510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Heartstream XL Defibrillator/Monitor, Model #M4735A

K021453510(k) TraditionalCleared Jul 3, 200258-day reviewClass III21 CFR 870.5310Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Jul 3, 2002 after a 58-day review. The FDA summary names its predicate without a number. 2 later devices descend from it. It has been recalled 8 times.

Decided
Jul 3, 2002
Review
58 days
Class
Class III
Product code
MKJ
Recalls
8
Predicates
0

2 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2020: 1 descendant cleared2021: 1 descendant cleared200020052010201520202025
  1. This deviceK021453 Heartstream XL Defibrillator/Monitor, Model #M4735APhilips Medical Systems · 2002
  2. Descendant · cites it directly · 0 of its ownK203231 Switched Internal PaddlesPhilips North America, LLC · 2021
  3. Descendant · cites it directly · 0 of its ownK200849 Tempus LS - ManualRemote Diagnostic Technologies, Ltd. A Philips Company · 2020 · Recalled

Most influential descendants

Showing 2 of 2 · See all
  • K203231Switched Internal PaddlesPhilips North America, LLC · 2021 · cites it directly0descendants
  • K200849Tempus LS - ManualRemote Diagnostic Technologies, Ltd. A Philips Company · 2020 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K021453Heartstream XL Defibrillator/Monitor, Model #M4735APhilips Medical Systems2002—
+1K203231Switched Internal PaddlesPhilips North America, LLC2021—
+1K200849Tempus LS - ManualRemote Diagnostic Technologies, Ltd. A Philips Company20204

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
202110
FIG. 1 — Lineage of K021453, 2002–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MKJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—LIFEPAK 12 Defibrillator/Monitor
Named without a submission number
—
*—Heartstream XL Defibrillator/Monitor
Named without a submission number
—

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K203231Switched Internal PaddlesPhilips North America, LLC20210
K200849Tempus LS - ManualRemote Diagnostic Technologies, Ltd. A Philips Company20200

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 20, 2020Class IIZ-2134-2020Device DesignInternal Paddles may wear over time and may not be safe or ready for use, unless the directions in the Instructions for Use (IFU) to perform routine operational checks are followed. The instructions for use rely on these checks rather…
Sep 11, 2012Class IIZ-0011-2013Nonconforming Material/ComponentHeartStart XL Defibrillator/Monitor (M4735A) used on a patient when AC power is not in use, with a battery that may fail, the device may shut down unexpectedly without providing the user with a low battery warning message or audible tone.
Jan 19, 2010Not classifiedZ-1135-2010Nonconforming Material/ComponentPaddle shock switch may fail to actuate and delay or prevent delivery of defibrillation therapy
Mar 30, 2009Not classifiedZ-1199-2009Nonconforming Material/ComponentRotary energy select switch in affected M4735A devices may fail and prevent the user from turning the device on, rendering the device unusable for monitoring and defibrillation therapy.
Sep 4, 2007Not classifiedZ-0062-2008Package design/selectionInadequate shipping package causing damage to control boards resulting in unit failure. Failure to deliver therapy (defibrillation, pacing or cardioversion) may occur as a result of undetected damage to the PCB assembly.
Jun 11, 2007Not classifiedZ-1128-2007Nonconforming Material/ComponentThe ECG signal may become unusable to assess patient condition and need for therapy.
Sep 2, 2003Not classifiedZ-1226-03OtherInstructions For Use include addtional information on configuration for selecting Power on Lead
Nov 25, 2002Not classifiedZ-0343-03OtherDefective energy switch may cause Defibrillator to turn off or charge to a different energy level than was selected.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K203231 (Philips North America, LLC, 2021) has gone furthest, with 0 descendants of its own.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

The Heartstream XL Defibrillator/Monitor is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. The SMART Biphasic waveform utilized in the Heartstream XL Defibrillator/Monitor has…

Read the full statement (254 words)

The Heartstream XL Defibrillator/Monitor is for use for the termination of ventricular tachycardia and ventricular fibrillation.

The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.

The SMART Biphasic waveform utilized in the Heartstream XL Defibrillator/Monitor has previously undergone clinical testing in adults. These trials support the waveform's effectiveness for defibrillation of ventricular tachyarrhythmias at 150J.

AED Therapy: To be used in the presence of a suspected cardiac arrest on patients of at least 8 years of age that are unresponsive, not breathing and pulseless.

Manual Defibrillation: Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients that are pulseless and unresponsive.

The SMART Biphasic waveform utilized in the Heartstream XL Defibrillator/Monitor incorporates some user selectable lower energy levels that were not used in the clinical trials.

Synchronized defibrillation is indicated for termination of atrial fibrillation. The SMART Biphasic waveform utilized in the Heartstream XL has undergone clinical testing demonstrating its safety and effectiveness for cardioversion of atrial fibrillation.

There are currently no clinical studies related to the use of SMART Biphasic waveform in pediatric applications.

Pacing: Noninvasive pacing is one method of treating patients with symptomatic bradycardia. It can also be helpful in patients with asystole, if performed early.

SpO2 Monitoring: SpO2 monitoring is indicated for use when it is beneficial to assess a patient's oxygen saturation level.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K021453: Heartstream XL Defibrillator/Monitor, Model #M4735A.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K021453

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026