510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Tempus LS - Manual

K200849510(k) TraditionalCleared Jul 24, 2020115-day reviewClass II21 CFR 870.5300Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Remote Diagnostic Technologies, Ltd. A Philips Company · LDD Dc-Defibrillator, Low-Energy, (Including Paddles)

Cleared on Jul 24, 2020 as substantially equivalent to K140502 (ZOLL Medical Corporation, 2014) and 2 other predicates after a 115-day review. Its family tree reaches back 13 generations to 1997. 5 of the 65 devices in its family, counting itself, have a Class I recall. It has been recalled 4 times.

Decided
Jul 24, 2020
Review
115 days
Class
Class II
Product code
LDD
Recalls
4
Predicates
3

Its longest line of descent runs back 13 generations to K962603 (1997). 64 ancestors in all, the oldest K893221 (1990). No later device cites it as a predicate yet. 5 of the 65 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2014: 1 other ancestor2002: 3 other ancestors, 1 with a Class I recall12012: 1 other ancestor2011: 2 other ancestors2009: 2 other ancestors2008: 1 other ancestor2006: 2 other ancestors, 1 with a Class I recall12005: 3 other ancestors, 1 with a Class I recall12004: 2 other ancestors2003: 3 other ancestors, 1 with a Class I recall12001: 3 other ancestors2000: 2 other ancestors1999: 6 other ancestors1996: 3 other ancestors1995: 5 other ancestors1990: 2 other ancestors1998: 4 other ancestors1997: 2 other ancestors, 1 with a Class I recall11991: 2 other ancestors1993: 1 other ancestor1992: 1 other ancestorK962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999×2: 2 generations cleared close together, K012891 (2002), K021090 (2002) K033048 · Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS, 740-2NL, 740M-2NL, 740-2NN, 740M-2NN, 740-3MS, 740M-3MS · Cas Medical Systems, Inc. · 2004 · recalledK050844 · 750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL, 750CM-2NL, 750C-3MS, 750CM-3MS, 750C-3NL, 750CM-3NL · Cas Medical Systems, Inc. · 2005K060065 · CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One Parameter Bedside Monitor · Oridion Capnography, Inc. · 2006 · recalledK072295 · Capnostream 20 with A2 Adaptive Averaging Software · Oridion Capnography, Inc. · 2007K082268 · CAPNOSTREAM20 with Integrated Pulmonary Index · Oridion Capnography, Inc. · 2009 · recalledK130153 · Hearstart MRX with Airway Confimration Assist · Philips Medical Systems · 2013K133659 · Heartstart XL+Defibrillator/Monitor with End-Tidal Carbon Dioxide Monitoring · Philips Medical Systems · 20151990199520002005201020152020
Other ancestors, by generation 1 generation back: 2 other ancestors2 generations back: 5 other ancestors3 generations back: 3 other ancestors, 1 with a Class I recall14 generations back: 10 other ancestors, 2 with a Class I recall25 generations back: 2 other ancestors, 1 with a Class I recall16 generations back: 7 other ancestors7 generations back: 9 other ancestors8 generations back: 6 other ancestors, 1 with a Class I recall19 generations back: 5 other ancestors10 generations back: 2 other ancestors K133659 · Heartstart XL+Defibrillator/Monitor with End-Tidal Carbon Dioxide Monitoring · Philips Medical Systems · 2015K130153 · Hearstart MRX with Airway Confimration Assist · Philips Medical Systems · 2013K082268 · CAPNOSTREAM20 with Integrated Pulmonary Index · Oridion Capnography, Inc. · 2009 · recalledK072295 · Capnostream 20 with A2 Adaptive Averaging Software · Oridion Capnography, Inc. · 2007K060065 · CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One Parameter Bedside Monitor · Oridion Capnography, Inc. · 2006 · recalledK050844 · 750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL, 750CM-2NL, 750C-3MS, 750CM-3MS, 750C-3NL, 750CM-3NL · Cas Medical Systems, Inc. · 2005K033048 · Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS, 740-2NL, 740M-2NL, 740-2NN, 740M-2NN, 740-3MS, 740M-3MS · Cas Medical Systems, Inc. · 2004 · recalledK021090 · Oximax N-550 Pulse Oximeter,Model N-550 Pulse Oximeter · Nellcor Puritan Bennett, Inc. · 2002K012891 · Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and Cables (Aka Accessories) · Nellcor Puritan Bennett, Inc. · 2002 · recalledK991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997−10−50
  1. Root · 13 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 9 generations between
  3. 3 generations backK082268 CAPNOSTREAM20 with Integrated Pulmonary IndexOridion Capnography, Inc. · 2009 · Recalled
  4. 2 generations backK130153 Hearstart MRX with Airway Confimration AssistPhilips Medical Systems · 2013
  5. 1 generation backK133659 Heartstart XL+Defibrillator/Monitor with End-Tidal Carbon Dioxide…Philips Medical Systems · 2015
  6. This deviceK200849 Tempus LS - ManualRemote Diagnostic Technologies, Ltd. A Philips Company · 2020
Show all 13 generations
  1. Root · 13 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 12 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 11 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
  4. 10 generations backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
  5. 9 generations backK012891 Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax…Nellcor Puritan Bennett, Inc. · 2002 · Recalled
  6. 8 generations backK021090 Oximax N-550 Pulse Oximeter,Model N-550 Pulse OximeterNellcor Puritan Bennett, Inc. · 2002
  7. 7 generations backK033048 Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS…Cas Medical Systems, Inc. · 2004 · Recalled
  8. 6 generations backK050844 750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL…Cas Medical Systems, Inc. · 2005
  9. 5 generations backK060065 CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One…Oridion Capnography, Inc. · 2006 · Recalled
  10. 4 generations backK072295 Capnostream 20 with A2 Adaptive Averaging SoftwareOridion Capnography, Inc. · 2007
  11. 3 generations backK082268 CAPNOSTREAM20 with Integrated Pulmonary IndexOridion Capnography, Inc. · 2009 · Recalled
  12. 2 generations backK130153 Hearstart MRX with Airway Confimration AssistPhilips Medical Systems · 2013
  13. 1 generation backK133659 Heartstart XL+Defibrillator/Monitor with End-Tidal Carbon Dioxide…Philips Medical Systems · 2015
  14. This deviceK200849 Tempus LS - ManualRemote Diagnostic Technologies, Ltd. A Philips Company · 2020

