the Pump may power down and a brief "chirp" will sound. This may…
Oct 15, 2004Not classifiedZ-0566-05OtherSmiths Medical MD, Inc. (SMMD) has received several complaints relating to two lots of Cozmo Cartridges describing leakage of insulin into the cartridge chamber. Firm expanded recall 03/17/05 to include one more lot, based on in-house…
Sep 18, 2004Not classifiedZ-0210-05OtherThe vibratory alarm motor installed in a small number of Deltec Cozmo Insulin Pumps may function intermittently.
Mar 14, 2003Not classifiedZ-0798-03OtherA software error could have resulted in the delivery of a larger bolus of insulin than anticipated when the pump was programmed in a particular manner.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K990801Minimed 508 Insulin PumpMedtronic Minimed · 1999 · Direct predicate · ◆ in the 1999 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • 2 of the 4 ancestors have a recall of any class; 1 has a Class I recall.
  • 3 other devices also cite its predicate K990801.

How does its category compare?

All LZG devices

LZG Pump, Infusion, Insulin: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LZG

2in 2024

2007: 32008: 62009: 22010: 22011: 42012: 32013: 22014: 42015: 12016: 42017: 12018: 12019: 22021: 12022: 12023: 12024: 22026 (partial year): 2
2007high 6 (2008) · low 0 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZG

116in 2016

2007: 112 days median2008: 129 days median2011: 189 days median2012: 73 days median2014: 123 days median2016: 116 days median
2007high 189 (2011) · low 73 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZG

1in 2022

2007: 12008: 132009: 22010: 42011: 22012: 42013: 3, 1 Class I2014: 2, 1 Class I2015: 2, 2 Class I2018: 3, 2 Class I2019: 52022: 1, 1 Class I2026 (partial year): 2, 2 Class I
2007high 13 (2008) · low 0 (2016)2026

Class I (9 since 2007) Class II/III not classified

Adverse event reports per year · LZG

12,004in 2025

2007: 5,1102008: 5,7112009: 9,4712010: 10,1752011: 12,8582012: 18,3672013: 30,0592014: 49,1882015: 69,1362016: 56,5812017: 44,4722018: 32,6522019: 29,0272020: 32,5882021: 31,5222022: 28,2252023: 21,9722024: 18,2282025: 12,004
2007high 69,136 (2015) · low 5,110 (2007)2025

MAUDE reports naming product code LZG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007311215,110
20086129135,711
20092—29,471
20102—410,175
20114189212,858
2012373418,367
20132—3 (1)30,059
201441232 (1)49,188
20151—2 (2)69,136
20164116056,581
20171—044,472
20181—3 (2)32,652
20192—529,027
20200—032,588
20211—031,522
20221—1 (1)28,225
20231—021,972
20242—018,228
20250—012,004
20262—2 (2)—

What is it indicated for?

The Deltec Cozmo Insulin Infusion Pump Model 1700 is a syringe infusion pump designed for Continuous Subcutaneous Insulin Infusion (CSII) for the control of diabetes.

The Deltec Cozmo 3-ml Cartridge is designed for use with the Deltec Cozmo Insulin Infusion Pump for Continuous Subcutaneous Insulin Infusion (CSII).

The Deltec Cozmo Treatment Assistant PC Communications System is designed for use with Deltec Cozmo Insulin Infusion Pump as a tool in optimizing diabetes management. It allows communications between the pump and a computer for accessing data stored in pump memory, and programming the pump.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K020655: Deltec Insulin Infusion Pump and Accessories.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K020655

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026