510(k) Report

510(k) · Cleared

Cleared

W & H Lisa Steam Sterilizer

K022816510(k) AbbreviatedCleared Feb 7, 2003165-day reviewClass II21 CFR 880.6880

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

A-Dec, Inc. · FLE Sterilizer, Steam

Cleared on Feb 7, 2003 after a 165-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 9 later devices descend from it.

Decided
Feb 7, 2003
Review
165 days
Class
Class II
Product code
FLE
Recalls
None
Predicates
0

9 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2006: 1 descendant cleared2016: 1 descendant cleared2009: 1 descendant cleared2015: 1 descendant cleared2022: 2 descendants cleared2017: 1 descendant cleared2023: 1 descendant cleared2025: 1 descendant cleared200020052010201520202025
  1. This deviceK022816 W & H Lisa Steam SterilizerA-Dec, Inc. · 2003
  2. Descendant · cites it directly · 8 of its ownK050263 Millennium Series, Including Millennium B, B+, B2 and BUM.O. Com S.R.L. · 2006
  3. Descendant · 2 generations on · 6 of its ownK143311 Pre-vacuum Tabletop AutoclaveTuttnauer U.S.A Co, Ltd. · 2015
  4. Descendant · 3 generations on · 2 of its ownK213080 T-Edge 10, T-Edge 11Tuttnauer, Ltd. · 2022
  5. Descendant · 4 generations on · 1 of its ownK232658 T-Top 10 & T-Top T-Top 11Tuttnauer U.S.A Co, Ltd. · 2023
  6. Descendant · 2 generations on · 1 of its ownK151597 Family of Steam Sterilizers B 28 seriesCefla S.C. · 2016

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2006: 12009: 12015: 12016: 12017: 12022: 22023: 12025: 1
2006high 2 (2022) · low 0 (2007)2026

Most influential descendants

Showing 5 of 9 · See all
  • K050263Millennium Series, Including Millennium B, B+, B2 and BUM.O. Com S.R.L. · 2006 · cites it directly8descendants
  • K143311Pre-vacuum Tabletop AutoclaveTuttnauer U.S.A Co, Ltd. · 2015 · 2 generations on6descendants
  • K213080T-Edge 10, T-Edge 11Tuttnauer, Ltd. · 2022 · 3 generations on2descendants
  • K232658T-Top 10 & T-Top T-Top 11Tuttnauer U.S.A Co, Ltd. · 2023 · 4 generations on1descendant
  • K151597Family of Steam Sterilizers B 28 seriesCefla S.C. · 2016 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K022816W & H Lisa Steam SterilizerA-Dec, Inc.2003—
+1K050263Millennium Series, Including Millennium B, B+, B2 and BUM.O. Com S.R.L.2006—
+2K143311Pre-vacuum Tabletop AutoclaveTuttnauer U.S.A Co, Ltd.2015—
+3K213080T-Edge 10, T-Edge 11Tuttnauer, Ltd.2022—
+4K232658T-Top 10 & T-Top T-Top 11Tuttnauer U.S.A Co, Ltd.2023—
+2K151597Family of Steam Sterilizers B 28 seriesCefla S.C.2016—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200910
201510
201610
201710
202220
202310
202510
FIG. 1 — Lineage of K022816, 2003–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FLE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FLE device

  1. #1 K250168 Life Ultra 25L Tabletop Mono Chamber Steam…score 12.7 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (9)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K050263Millennium Series, Including Millennium B, B+, B2 and BUM.O. Com S.R.L.20068

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K050263 (M.O. Com S.R.L., 2006) has gone furthest, with 8 descendants of its own.

How does its category compare?

All FLE devices

FLE Sterilizer, Steam: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FLE

4in 2025

2007: 52008: 12009: 82010: 22011: 52012: 52013: 52014: 22015: 32016: 12017: 32018: 82019: 52020: 22021: 52022: 102023: 72024: 22025: 42026 (partial year): 3
2007high 10 (2022) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FLE

184in 2025

2007: 67 days median2009: 109 days median2011: 141 days median2012: 111 days median2013: 241 days median2015: 302 days median2017: 233 days median2018: 146 days median2019: 192 days median2021: 147 days median2022: 133 days median2023: 88 days median2025: 184 days median2026 (partial year): 85 days median
2007high 302 (2015) · low 67 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FLE

1in 2021

2007: 12008: 12011: 12012: 12013: 22015: 12018: 12019: 22021: 12026 (partial year): 1
2007high 2 (2013) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FLE

71in 2025

2007: 192008: 212009: 192010: 202011: 632012: 692013: 1122014: 1132015: 682016: 762017: 632018: 902019: 912020: 872021: 812022: 702023: 662024: 642025: 71
2007high 113 (2014) · low 19 (2007)2025

MAUDE reports naming product code FLE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007567119
20081—121
20098109019
20102—020
20115141163
20125111169
201352412112
20142—0113
20153302168
20161—076
20173233063
20188146190
20195192291
20202—087
20215147181
202210133070
2023788066
20242—064
20254184071
20263851—

What is it indicated for?

The W&H Lisa Water Steam Sterilizer is designed for pressurized steam sterilization of medical and dental instruments (wrapped and unwrapped), including complex lumened devices (such as dental handpieces), porous and hollow loads. It has the following automated program cycles: Two Sterilization Cycles (summaries): B Standard 273 Cycle | B Standard 250 Cycle Max Temperature 273 °F 250 °F Duration of sterilization 4 min 30 min Pressure 31.5 psi 17 psi…

Read the full statement (170 words)

The W&H Lisa Water Steam Sterilizer is designed for pressurized steam sterilization of medical and dental instruments (wrapped and unwrapped), including complex lumened devices (such as dental handpieces), porous and hollow loads.

It has the following automated program cycles:

Two Sterilization Cycles (summaries):

B Standard 273 Cycle | B Standard 250 Cycle

Max Temperature 273 °F 250 °F

Duration of sterilization 4 min 30 min

Pressure 31.5 psi 17 psi

Duration of drying 15 min 20 min

Total cycle time empty 30 min 65 min

Total cycle time full load 40 min 75 min

Uses:

Full solid load Yes Yes

Full porous textile load Yes Yes

Small porous items Yes Yes

Dental Handpieces Yes Yes

Unwrapped material Yes Yes

Single/double wrapped 4.5 lbs max. wrapped 4.5 lbs max. wrapped

Liquids No No

Two Test Cycles (summaries):

Bowie & Dick Vacuum

Max Temperature 275.9 °F None

Duration of sterilization 3 min 20 sec None

Max Pressure 31.5 psi -12.3 psi

Duration of drying 4 min None

Test period N/A 10 min

Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K022816: W & H Lisa Steam Sterilizer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K022816

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026