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−13K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−12K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−11K982255Model 2000 Pulse OximeterIvy Biomedical Systems, Inc.1998—
−10K991823N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc.1999—
−9K012891Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and…Nellcor Puritan Bennett, Inc.20021
−8K021090Oximax N-550 Pulse Oximeter,Model N-550 Pulse OximeterNellcor Puritan Bennett, Inc.2002—
−7K033048Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS, 740-2NL…Cas Medical Systems, Inc.20041
−6K050844750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL, 750CM-2NL…Cas Medical Systems, Inc.2005—
−5K060065CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One Parameter…Oridion Capnography, Inc.20063
−4K072295Capnostream 20 with A2 Adaptive Averaging SoftwareOridion Capnography, Inc.2007—
−3K082268CAPNOSTREAM20 with Integrated Pulmonary IndexOridion Capnography, Inc.20099
−2K130153Hearstart MRX with Airway Confimration AssistPhilips Medical Systems2013—
−1K133659Heartstart XL+Defibrillator/Monitor with End-Tidal Carbon Dioxide MonitoringPhilips Medical Systems2015—
0 (this device)K200849Tempus LS - ManualRemote Diagnostic Technologies, Ltd. A Philips Company2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199020
199120
199210
199310
199550
199630
199721
199840
199960
200020
200130
200231
200331
200420
200531
200621
200810
200920
201120
201210
201410
FIG. 1 — Lineage of K200849, 1997–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LDD device

  1. #1 K203231 Switched Internal Paddlesscore 1.0 · 2 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K140502Zoll E Series Als
Summary page 5
…es: DRO, LIX, DPS Regulation Number: 21 CFR 870. 5300 Predicate Device(s) Primary Predicate: ZOLL E-Series ALS (K140502) Secondary Predicate: Philips HeartStart XL (K021453) Philips HeartStar…
ZOLL Medical Corporation2014—
*K021453Heartstream XL Defibrillator/Monitor, Model #M4735A
Summary page 5
…R 870. 5300 Predicate Device(s) Primary Predicate: ZOLL E-Series ALS (K140502) Secondary Predicate: Philips HeartStart XL (K021453) Philips HeartStart XL+ (K133659) Device Description The TEMPUS LS – MANUAL is a compac…
Philips Medical Systems20028
*K133659Heartstart XL+Defibrillator/Monitor with End-Tidal Carbon Dioxide Monitoring
Summary page 5
…ZOLL E-Series ALS (K140502) Secondary Predicate: Philips HeartStart XL (K021453) Philips HeartStart XL+ (K133659) Device Description The TEMPUS LS – MANUAL is a compact, lightweight, portable electromedical device that contains a direct cu…
Philips Medical Systems2015—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 28, 2024Class IIIZ-0557-2025Labeling Change ControlDue to incorrect labels applied to devices during servicing. Labels contain the incorrect product name and CAT number.
May 9, 2024Class IIZ-2028-2024Labeling designSimultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may lead to no or…
May 8, 2023Class IIZ-1825-2023Device DesignDefibrillator/pacemaker module (DPM) may encounter a problem causing communication from the Mainboard to the DPM to fail, leading to no or ineffective pacing with displayed error message "HW failure DPM".
Nov 17, 2020Class IIZ-0564-2021Software designA software error was detected within software version 1.3.4 for the Tempus LS device. When exiting pacer back into manual mode the pacer pulse wave may put the ECG calibration out of calibration. This software error may result in a delay…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K061707Heartstart MRX, Invasive Pressure & Temperature OptionsPhilips Medical Systems · 2006 · 3 generations back · ◆ in the 2006 ancestors bar of FIG. 12 Class I recalls
  • K051134Heartstart MRX with Q-CPR Option, Models M3535A or M3536APhilips Medical Systems · 2005 · 4 generations back · ◆ in the 2005 ancestors bar of FIG. 12 Class I recalls
  • K031187Heartstart MRX, Model M3535APhilips Medical Systems · 2003 · 4 generations back · ◆ in the 2003 ancestors bar of FIG. 12 Class I recalls
  • K022732Infant/Child Reduced Energy ElectrodesMedtronic Physio-Control Corp. · 2002 · 5 generations back · ◆ in the 2002 ancestors bar of FIG. 12 Class I recalls
  • K970209Capnocheck PlusBci Intl., Inc. · 1997 · 8 generations back · ◆ in the 1997 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 23 years and passes through 7 companies; a new generation was cleared every 1.3 years on median.
  • 17 of the 64 ancestors have a recall of any class; 5 have a Class I recall.

