510(k) Report

510(k) · Cleared

Cleared

AMSCO 700 Steam Sterilizer

K261055510(k) TraditionalCleared Jun 24, 202685-day reviewClass II21 CFR 880.6880

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · FLE Sterilizer, Steam

Cleared on Jun 24, 2026 as substantially equivalent to K232485 (Steris, 2023) after an 85-day review. Its family tree reaches back 1 generation to 2023.

Decided
Jun 24, 2026
Review
85 days
Class
Class II
Product code
FLE
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K232485 (2023). No later device cites it as a predicate yet.

K232485 · AMSCO 600 Steam Sterilizer · Steris · 202320202022202420262028
K232485 · AMSCO 600 Steam Sterilizer · Steris · 20230
  1. Root · 1 generation backK232485 AMSCO 600 Steam SterilizerSteris · 2023
  2. This deviceK261055 AMSCO 700 Steam SterilizerSTERIS Corporation · 2026
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K232485AMSCO 600 Steam SterilizerSteris2023—
0 (this device)K261055AMSCO 700 Steam SterilizerSTERIS Corporation2026—
FIG. 1 — Lineage of K261055, 2023–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FLE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FLE device

  1. #1 K250168 Life Ultra 25L Tabletop Mono Chamber Steam…score 12.7 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K232485AMSCO 600 Steam Sterilizer
Summary page 7
…Sterilizer Automated Loading System Classification Number: 21 CFR 880.6880 Product Code: FLE, PEC 2. Predicate Device K232485 AMSCO 600 Steam Sterilizer 3. Description of Device The AMSCO 700 Steam Sterilizer uses saturated steam, generated from a ho…
Steris2023—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.

How does its category compare?

All FLE devices

FLE Sterilizer, Steam: FDA decisions, review times, recalls and adverse event reports over 20 years. 2026, this device’s year, is highlighted.

FDA decisions per year · FLE

4in 2025

2007: 52008: 12009: 82010: 22011: 52012: 52013: 52014: 22015: 32016: 12017: 32018: 82019: 52020: 22021: 52022: 102023: 72024: 22025: 42026 (partial year): 3
2007high 10 (2022) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FLE

184in 2025

2007: 67 days median2009: 109 days median2011: 141 days median2012: 111 days median2013: 241 days median2015: 302 days median2017: 233 days median2018: 146 days median2019: 192 days median2021: 147 days median2022: 133 days median2023: 88 days median2025: 184 days median2026 (partial year): 85 days median
2007high 302 (2015) · low 67 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FLE

1in 2021

2007: 12008: 12011: 12012: 12013: 22015: 12018: 12019: 22021: 12026 (partial year): 1
2007high 2 (2013) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FLE

71in 2025

2007: 192008: 212009: 192010: 202011: 632012: 692013: 1122014: 1132015: 682016: 762017: 632018: 902019: 912020: 872021: 812022: 702023: 662024: 642025: 71
2007high 113 (2014) · low 19 (2007)2025

MAUDE reports naming product code FLE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007567119
20081—121
20098109019
20102—020
20115141163
20125111169
201352412112
20142—0113
20153302168
20161—076
20173233063
20188146190
20195192291
20202—087
20215147181
202210133070
2023788066
20242—064
20254184071
20263851—

What is it indicated for?

The AMSCO 700 Steam Sterilizers are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are equipped with the following factory-programmed cycles: Table 1. AMSCO 700 Steam Sterilizer sterilization cycles and cycle values Sterilize Cycles Sterilize Time Dry Time Maximum Recommended Load Temperature Fabric Packs. Refer to Table 2 for Prevac 270°F (132°C) 4 minutes 20 minutes recommended quantities. Double…

Read the full statement (352 words)

The AMSCO 700 Steam Sterilizers are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are equipped with the following factory-programmed cycles:

Table 1. AMSCO 700 Steam Sterilizer sterilization cycles and cycle values

Sterilize

Cycles Sterilize Time Dry Time Maximum Recommended Load

Temperature

Fabric Packs. Refer to Table 2 for

Prevac 270°F (132°C) 4 minutes 20 minutes

recommended quantities.

Double wrapped instrument trays, maximum

Prevac 270°F (132°C) 4 minutes 30 minutes weight 25 lbs. (11.3 kg) each and Fabric Packs.

See Table 2 for maximum quantities.

Prevac 270°F (132°C) 4 minutes 5 minutes Single Fabric Pack.

Double wrapped instrument trays, maximum

Prevac 275°F (135°C) 3 minutes 30 minutes weight 25 lbs. (11.3 kg) each. Refer to Table 2

for recommended quantities.

Double wrapped instrument trays, maximum

Prevac 270°F (132°C) 10 minutes 30 minutes weight 25 lbs. (11.3 kg) each. Refer to Table 2

for recommended quantities.

Prevac-

270°F (132°C) 4 minutes 1 minute Immediate use – single unwrapped tray

IUSS

Double wrapped instrument trays, maximum

Gravity 270°F (132°C) 15 minutes 30 minutes weight 25 lbs. (11.3 kg) each. Refer to Table 2

for recommended quantities.

Double wrapped instrument trays, maximum

Gravity 250°F (121°C) 30 minutes 30 minutes weight 25 lbs. (11.3 kg) each. Refer to Table 2

for recommended quantities.

Warm-Up 270°F (132°C) 3 minutes 1 minute N/A

DART 270°F (132°C) 3.5 minutes 1 minute Bowie-Dick Test Pack, DART Test Pack

Leak Test N/A N/A N/A N/A

The maximum quantities of wrapped instrument trays or fabric packs for each proposed model are as follows:

Table 2. AMSCO 700 Steam Sterilizer maximum load per sterilizer size

Wrapped

Sterilizer Size Fabric Packs

Instrument Trays

26” x 26” x 39” 9 12

26” x 26” x 51” 12 16

26” x 26” x 63” 15 20

The Automated Load and Unload System (ALUS) provides semi-automated loading and unloading from an AMSCO 700 steam sterilizer when a cycle is complete. Alternatively, the ALUS may also be used to provide automatic unloading only in combination with manual loading. The ALUS can start a cycle automatically when equipped with the optional bar code reader.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K261055: AMSCO 700 Steam Sterilizer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K261055

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded Oct 30, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026