510(k) · Cleared
Recalled · Class II · 2016EB-1830T3, Video Bronchoscope
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Pentax Precision Instrument Corp. · EOQ Bronchoscope (Flexible Or Rigid)
Cleared on Oct 16, 2002 as substantially equivalent to K934920 (Pentax Precision Instrument Corp., 1994) after an 8-day review. Its family tree reaches back 1 generation to 1994. 6 later devices descend from it. It has been recalled 3 times.
- Decided
- Oct 16, 2002
- Review
- 8 days
- Class
- Class II
- Product code
- EOQ
- Recalls
- 3
- Predicates
- 1
Its longest line of descent runs back 1 generation to K934920 (1994). 6 later devices descend from it, 1 citing it directly.
- Root · 1 generation backK934920 VB-1830/VB-1530, Video BronchoscopePentax Precision Instrument Corp. · 1994
- This deviceK023376 EB-1830T3, Video BronchoscopePentax Precision Instrument Corp. · 2002
- Descendant · cites it directly · 5 of its ownK081518 Ultrasound Video Bronchoscope, Model EB-1970UKPentax Medical Company · 2008
- Descendant · 2 generations on · 3 of its ownK131946 Entax Ultrasound Video Bronchoscope EB-1970UK + HI Visin PreirusPentax Medical Company · 2014 · Recalled
- Descendant · 3 generations on · 1 of its ownK183516 PENTAX Medical Ultrasound Video Bronchoscope EB19-J10UPentax of America, Inc. · 2019
- Descendant · 4 generations on · 0 of its ownK203166 PENTAX Medical Ultrasound Upper GI Video Scope EG34-J10U, PENTAX…Pentax of America, Inc. · 2020
- Descendant · 3 generations on · 0 of its ownK183654 PENTAX Medical EB-1970UK Ultrasound Video Bronchoscope + Hitachi…Pentax of America, Inc. · 2019
Showing 5 of 6 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 6 · See all- K081518Ultrasound Video Bronchoscope, Model EB-1970UKPentax Medical Company · 2008 · cites it directly5descendants
- K131946Entax Ultrasound Video Bronchoscope EB-1970UK + HI Visin PreirusPentax Medical Company · 2014 · 2 generations on3descendants
- K183516PENTAX Medical Ultrasound Video Bronchoscope EB19-J10UPentax of America, Inc. · 2019 · 3 generations on1descendant
- K203166PENTAX Medical Ultrasound Upper GI Video Scope EG34-J10U, PENTAX Medical Ultrasound…Pentax of America, Inc. · 2020 · 4 generations on0descendants
- K183654PENTAX Medical EB-1970UK Ultrasound Video Bronchoscope + Hitachi NOBLUS and HI VISION…Pentax of America, Inc. · 2019 · 3 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K934920 | VB-1830/VB-1530, Video Bronchoscope | Pentax Precision Instrument Corp. | 1994 | — |
| 0 (this device) | K023376 | EB-1830T3, Video Bronchoscope | Pentax Precision Instrument Corp. | 2002 | — |
| +1 | K081518 | Ultrasound Video Bronchoscope, Model EB-1970UK | Pentax Medical Company | 2008 | — |
| +2 | K131946 | Entax Ultrasound Video Bronchoscope EB-1970UK + HI Visin Preirus | Pentax Medical Company | 2014 | 4 |
| +3 | K183516 | PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U | Pentax of America, Inc. | 2019 | — |
| +4 | K203166 | PENTAX Medical Ultrasound Upper GI Video Scope EG34-J10U, PENTAX Medical… | Pentax of America, Inc. | 2020 | — |
| +3 | K183654 | PENTAX Medical EB-1970UK Ultrasound Video Bronchoscope + Hitachi NOBLUS and HI… | Pentax of America, Inc. | 2019 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2008 | 1 | 0 |
| 2014 | 1 | 0 |
| 2018 | 1 | 0 |
| 2019 | 2 | 0 |
| 2020 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in EOQ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new EOQ device
- #1 K223299 Ambu® aScope™ 5 Broncho HD 5.0/2.2, Ambu®…score 18.4 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K934920 | VB-1830/VB-1530, Video BronchoscopeSummary page 1…Manufacturer PMN# VB-1830, Video Bronchoscope PPIC K934920 Intended Use: The EB-l83OT3, Video Bronchoscope, is intended to provide optical visualization (via a video monitor) of, and ther… | 1994 |
What descends from it? (6)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K081518 | Ultrasound Video Bronchoscope, Model EB-1970UK | 2008 | 5 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Apr 6, 2016 | Class II | Z-2076-2016 | Device Design | Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel. |
| Apr 6, 2016 | Class II | Z-2065-2016 | Device Design | Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel. |
| Sep 30, 2011 | Not classified | Z-0818-2012 | Employee error | The wrong silicone sealant was used and excess sealant was neglected to be removed during repair of the bronchoscope. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All EOQ devicesEOQ Bronchoscope (Flexible Or Rigid): FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · EOQ
20in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · EOQ
147in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · EOQ
25in 2025
Class I (28 since 2007) Class II/III not classified
Adverse event reports per year · EOQ
6,546in 2025
MAUDE reports naming product code EOQ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 0 | 39 |
| 2008 | 2 | — | 0 | 87 |
| 2009 | 3 | 24 | 0 | 114 |
| 2010 | 0 | — | 0 | 100 |
| 2011 | 0 | — | 1 | 96 |
| 2012 | 2 | — | 4 | 93 |
| 2013 | 4 | 178 | 0 | 93 |
| 2014 | 3 | 302 | 0 | 98 |
| 2015 | 6 | 206 | 0 | 132 |
| 2016 | 3 | 262 | 1 | 297 |
| 2017 | 4 | 136 | 0 | 337 |
| 2018 | 7 | 212 | 0 | 279 |
| 2019 | 11 | 147 | 1 | 654 |
| 2020 | 10 | 103 | 1 | 515 |
| 2021 | 15 | 169 | 9 | 1,358 |
| 2022 | 11 | 134 | 5 | 2,541 |
| 2023 | 12 | 160 | 18 (4) | 5,064 |
| 2024 | 13 | 130 | 4 | 6,051 |
| 2025 | 20 | 147 | 25 (24) | 6,546 |
| 2026 | 9 | 206 | 0 | — |
What is it indicated for?
The EB-1830T3, Video Bronchoscope, is intended to provide optical visualization (via a video monitor) of, and therapeutic access to, the Pulmonary Tract. The Pulmonary Tract includes, but is not restricted to, the organs; tissues; and subsystems: Nasal Passage, Pharynx, Larynx, Trachea, and bronchial Tree (including access beyond the stem). The instrument is introduced per oral or per nasal when indications consistent with the requirement for the procedure are observed in adult and pediatric patient populations.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K023376 · printed Nov 6, 2026
Cite this page
510(k) Report. “K023376: EB-1830T3, Video Bronchoscope.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K023376
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026