510(k) · Cleared
Recalled · Class II · 2019Gripper Plus Needle
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Deltec, Inc. · FPA Set, Administration, Intravascular
Cleared on Aug 13, 2002 after a 55-day review. The FDA summary names its predicate without a number. 21 later devices descend from it. It has been recalled once.
- Decided
- Aug 13, 2002
- Review
- 55 days
- Class
- Class II
- Product code
- FPA
- Recalls
- 1
- Predicates
- 0
21 later devices descend from it, 2 citing it directly.
- This deviceK021999 Gripper Plus NeedleDeltec, Inc. · 2002
- Descendant · cites it directly · 18 of its ownK070116 Power Port-a-Cath and Power Port-a-Cath II; Implantable Venous…Smiths Medical MD, Inc. · 2007
- Descendant · 2 generations on · 13 of its ownK073050 Safestep Max Power-Injectable Infusion SetSpecialized Health Products International, Inc. · 2008 · Recalled
- Descendant · 3 generations on · 6 of its ownK170897 Surecan Safety II Needle, Surecan Safety II Needle, Surecan Safety…B.Braun Medical, Inc. · 2017
- Descendant · 4 generations on · 2 of its ownK230715 Promisemed Safety Huber NeedlePromisemed Hangzhou Meditech Co., Ltd. · 2023
- Descendant · 4 generations on · 2 of its ownK200463 Huber Needle Infusion Set, Safety Huber Needle Infusion SetJiangsu Caina Medical Co.,Ltd · 2021
Showing 5 of 21 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 21 · See all- K070116Power Port-a-Cath and Power Port-a-Cath II; Implantable Venous Access Systems and Power…Smiths Medical MD, Inc. · 2007 · cites it directly18descendants
- K073050Safestep Max Power-Injectable Infusion SetSpecialized Health Products International, Inc. · 2008 · 2 generations on13descendants
- K170897Surecan Safety II Needle, Surecan Safety II Needle, Surecan Safety II Needle with…B.Braun Medical, Inc. · 2017 · 3 generations on6descendants
- K230715Promisemed Safety Huber NeedlePromisemed Hangzhou Meditech Co., Ltd. · 2023 · 4 generations on2descendants
- K200463Huber Needle Infusion Set, Safety Huber Needle Infusion SetJiangsu Caina Medical Co.,Ltd · 2021 · 4 generations on2descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K021999 | Gripper Plus Needle | Deltec, Inc. | 2002 | — |
| +1 | K070116 | Power Port-a-Cath and Power Port-a-Cath II; Implantable Venous Access Systems… | Smiths Medical MD, Inc. | 2007 | — |
| +2 | K073050 | Safestep Max Power-Injectable Infusion Set | Specialized Health Products International, Inc. | 2008 | 2 |
| +3 | K170897 | Surecan Safety II Needle, Surecan Safety II Needle, Surecan Safety II Needle… | B.Braun Medical, Inc. | 2017 | — |
| +4 | K230715 | Promisemed Safety Huber Needle | Promisemed Hangzhou Meditech Co., Ltd. | 2023 | — |
| +4 | K200463 | Huber Needle Infusion Set, Safety Huber Needle Infusion Set | Jiangsu Caina Medical Co.,Ltd | 2021 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2007 | 2 | 0 |
| 2008 | 1 | 0 |
| 2009 | 1 | 0 |
| 2010 | 1 | 0 |
| 2011 | 2 | 0 |
| 2012 | 1 | 0 |
| 2014 | 1 | 0 |
| 2016 | 1 | 0 |
| 2017 | 2 | 0 |
| 2020 | 1 | 0 |
| 2021 | 3 | 0 |
| 2023 | 1 | 0 |
| 2024 | 1 | 0 |
| 2025 | 2 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FPA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FPA device
- #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | GRIPPER Needle Named without a submission number | |
| * | — | Huber Plus Safety Infusion Set Named without a submission number |
What descends from it? (21)
See all 212 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jul 15, 2019 | Class II | Z-0253-2020 | Nonconforming Material/Component | Supplier initiated recall of the GRIPPER Needles |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K070116 (Smiths Medical MD, Inc., 2007) has gone furthest, with 18 descendants of its own.
How does its category compare?
All FPA devicesFPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · FPA
11in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FPA
225in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FPA
98in 2025
Class I (17 since 2007) Class II/III not classified
Adverse event reports per year · FPA
34,506in 2025
MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 16 | 75 | 17 | 1,499 |
| 2008 | 18 | 89 | 15 | 1,941 |
| 2009 | 10 | 166 | 166 | 2,512 |
| 2010 | 23 | 120 | 58 | 2,537 |
| 2011 | 12 | 87 | 28 | 4,407 |
| 2012 | 21 | 96 | 16 (4) | 4,930 |
| 2013 | 13 | 90 | 64 | 4,651 |
| 2014 | 22 | 89 | 24 | 4,695 |
| 2015 | 17 | 123 | 6 | 6,625 |
| 2016 | 12 | 195 | 10 | 8,470 |
| 2017 | 17 | 240 | 13 | 6,462 |
| 2018 | 4 | 177 | 4 | 6,574 |
| 2019 | 7 | 161 | 13 | 11,227 |
| 2020 | 10 | 327 | 107 | 9,563 |
| 2021 | 13 | 186 | 1 | 10,414 |
| 2022 | 14 | 213 | 9 (5) | 13,393 |
| 2023 | 14 | 340 | 12 (4) | 10,916 |
| 2024 | 7 | 244 | 13 (1) | 47,210 |
| 2025 | 11 | 225 | 98 (1) | 34,506 |
| 2026 | 5 | 266 | 10 (2) | — |
What is it indicated for?
The GRIPPERPLUS™ Needle is indicated for the administration into or withdrawal of fluids from implanted ports. It is designed to help protect against exposure to bloodborne pathogens caused by accidental needlestick injuries.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K021999 · printed Nov 6, 2026
Cite this page
510(k) Report. “K021999: Gripper Plus Needle.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K021999
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026