How does its category compare?

All LDD devices

LDD Dc-Defibrillator, Low-Energy, (Including Paddles): FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · LDD

1in 2021

2008: 22010: 22011: 42014: 12015: 12019: 12020: 12021: 1
2007high 4 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LDD

124in 2011

2011: 124 days median
2007high 124 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LDD

2in 2024

2008: 112009: 12011: 12014: 2, 2 Class I2020: 12022: 12024: 2
2007high 11 (2008) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · LDD

74in 2025

2007: 5502008: 1,2142009: 1,0572010: 6352011: 3862012: 3982013: 6022014: 4812015: 862016: 582017: 492018: 282019: 342020: 192021: 222022: 802023: 2642024: 1602025: 74
2007high 1,214 (2008) · low 19 (2020)2025

MAUDE reports naming product code LDD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—0550
20082—111,214
20090—11,057
20102—0635
201141241386
20120—0398
20130—0602
20141—2 (2)481
20151—086
20160—058
20170—049
20180—028
20191—034
20201—119
20211—022
20220—180
20230—0264
20240—2160
20250—074
20260—0—

What is it indicated for?

The TEMPUS LS - MANUAL is a multi-function, portable device that is capable of providing defibrillation therapy, non-invasive external pacing, displaying ECG, and CPR feedback in hospital and pre-hospital settings. It is to be used by qualified medical health care professionals trained in the use and operation of the device and qualified by training in advanced cardiovascular life (ACLS) support. Manual Defibrillation In Manual Defibrillation mode, the…

Read the full statement (438 words)

The TEMPUS LS - MANUAL is a multi-function, portable device that is capable of providing defibrillation therapy, non-invasive external pacing, displaying ECG, and CPR feedback in hospital and pre-hospital settings. It is to be used by qualified medical health care professionals trained in the use and operation of the device and qualified by training in advanced cardiovascular life (ACLS) support.

Manual Defibrillation

In Manual Defibrillation mode, the TEMPUS LS - MANUAL is indicated for victims of cardiac arrest where there is apparent lack of circulation as indicated by these three conditions:

• Unconsciousness

• Absence of breathing, and

• Absence of pulse.

This product should be used only by qualified medical personnel for terminating ventricular fibrillation and ventricular tachycardia.

Defibrillation is indicated for adult and pediatric patients. When a victim is less than 8 years of age, or weighs less than 55 lbs. (25kg), the TEMPUS LS - MANUAL should be used with the Tempus Pediatric Electrodes or in the device’s pediatric mode with the Tempus Adult Electrodes. Therapy should not be delayed to determine the patient’s exact age or weight.

Cardioversion

The TEMPUS LS – MANUAL may be used for synchronized cardioversion to terminate atrial fibrillation (AF) or ventricular tachycardia (VT) by using the R-wave of the patient’s ECG as a timing reference. A qualified user trained in the operation of the device must decide when synchronized cardioversion is appropriate. Cardioversion is indicated for adult and pediatric patients.

External Pacing

The TEMPUS LS – MANUAL is indicated for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation. The purposes of pacing include:

• Resuscitation from standstill or bradycardia of any etiology.

• As a standby when standstill or bradycardia might be expected.

External pacing is indicated for adult and pediatric patients.

ECG Display

The TEMPUS LS – MANUAL is intended to display a patient’s ECG from a 3- or 4- lead wire ECG patient cable. It is indicated for use in adult or pediatric patients for attended monitoring of the patient’s ECG signal (e.g., before starting demand pacing or after successful defibrillation).

CPR Feedback

The TEMPUS LS – MANUAL may be supplied with an optional CPR feedback sensor, which is intended to provide visual and audio feedback to encourage rescuers to perform CPR in accordance with AHA/ERC guidelines for chest compression rate and depth. It is indicated for use on adult and pediatric patients ≥ 8 years of age or weighing ≥ 25 kg / 55 lbs.

CPR Feedback is not indicated for use on neonatal and pediatric patients under 8 years of age or weighing less than 25 kg/55 lbs.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K200849: Tempus LS - Manual.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K200849

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026