510(k) Report

510(k) · Cleared

Recalled · Class I · 2016

Inratio

K020679510(k) TraditionalCleared May 6, 200263-day reviewClass II21 CFR 864.7750Recalls 6 (4 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hemosense, Inc. · GJS Test, Time, Prothrombin

Cleared on May 6, 2002 after a 63-day review. The FDA summary names its predicate without a number. 4 of the 7 devices in its family, counting itself, have a Class I recall. 6 later devices descend from it. It has been recalled 6 times, including 4 Class I recalls.

Decided
May 6, 2002
Review
63 days
Class
Class II
Product code
GJS
Recalls
6 (4 Class I)
Predicates
0

6 later devices descend from it, 2 citing it directly. 4 of the 7 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2007: 1 descendant cleared, 1 with a Class I recall12012: 1 descendant cleared, 1 with a Class I recall12010: 2 descendants cleared, 1 with a Class I recall12019: 1 descendant cleared2022: 1 descendant cleared200020052010201520202025
  1. This deviceK020679 InratioHemosense, Inc. · 2002
  2. Descendant · cites it directly · 5 of its ownK072727 Inratio 2 PT Monitoring SystemHemosense, Inc. · 2007 · Recalled (Class I)
  3. Descendant · 2 generations on · 2 of its ownK093243 Coagusense Self-Test Prothrombin Time/Inr Monitoring SystemCoagusense, Inc. · 2010 · Recalled
  4. Descendant · 3 generations on · 1 of its ownK183255 Coag-Sense Prothrombin Time (PT) / INR Monitoring DeviceCoagusense, Inc. · 2019 · Recalled
  5. Descendant · 2 generations on · 1 of its ownK092987 Inratio/INRATIO2 Test StripsBiosite Incorporated · 2010 · Recalled (Class I)
  6. Descendant · 4 generations on · 0 of its ownK212779 Coag-Sense Prothrombin (PT) / INR Monitoring System for Patient…Coagusense, Inc. · 2022

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2007: 12010: 22012: 12019: 12022: 1
2007high 2 (2010) · low 0 (2008)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 6 · See all
  • K072727Inratio 2 PT Monitoring SystemHemosense, Inc. · 2007 · cites it directly5descendants
  • K093243Coagusense Self-Test Prothrombin Time/Inr Monitoring SystemCoagusense, Inc. · 2010 · 2 generations on2descendants
  • K183255Coag-Sense Prothrombin Time (PT) / INR Monitoring DeviceCoagusense, Inc. · 2019 · 3 generations on1descendant
  • K092987Inratio/INRATIO2 Test StripsBiosite Incorporated · 2010 · 2 generations on1descendant
  • K212779Coag-Sense Prothrombin (PT) / INR Monitoring System for Patient Self-TestingCoagusense, Inc. · 2022 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K020679InratioHemosense, Inc.2002—
+1K072727Inratio 2 PT Monitoring SystemHemosense, Inc.2007Class I
+2K093243Coagusense Self-Test Prothrombin Time/Inr Monitoring SystemCoagusense, Inc.20101
+3K183255Coag-Sense Prothrombin Time (PT) / INR Monitoring DeviceCoagusense, Inc.20191
+2K092987Inratio/INRATIO2 Test StripsBiosite Incorporated2010Class I
+4K212779Coag-Sense Prothrombin (PT) / INR Monitoring System for Patient Self-TestingCoagusense, Inc.2022—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200711
201021
201211
201910
202210
FIG. 1 — Lineage of K020679, 2002–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GJS device

  1. #1 K243543 microINR Systemscore 13.4 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—CoaguChek S
Named without a submission number
—

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K072727Inratio 2 PT Monitoring System Class IHemosense, Inc.20075
K110212INRATIO2 PI/NR Monitoring System; Inratio PT/Inr Test Strip Class IAlere San Diego, Inc (Formally Biosite Incorporate20120

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 11, 2016Class IZ-2356-2016OtherAlere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR…
Jul 11, 2016Class IZ-2355-2016OtherAlere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR…
Dec 5, 2014Class IZ-0881-2015Under Investigation by firmIn certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical…
Dec 5, 2014Class IZ-0880-2015Under Investigation by firmIn certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical…
Oct 11, 2011Not classifiedZ-0188-2012Material/Component ContaminationAlere has initiated a recall of the Professional Disposables International (PDI) alcohol prep products due to the potential presence of low levels of the microorganism Bacillus cereus.
Sep 7, 2005Not classifiedZ-0132-06OtherA software problem that may cause the INRatio meter to incorrectly display 'INR>7.5' test message under a particular use condition.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K072727Inratio 2 PT Monitoring SystemHemosense, Inc. · 2007 · Cites it directly · ◆ in the 2007 descendants bar of FIG. 14 Class I recalls
  • K110212INRATIO2 PI/NR Monitoring System; Inratio PT/Inr Test StripAlere San Diego, Inc (Formally Biosite Incorporate · 2012 · Cites it directly · ◆ in the 2012 descendants bar of FIG. 15 Class I recalls
  • K092987Inratio/INRATIO2 Test StripsBiosite Incorporated · 2010 · 2 generations on · ◆ in the 2010 descendants bar of FIG. 16 Class I recalls

What does this lineage say?

  • Of its descendants, K072727 (Hemosense, Inc., 2007) has gone furthest, with 5 descendants of its own.

How does its category compare?

All GJS devices

GJS Test, Time, Prothrombin: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GJS

3in 2025

2007: 42008: 12010: 42012: 12013: 12014: 12016: 12017: 12018: 22019: 22020: 12022: 32023: 22024: 12025: 3
2007high 4 (2007) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GJS

60in 2025

2007: 77 days median2010: 168 days median2022: 399 days median2025: 60 days median
2007high 399 (2022) · low 60 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GJS

1in 2022

2007: 52008: 32009: 42010: 22011: 12012: 12013: 12014: 4, 3 Class I2017: 52018: 3, 2 Class I2020: 22021: 12022: 1
2007high 5 (2007) · low 0 (2015)2026

Class I (5 since 2007) Class II/III not classified

Adverse event reports per year · GJS

105in 2025

2007: 4392008: 4202009: 8762010: 2,1772011: 2,9652012: 2,1982013: 2,5842014: 2,3972015: 1,2602016: 8642017: 5322018: 1,9582019: 2,3852020: 1,3902021: 1,3592022: 1,2102023: 8092024: 1772025: 105
2007high 2,965 (2011) · low 105 (2025)2025

MAUDE reports naming product code GJS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074775439
20081—3420
20090—4876
2010416822,177
20110—12,965
20121—12,198
20131—12,584
20141—4 (3)2,397
20150—01,260
20161—0864
20171—5532
20182—3 (2)1,958
20192—02,385
20201—21,390
20210—11,359
2022339911,210
20232—0809
20241—0177
20253600105
20260—0—

What is it indicated for?

The HemoSense INRatio is an in vitro diagnostic system that provides a quantitative prothrombin time results, expressed in seconds and as an International Normalized Ratio (INR). It uses fresh capillary whole blood and is intended for use by health care professionals in monitoring patients who are on warfarin-type (coumarin) anticoagulation therapy. This device is not intended to be used for screening purposes.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K020679: Inratio.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K020679

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026