510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Philips M5066A and M5068A

K020715510(k) AbbreviatedCleared Nov 8, 2002248-day reviewClass III21 CFR 870.5310Recalls 12

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Nov 8, 2002 after a 248-day review. The FDA summary names its predicate without a number. 1 of the 13 devices in its family, counting itself, has a Class I recall. 12 later devices descend from it. It has been recalled 12 times.

Decided
Nov 8, 2002
Review
248 days
Class
Class III
Product code
MKJ
Recalls
12
Predicates
0

12 later devices descend from it, 7 citing it directly. 1 of the 13 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2008: 2 descendants cleared2004: 3 descendants cleared, 1 with a Class I recall12009: 1 descendant cleared2005: 1 descendant cleared2012: 1 descendant cleared2013: 2 descendants cleared2011: 1 descendant cleared2015: 1 descendant cleared200020052010201520202025
  1. This deviceK020715 Philips M5066A and M5068APhilips Medical Systems · 2002
  2. Descendant · cites it directly · 5 of its ownK050004 Philips Heartstart FRX Defibrillator, Model 861304Philips Medical Systems · 2005 · Recalled
  3. Descendant · 2 generations on · 3 of its ownK072812 Philips Heartstart Smartpads IiiPhilips Medical Systems · 2008
  4. Descendant · 3 generations on · 2 of its ownK111693 Philips Heartstart FR3 Automated External DefibrillatorPhilips Medical Systems · 2011 · Recalled
  5. Descendant · cites it directly · 2 of its ownK040904 Philips Heartstart Home Defibrillator, Model M5068APhilips Medical Systems · 2004 · Recalled
  6. Descendant · cites it directly · 2 of its ownK041067 Heartsine Samaritan PadHeartsine Technologies, Inc. · 2004 · Recalled (Class I)

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2004: 32005: 12008: 22009: 12011: 12012: 12013: 22015: 1
2004high 3 (2004) · low 0 (2006)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 12 · See all
  • K050004Philips Heartstart FRX Defibrillator, Model 861304Philips Medical Systems · 2005 · cites it directly5descendants
  • K072812Philips Heartstart Smartpads IiiPhilips Medical Systems · 2008 · 2 generations on3descendants
  • K111693Philips Heartstart FR3 Automated External DefibrillatorPhilips Medical Systems · 2011 · 3 generations on2descendants
  • K040904Philips Heartstart Home Defibrillator, Model M5068APhilips Medical Systems · 2004 · cites it directly2descendants
  • K041067Heartsine Samaritan PadHeartsine Technologies, Inc. · 2004 · cites it directly2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K020715Philips M5066A and M5068APhilips Medical Systems2002—
+1K050004Philips Heartstart FRX Defibrillator, Model 861304Philips Medical Systems20057
+2K072812Philips Heartstart Smartpads IiiPhilips Medical Systems2008—
+3K111693Philips Heartstart FR3 Automated External DefibrillatorPhilips Medical Systems20111
+1K040904Philips Heartstart Home Defibrillator, Model M5068APhilips Medical Systems20047
+1K041067Heartsine Samaritan PadHeartsine Technologies, Inc.2004Class I

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200431
200510
200820
200910
201110
201210
201320
201510
FIG. 1 — Lineage of K020715, 2002–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Heartstream FR2 AED
Named without a submission number
—
*—Cardiac Science/SurVivaLink Firstsave AED
Named without a submission number
—
*—Katecho Adult/Defibrillator/Pacing Electrodes
Named without a submission number
—

What descends from it? (12)

See all 12

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K050004Philips Heartstart FRX Defibrillator, Model 861304Philips Medical Systems20055
K040904Philips Heartstart Home Defibrillator, Model M5068APhilips Medical Systems20042
K041067Heartsine Samaritan Pad Class IHeartsine Technologies, Inc.20042
K120350Extended Life BatteryAed Battery Exchange, LLC20120
K091548Amco Replacement Battery for Philips Medical Models BT1 and M5070AAmco International Manufacturing & Design, Inc.20090
K062970I-PADCu Medical Systems, Inc.20080
K042088Samaritan Pad Pak Defibrillation Electrode, Model SPP-301Heartsine Technologies, Inc.20040

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 10, 2019Class IIZ-1272-2020Process controlAffected devices contain a Printed Circuit Assembly (PCA) that was not manufactured/repaired in accordance to quality standards.
Sep 10, 2019Class IIZ-1271-2020Process controlAffected devices contain a Printed Circuit Assembly (PCA) that was not manufactured/repaired in accordance to quality standards.
Aug 8, 2018Class IIZ-2549-2019Labeling False and MisleadingThe recall is being initiated due to incorrect age range labeling (8+ yrs) on the foil packaging for infant/child pad cartridges. The infant/child pad cartridge is designed for use with the defibrillator An Infant/Child pad cartridge…
Mar 14, 2018Class IIZ-1914-2018Nonconforming Material/ComponentAn issue with one of the electric components (a resistor) could result in the device not delivering a shock when needed.
Feb 6, 2014Class IIZ-1232-2014Vendor change controlPhilips is recalling HeartStart Home and OnSite (HS1) automated external defibrillators (AEDs) because one or both contact pins may be contaminated by residue from the soldering process. This contamination could prevent an adequate…
Jul 26, 2012Class IIZ-0643-2013Nonconforming Material/ComponentPhilips determined that it is important to clarify information in the Owner's Manual and keep customers informed about the maintenance of their FRx (models 861304 and 861305), OnSite (models M5066A and M5067A), or HeartStart Home (model…
Apr 30, 2010Not classifiedZ-1916-2010Nonconforming Material/ComponentSeventeen AEDs failed production line testing (Final Acceptance Test).
Apr 30, 2010Not classifiedZ-1917-2010Nonconforming Material/ComponentSeventeen AEDs failed production line testing (Final Acceptance Test).
Nov 2, 2009Not classifiedZ-1194-2010Nonconforming Material/ComponentPotential for failure of a capacitor during use which would prevent the AED from delivering effective defibrillation therapy when indicated.
Nov 2, 2009Not classifiedZ-1195-2010Nonconforming Material/ComponentPotential for failure of a capacitor during use which would prevent the AED from delivering effective defibrillation therapy when indicated.
Apr 30, 2008Not classifiedZ-1780-2008Component design/selectionButtons Sticking - The On/Off and Shock buttons on the front panel of the device may stick in place when pressed and fail to respond to the button press, causing a delay in the delivery of a defibrillation shock when the system advises a…
Apr 30, 2008Not classifiedZ-1779-2008Component design/selectionButtons Sticking - The On/Off and Shock buttons on the front panel of the device may stick in place when pressed and fail to respond to the button press, causing a delay in the delivery of a defibrillation shock when the system advises a…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K041067Heartsine Samaritan PadHeartsine Technologies, Inc. · 2004 · Cites it directly · ◆ in the 2004 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K050004 (Philips Medical Systems, 2005) has gone furthest, with 5 descendants of its own.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

The M5066A is designed to be used on a person in sudden cardiac arrest, who is: unresponsive when shaken, and not breathing normally. If in doubt, apply the pads. If the victim is an infant or child younger than eight years old or weighs less than 55 lbs (25 kg), you should use the special infanuchild pads. If the child appears olded/larger, use the adult pads. Do not delay treatment to determine the child’s exact age or weight. The M5066A is intended…

Read the full statement (316 words)

The M5066A is designed to be used on a person in sudden cardiac arrest, who is:

unresponsive when shaken, and

not breathing normally.

If in doubt, apply the pads.

If the victim is an infant or child younger than eight years old or weighs less than 55 lbs (25 kg), you should use the special infanuchild pads. If the child appears olded/larger, use the adult pads. Do not delay treatment to determine the child’s exact age or weight.

The M5066A is intended for use by people who have been specifically trained in its operation. A M5066A user should also have training in cardiopulmonary resuscitation (CPR) or another physician-authorized emergency medical response program in accordance with local and state requirements.

Caution: Federal Law (USA) restricts this device to the sale by or on the order of a physician.

The M5068A is designed to be used on a person who is in sudden cardiac arrest and who is:

unresponsive when shaken, and

not breathing normally.

The defibrillator should not be used on a person who is:

responsive when shaken, or

breathing normally.

If you are not certain if the person is in sudden cardiac arrest, apply the defibrillator and follow its instructions.

For children 8 years or older, or who weigh 55 pounds or more, use the M5068A with the adult pads that come with it. For younger children to those who weigh less than 55 pounds, the special infadchild pads should be used. When used with these pads, the M5068A delivers a lower energy appropriate for infants and small children.

The M5068A is intended for use by people who have been specifically trained in its operation. The user should also have training in cardiopulmonary resuscitation (CPR) or another physician-authorized emergency response program in accordance with local and state requirements.

Caution: Federal Law (USA) restricts this device to the sale by or on the order of a physician.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K020715: Philips M5066A and M5068A.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K020715

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026